BioCompact II
INSTRUCTIONS FOR USE
MODELS: 210/310/410/610
Original Instructions for use
Original instructions for use
Item nr. 765042589
Revision nr. 20230412
English
b i o l i n e

Advertisement

Table of Contents
loading

Summary of Contents for Gram bioline BioCompact II

  • Page 1 b i o l i n e BioCompact II INSTRUCTIONS FOR USE MODELS: 210/310/410/610 Original Instructions for use Original instructions for use Item nr. 765042589 Revision nr. 20230412 English...
  • Page 2 Quick guide – BioCompact II b i o l i n e Example: Setting the upper limits for the alarms; LHL Press and hold until the display shows LAL Press to select LAL, LHL is now shown in the display Press to select LHL, 25 is shown in the display Press...
  • Page 3: Table Of Contents

    BioCompact II RR310G ........65 easy to access. If the instructions for use is lost, please refer to your local distributor or Gram Scientific for a replacement.
  • Page 4: Cabinet Components

    In the event of technical difficulties or If parts show signs of damage; do not breakdown, always contact Gram BioLine use cabinet and contact Gram BioLine or technical support or a Gram BioLine supplier for further aid. authorized service partner.
  • Page 5: Biocompact Ii 210, 310, 410, 210/210, 310/210

    In the event of technical difficulties or If parts show signs of damage; do not breakdown, always contact Gram BioLine use cabinet and contact Gram BioLine or technical support or a Gram BioLine supplier for further aid. authorized service partner.
  • Page 6: Biocompact Ii 610

    Installation b i o l i n e b i o l i n e Initial setup steps I-1*: Due to safety and operating considerations, the cabinet must not be used outdoors. I-4*: Ambient operating temperature range I-2*: The cabinet should be installed in a dry and sufficiently ventilated area. Minimum ambient Maximum ambient Cabinet...
  • Page 7 I-10-11*: For cabinets equipped with castors, the floor must be level to ensure stable positioning and safe use. When the cabinet is positioned, the 2 front castors should be locked. For replacement brackets, please contact your local Gram BioLine distributor *): Paragraph designation (I-1, I-2 etc) refers to the IQ (Installation Qualification)
  • Page 8 b i o l i n e The cabinet must not be connected to a power source changing door hinging.
  • Page 9: Anti Tilt Bracket

    b i o l i n e Anti tilt bracket I-12*: Cabinets with drawers and/or glass door must be secured to a stable vertical surface, ensuring that the cabinet can The anti-tilt brackets must be fitted when installing the cabinet, ensuring that the users, surroundings and not tip over when the drawers are drawn to the outermost position, or the door is open.
  • Page 10: Surroundings

    b i o l i n e Surroundings I-15*: Do not cover the upper part of the cabinet. I-13-14*: The cabinet must be installed in accordance with the illustrations below. I-16*: Do not use electrical appliances inside the cabinet. The cabinet is not suited for storing items that emit vapours, as this might lead to deterioration of cabinet performance and/or longevity.
  • Page 11: Voltage Free Contact

    b i o l i n e Voltage free contact Placement of voltage free contact I-17*: The illustration shows the three connectors for the relay (used ex. in connecting to CTS or other external monitoring systems). The three connections, are respectively. Common, NO and NC. BioCompact II 610 The moment when voltage is applied the controller draws the relay, this makes it possible for the controller to respond to both high and low alarms, door alarms and power failures.
  • Page 12: Connection To Power

    Never use the cabinet if the plug is damaged. The cabinet should be examined by a Gram BioLine service technician in I-19*: The cabinet is intended for connection to alternating such cases.
  • Page 13: Equipotential Bonding

    b i o l i n e Equipotential bonding Exclusively for the models – BioCompact II 210, 310, 410, 210/210, 310/210 I-22*: For installation in ATEX Cat. 3 Zone 2 areas, it is mandatory to have a equipotential bonding, it is not sufficient to use protective earth through the mains connection.
  • Page 14: Start-Up

    Start-up b i o l i n e The digital display Parameter settings Standby indicator Temperature settings Standby Defrosting Alarms A2, A3, A4, A5 Key pad lock is activated O-1*: Standby • Parameter setting Gives access to the cabinets configurable parameters, such as Press to turn the cabinet on.
  • Page 15: Walkthrough Of Menu

    Error codes b i o l i n e Walkthrough of menu Menu Access Display code Explanation Local alarm settings [° C] Upper alarm limit. Code for activated alarm [A2] - 0 - Door is open [° C] Lower alarm limit. Code for activated alarm [A3] Door alarm “dAd”...
  • Page 16: Local Alarm Settings

