Cal Test Electronics, Inc warrants these products to be free from defective material or workmanship for a period of 1 year from the date of original purchase. Under this warranty, Cal Test Electronics, Inc is lim- ited to repairing this product when returned to the factory, shipping charges prepaid, within the warranty period.
Directive, is marked as being put on the market after August 13, 2005, and should not be disposed of as unsorted municipal waste. Please utilize your local WEEE collection facilities in the disposition of this product. CT2700A 40 kV High-Voltage DMM Probe User Manual...
WARNING - HIGH VOLTAGE - possibility of electric shock. The Conformitè Europëenne (CE) Mark is the European Union’s (EU) mandatory conformity marking for regulating the goods sold within the European Economic Area (EEA). CT2700A 40 kV High-Voltage DMM Probe User Manual...
Disconnect the probe input from the circuit before disconnecting the probe from the oscilloscope. • Do not exceed the voltage rating category rating of the probe. Use only accessories provided with probe. CT2700A 40 kV High-Voltage DMM Probe User Manual...
Pollution - addition of foreign matter, solid, liquid, or gaseous (ionized gases) that may produce a reduc- tion of dielectric strength or surface resistivity. Pollution Degree 2 (P2)- only non-conductive pollution occurs except that occasionally a temporary conductivity caused by condensation is expected. CT2700A 40 kV High-Voltage DMM Probe User Manual...
Introduction Overview The Cal Test CT2700A 40 kV High-Voltage Digitial Multimeter (DMM) probe can be used with anlalog or digital multimeters (DMMs) which accept 4 mm sheathed banana plugs and have an input resistance of 10 MΩ (±1%). The CT2700A is a 1000:1 divider which extends the voltage meansurement capability to 28 kV ACrms.
The end of the Flat Bar Tip can be clamped to the device-under-test (DUT). To secure the Flat Bar Tip to the probe, use the supplied M5 flange nut (8 mm wrench or nut driver) as show in Figure 3. Figure 3 Attaching the Flat Bar Tip CT2700A 40 kV High-Voltage DMM Probe User Manual...
Disconnect the probe tip from the high-voltage source. Disconnect the probe’s ground clip lead. WARNING Disconnect the probe tip from the high voltage source BEFORE disconnecting the alliga- tor ground clip lead. CT2700A 40 kV High-Voltage DMM Probe User Manual...
Dry the probe thoroughly before attempting to make voltage measurements. CAUTION Do not subject the probe to solvents or solvent fumes as these can cause deterioration of the probe body and cables. CT2700A 40 kV High-Voltage DMM Probe User Manual...
X: Indicates that this hazardous substance contained in at least one of the homogeneous materials used for this part is above the limit requirement in GB/T 26572. X: 表示該有害物質在用於該部件的至少一種均質材料中的含量高於 GB/T 26572 中的限量要求。 CT2700A 40 kV High-Voltage DMM Probe User Manual...
Need help?
Do you have a question about the CT2700A and is the answer not in the manual?
Questions and answers