  • Page 1 INC-H Refrigerated Incubator with Humidity Control Please read the User Manual carefully before use, and follow all operating and safety instructions! user manual english...
  • Page 2: Warranty

    User Manual INC-H Refrigerated Incubator with Humidity Control Preface Thank you for purchasing our product. Users should read this manual carefully, follow the instructions and procedures, and be aware of all the preventive measures when using this instrument. Service If help is needed, you can always contact your dealer or Labbox via www.labbox.com.
  • Page 3 1. Introduction The INC-H Refrigerated Incubator with Humidity Control is the ideal testing equipment for simulating natural environments. It comes with 30 segments of program memory and functions such as constant temperature, auto humidity control, day/night cycle, and over-temperature protection. This equipment can be used for a wide...
  • Page 4: Main Technical Parameters

    3. Main Technical Parameters Model INCH-070-001 INCH-150-001 Supply Power 220/50Hz Temperature Range With humidification 10-65 ºC; without humidification: 0-65 ºC Temperature 0.1 ºC Distinguishability Temperature Fluctuation High temperature ±1 ºC; low temperature ±0.5 ºC Humidity Range 45 % ~ 90 % RH Humidity Error ±5 % RH 1800W...
  • Page 5: Working Conditions

    2. Temperature and Humidity Diagram 4. Working Conditions 1. Ambient Temperature: 5 ºC ~ 32 ºC (e.g., if the set temperature is ≤10 ºC, the ambient temperature should be ≤28 ºC). 2. Relative Humidity: ≤80 % RH. 3. Air Pressure: 86 - 10 Kpa. 4.
  • Page 6: Operation Steps

    Operation 1. If starting up the machine for the first time, do not change any parameters unless allowed in the operation manual. 2. Turn off any unnecessary lights or humidifiers. 3. During testing, if the temperature or humidifier is shut off for protection, set the humidifying parameters to 0 when not necessary.
  • Page 7 3.1 Definition T1 Reservation: T1 flashes when entering the reservation timing. The【(2)display window】shows the reservation countdown. T2 Heating: T2 lights up when heating is output. T3 Compressor: T3 lights up when the compressor starts up, and flashes when the compressor is waiting for the start-up delay.
  • Page 8 3.3 Definition of Keys Under normal display, press and hold this button for 2 seconds to manually 【LOCK】key lock or unlock the screen. Under normal display, clicking this button switches the illumination. 【ILLUMINATION】key Under normal display, clicking the button enters the setting value modification interface.
  • Page 9 3.4.3 Set Segment Number or Cycles In the no constant mode (see parameter list -1-U1 for details), when the controller is stopped, press the 【SET】 button for 3 seconds. The 【(1) display window】 will then show “Lc,” and the 【(2) display window】...
  • Page 10 3.4.7 Defrost Function The defrost function can be turned on either automatically or manually. If you want to turn it on automatically (check the parameters in table-4 for details) and then set the interval time and duration of the defrost cycle. On the other hand, if you want to turn it on manually, click the 【DEFROST】...
  • Page 11: Parameters Table

    4. Parameters Table User Parameters Table -1 Parameter Parameter (Range) Factory Parameter Function Description Indication Name Value When “Lc=9”, you can check and modify the Password parameters. 0: Fixed value running mode. 1: Day or night mode, 0 ~ 99 cycles. (0~2) Running mode 2: Program mode, programmable 1 ~ 30 period,...
  • Page 12 Temperature Parameter Table -2 Parameter Parameter (Range) Factory Parameter Function Description Indication Name Value “Lc=103”, allows you to check and modify the Password parameters. If the “measured value > setting value + TH”, then the upper deviation alarm will sound, and the Upper temperature and humidity output will be turned off.
  • Page 13 Humidity Parameter Table -3 Parameter Parameter (Range) Factory Parameter Function Description Indication Name Value To view and modify parameter values, Password enter “Lc=203”. If the “humidity measurement value > set value + HH”, an upper deviation alarm is triggered, and the super-wet alarm Upper deviation relay and humidity output are turned off.
  • Page 14 When controlling for low humidity, if the Low humidity humidity measurement value is ≥ (-10.0~10.0%) No humidification point humidity set value + HF, the controller -5.0 will prevent humidification. Compressor Parameter Table -4 Parameter Parameter (Range) Factory Parameter Function Description Indication Name Value...
  • Page 15 Defrost time interval when “10.0 ºC < temperature set value ≤ 20.0 ºC”. (0~9999min) Defrost interval 2 Note: 0 means there is no automatic defrosting in this section and it can be done manually. Defrost time interval when “20.0 ºC < temperature set value ≤...
  • Page 16 Internal Parameter Table -5 (Range) Parameter Parameter Parameter Function Description Factory Indication Name Value The parameter values can be viewed and Password modified when “Lc=209”. 0: no light; 1: total 3 levels. 2: total of 4; 3: total of 5. Illuminance 4: total of 6 levels.
  • Page 17: Wiring Diagram

    If Pd>0, extend the Pd time and close the water Water level delay after adding water. (-20~20s) Adding water time If Pd < 0, water shortage is detected, and water is added after a delay of Pd time. Humidity 0: No decimals. (0~1) Decimal place 1: Decimal display.
  • Page 18 6. General Faults and Troubleshooting Fault Phenomenon Failure Analysis Troubleshooting 1. Replace the sensor. 1. The sensor is broken. Temperature control instrument 2. Check wiring and connect 2. The sensor wiring is off. displays 0000 or ----- securely. 3. The controller is broken 3.
  • Page 20 www.labbox.com...