    Local alarm settings b i o l i n e b i o l i n e Local high alarm Local alarm settings Local low alarm O-3*: LHL – Setting the upper alarm limit [° C] O-5*: LHd – Setting the delay of the local upper alarm limit [min.] Press and hold for more than 3 seconds Press and hold...
  • Page 17: On/Off Local Door Alarm

    b i o l i n e b i o l i n e On/off local door alarm Local acoustic settings Delay for local door alarm O-7*: dA – Activate/deactivate of local door alarm O-9*: bU – Activation/deactivation of the acoustic local alarms Press and hold for more than 3 seconds Press and hold...
  • Page 18: External Alarm Settings

    External alarm settings b i o l i n e b i o l i n e External high alarm External high alarm delay External low alarm External low alarm delay O-10*: EHL – Setting the external upper alarm limit [° C] O-12*: EHd –...
  • Page 19: On/Off External Door

    b i o l i n e On/off external door External acoustic settings External door alarm delay alarm O-14*: dA – Activation/deactivation of external door alarm O-16*: bU – Activation/deactivation of the acoustic external alarms Press and hold for more than 3 seconds Press and hold for more than 3 seconds Press...
  • Page 20: Parameter Settings

    Parameter settings b i o l i n e Sensor offset Offset of the A-sensor The temperature sensors connected to MPC controller can be offset independently of each other in the parameter cAL. Offset is used in cases where there are deviations in the cabinets actual operation compared to the display and/or control Press and hold for more than 3 seconds measurements by independent temperature monitoring.
  • Page 21: Escorted/Set Alarm Limits

    b i o l i n e Escorted/set alarm limits Defrosts/24 hours ALL – Setting of escorted/set alarm limits O-17*: dEF – Number of defrosts Press and hold for more than 3 seconds Press and hold for more than 3 seconds Press several times until “ALL”...
  • Page 22: Display Sensor

    b i o l i n e Display sensor Electric low temperature protection O-18*: dPS – Display sensor FP – Activation/deactivation of low temperature protection Press and hold for more than 3 seconds Press and hold for more than 3 seconds Press several times until “dPS”...
  • Page 23: Ordinary Use

    Ordinary use b i o l i n e BioCompact II 410 Keep the marked areas in the cabinet (shown on next page) clear of all items, thereby ensuring adequate air circulation, and therein cooling. Do not place items beneath the lowest shelf bracket. All items in the cabinet that are not encapsulated, or wrapped, should be covered to reduce the risk of corrosion of the cabinet and its components.
  • Page 24: Regular Maintenance

    The door gasket should be cleaned regularly with a mild soap solution. If a gasket is to be replaced, please contact your local Gram The compressor compartment and in particular the condenser must be kept free from dust and dirt. This is best done BioLine distributor.
  • Page 25: General Info

    Defective parts must be replaced with original parts from Gram BioLine. Gram BioLine can only guarantee functional and safety requirements on the cabinets, if above mentioned is adhered to.
  • Page 26: Defrost Water

    b i o l i n e Defrost water Access port All BioCompact ll cabinets are equipped with an access port on the back of the cabinets, this can be used to easily fit external sensors. All access ports are constructed in the same fashion, with a conical polystyrene plug (fitted from the back of the cabinet) and a plastic The cabinet creates defrost water, that is directed out into a re-evaporation tray at the back of the cabinet.
  • Page 27: Important

    Disposal b i o l i n e Important Electrical and electronic equipment (EEE) contains materials, components and substances that can be dangerous and harmful to human health and the environment if the waste (WEEE) is not disposed of properly. IMPORTANT! Contact you local BioLine distributor when the cabinet needs to be disposed of.
  • Page 28: Datasheet

    230 VAC, 50 Hz Temperature range +2/+20 °C Temperature range +2/+20 °C Control Unit Gram Control Unit with voltage free contact, alarms and offset function Ambient temperature range +10/+35 °C Ambient temperature range +10/+32 °C Alarms Acoustic and visual, High/Low temperature alarms and door alarm...
  • Page 29: Biocompact Ii Rr310H

    b i o l i n e BioCompact II RR310H BioCompact II RR410H BioCompact II RR310H BioCompact II RR310H BioCompact II RR410H BioCompact II RR410H Solid door Glass door Solid door Glass door Technical specifications Data Technical specifications Data Technical specifications Data Technical specifications Data...
  • Page 30: Biocompact Ii Rf210H

    b i o l i n e BioCompact II RF210H BioCompact II RF310H BioCompact II RF410H BioCompact II RF210H BioCompact II RF310H BioCompact II RF410H Solid door Solid door Solid door Technical specifications Data Technical specifications Data Technical specifications Data Temperature range -25/-5 °C Temperature range...
  • Page 31: Biocompact Ii Rr210/Rr210H

    b i o l i n e BioCompact II RR210/RR210H BioCompact II RR210/RF210H BioCompact II RR210/RR210H BioCompact II RR210/RR210H BioCompact II RR210/RF210H BioCompact II RR210/RF210H Solid door Glass door Solid door Glass door (RR) Technical specifications Data Technical specifications Data Technical specifications Data Technical specifications...
  • Page 32: Biocompact Ii Rr310/Rf210H

    b i o l i n e BioCompact II RF210/RF210H BioCompact II RR310/RF210H BioCompact II RF210/RF210H BioCompact II RR310/RF210H BioCompact II RR310/RF210H Solid door Solid door Glass door (RR) Technical specifications Data Technical specifications Data Technical specifications Data Temperature range -25/-5 °C Temperature range +2/+20 °C and -25/-5 °C...
  • Page 33: Biocompact Ii Rr210G

    b i o l i n e BioCompact II RR210G BioCompact II RR310G BioCompact II RR210G BioCompact II RR210G BioCompact II RR310G BioCompact II RR310G Solid door Glass door Solid door Glass door Technical specifications Data Technical specifications Data Technical specifications Data Technical specifications Data...
  • Page 34: Biocompact Ii Rr410G

    b i o l i n e b i o l i n e BioCompact II RR410G BioCompact II RF210G BioCompact II RF310G BioCompact II RR410G BioCompact II RR410G BioCompact II RF210G BioCompact II RF310G Solid door Glass door Solid door Solid door Technical specifications Data...
  • Page 35: Biocompact Ii Rf410G

    b i o l i n e BioCompact II RF410G BioCompact II RR210/RR210G BioCompact II RF410G BioCompact II RR210/RR210G BioCompact II RR210/RR210G Solid door Solid door Glass door Technical specifications Data Technical specifications Data Technical specifications Data Temperature range -25/-5 °C Temperature range +2/+20 °C Temperature range...
  • Page 36: Biocompact Ii Rr210/Rf210G

    b i o l i n e BioCompact II RR210/RF210G BioCompact II RF210/RF210G BioCompact II RR210/RF210G BioCompact II RR210/RF210G BioCompact II RF210/RF210G Solid door Glass door (RR) Solid door Technical specifications Data Technical specifications Data Technical specifications Data Temperature range +2/+20 °C and -25/-5 °C Temperature range +2/+20 °C and -25/-5 °C...
  • Page 37: Biocompact Ii Rr310/Rf210G

    b i o l i n e BioCompact II RR310/RF210G BioCompact II RR310/RF210G BioCompact II RR310/RF210G Solid door Glass door (RR) Technical specifications Data Technical specifications Data Temperature range +2/+20 °C and -25/-5 °C Temperature range +2/+20 °C and -25/-5 °C Ambient temperature range +10/+35 °C Ambient temperature range...
  • Page 38: Biocompact Ii Rr610H

    230 VAC, 50 Hz Temperature range +2/+20 °C Temperature range +2/+20 °C Control Unit Gram Control Unit with voltage free contact, alarms and offset function Ambient temperature range +10/+43 °C Ambient temperature range +10/+38 °C Alarms Acoustic and visual, High/Low temperature alarms and door alarm...
  • Page 39: Biocompact Ii Rr610G

    b i o l i n e BioCompact II RR610G BioCompact II RF610H BioCompact II RR610G BioCompact II RR610G Solid door Solid door Glass door Technical specifications Data Technical specifications Data Technical specifications Data Temperature range -25/-5 °C Temperature range +2/+20 °C Temperature range +2/+20 °C...
  • Page 40: Biocompact Ii Rf610G

    b i o l i n e b i o l i n e BioCompact II RF610G BioCompact II RF610G Solid door Technical specifications Data Temperature range -25/-5 °C Ambient temperature range +10/+43 °C Software variant K-Value 0.31 W/(m2*K) ATEX marking II 3G Ex nA nC nL llB T2 Gc ATEX certificate –...
  • Page 41: Declaration Of Conformity

    BioCompact II 610 English EC Declaration of Conformity English EC Declaration of Conformity We, Gram Scientific ApS declare under sole responsibility that the following products: We, Gram Scientific ApS declare under sole responsibility that the following products: Range: BioCompact II...
  • Page 42: Accessories Code 69

    Accessories code 69 English EC Declaration of Conformity English EC Declaration of Conformity We, Gram Scientific ApS declare under sole responsibility that the following products: We, Gram Scientific ApS declare under sole responsibility that the following products: Range: BioCompact II (Accessorie code 69)
  • Page 43: Wiring Diagram

    Wiring diagram b i o l i n e BioCompact II 210/310/410 BioCompact II 210/310/410 - with LTP...
  • Page 44: Biocompact Ii Rr610 - With Solid Door

    b i o l i n e BioCompact II RR610 – with solid door BioCompact II RF610...
  • Page 45: Biocompact Ii Rr610 - With Glass Door

    b i o l i n e BioCompact II RR610 – with glass door BioCompact II RR610 – with solid door, with LTP...
  • Page 46: Piping Diagram

    6500 Vojens All rights to this drawing belongs to Gram Scientific ApS, cf. law of copyright. Thus the drawing should not without our written permission be copied, presented or passed over to a third person. Misuse will lead to prosecution.
  • Page 47: Iq & Oq

    Organisation: procedures can vary depending on 2. Operational test of the cabinet Date: application and items stored in the Gram BioLine cabinet. 3. Responsible party Tel: Instructions to users: Deviations from the specifications dictated in the PQ are to be...
  • Page 48 b i o l i n e Installation Qualification – IQ Installation Qualification – IQ Comply Comply Description of installation Reference Attachment Description of installation Notes Reference Attachment Notes in manual in manual Ensure that the inner doors can Ensure the cabinet is installed indoors. page 10 I-18 operate in accordance with the...
  • Page 49 b i o l i n e Operation Qualification – OQ Deviation Report: Comply Description of operation Deviations to the criteria of acceptance are to be documented in the deviation report. A separate deviation Reference Attachment Notes in manual report shall be made for each deviation. Mark the entry with the relevant “-ID” specified in the left column in the test specifications.
  • Page 50 b i o l i n e NOTES: Approval of test results – Installation Qualification (IQ) The steps in the Installation Qualification – IQ were completed with positive results The steps in the Installation Qualification – IQ were completed with negative results ID of steps with negative results: Approval of test results –...
  • Page 51: Performance Qualification

    b i o l i n e Performance Qualification List of names – Persons involved in the test procedure and subsequent report Organisation: Location of installation: Date Name Organisation Signature Model: ________________________ SN: ________________________ Item number: ________________________ (manual) The PQ consists of inspections of the correct operation of the cabinet Person responsible for the cabinet: under predefined conditions and...
  • Page 52 b i o l i n e Deviations from the specifications dictated in the PQ, are to be reported in the deviation report. Deviations from the specifications dictated in the PQ, are to be reported in the deviation report. The PQ is concluded if all criteria of acceptance are approved and the possible deviations are rectified or The PQ is concluded if all criteria of acceptance are approved and the possible deviations are rectified or accepted.
  • Page 53 b i o l i n e Deviations from the specifications dictated in the PQ are to be reported in the deviation report. Deviations from the specifications dictated in the PQ are to be reported in the deviation report. The PQ is concluded if all criteria of acceptance are approved and the possible deviations are rectified or accepted. The PQ is concluded if all criteria of acceptance are approved and the possible deviations are rectified or accepted.
  • Page 54 b i o l i n e Deviations from the specifications dictated in the PQ, are to be reported in the deviation report. Deviations from the specifications dictated in the PQ, are to be reported in the deviation report. The PQ is concluded if all criteria of acceptance are approved and the possible deviations are rectified or accepted. The PQ is concluded if all criteria of acceptance are approved and the possible deviations are rectified or accepted.
  • Page 55 b i o l i n e Deviation Report Approval of test results – Performance Qualification (PQ) Deviations to the criteria of acceptance are to be documented in the deviation report. A separate deviation report shall be made for each deviation. Mark the entry with the relevant “P-ID” specified in the left column The steps in the Performance Qualification –...
  • Page 56 b i o l i n e NOTES: Appendix BioCompact II Tolerances Door Permissable Pull- Permissable Hold-over Hold- Permissable opening deviation down deviation range* over deviation recovery (minutes) time (minutes) REFRIGERATORS (RR) 210 (Solid door) 35 % 210 (Glass door) 10 % 310 (Solid door) 35 %...
  • Page 57 Gram Scientific ApS Aage Grams Vej 1 · 6500 Vojens · Danmark Tel: +45 73 20 13 00 e-mail: info@gram-bioline.com www.gram-bioline.com...