Manual Part Number E8770-90037
Agilent TS-5400
Functional Test System
Series IIB
Site Preparation and Getting
Started Manual
Agilent Technologies

Advertisement

Table of Contents
loading

Summary of Contents for Agilent Technologies TS-5400

  • Page 1 Agilent TS-5400 Functional Test System Series IIB Site Preparation and Getting Started Manual *E8770-90037* *E8770-90037* *E8770-90037* *E8770-90037* Manual Part Number E8770-90037 Agilent Technologies...
  • Page 2 Notices © Agilent Technologies, Inc. 1999 - 2003 Manual Part Number subject to Agilent Technologies’ standard commercial license terms, and non-DOD No part of this manual may be reproduced E8770-90037 Departments and Agencies of the U.S. Gov- in any form or by any means (including...
  • Page 3 Safety Summary The following general safety precautions must be observed WARNING: DO NOT REMOVE during all phases of operation of ANY SYSTEM COVER this system. Failure to comply Operating personnel must not with these precautions or with remove system covers. Component specific warnings elsewhere in replacement and internal this manual violates safety...
  • Page 4 Ground the System ASSURE THE EQUIPMENT UNDER TEST HAS To minimize shock hazard, the ADEQUATE INSULATION system chassis must have a BETWEEN THE CABLE hard-wired connection to an CONNECTIONS AND ANY electrical protective earth OPERATOR-ACCESSIBLE ground. The system must also be PARTS (DOORS, COVERS, connected to the ac power mains PANELS, SHIELDS, CASES,...
  • Page 5 Table 1 Safety Symbols and Markings Safety symbols Warning: risk of electric shock. Caution: refer to accompanying documents. Alternating current. Both direct and alternating current. Earth (ground) terminal Protective earth (ground) terminal Frame or chassis terminal Terminal is at earth potential. Used for measurement and control circuits designed to be operated with one terminal at earth potential.
  • Page 6 Measurement. Select your personnel. Contact your country from the drop-down customer engineer through your menus. The Web page that local Agilent Technologies appears next has contact Service Center. information specific for your country. Agilent on the Web...
  • Page 7: Declaration Of Conformity

    Date Ray Corson Product Regulations Program Manager For further information, please contact your local Agilent Technologies sales office, agent or distributor. Authorized EU-representative: Agilent Technologies Deutschland GmbH, Herrenberger Straβe 130, D 71034 Böblingen, Germany Revision: B.02 Issue Date: 16 August 2001...
  • Page 8 Notes:...
  • Page 9: Table Of Contents

    Chapter 3 Connecting Power and Getting Started ..............27 Tools Required ....................27 Connecting AC Mains Power to the Agilent TS-5400 System ........27 Procedure ......................28 Determine the Power Cord Type Needed ............29 Locate the Agilent E1135C PDU ................ 31 Set PDU Mains Disconnect to 0 (Off), Attach Strain Relief to the PDU’s Access Panel ..................
  • Page 10 Contents...
  • Page 11: Preparing Your Site For The System

    Chapter 1 Preparing Your Site for the System The following gives information to prepares your site for using the Agilent TS-5400 Series IIB System. This chapter is separated as follows: • Typical System ....... . page 11 •...
  • Page 12: The System Plan Drawing

    It can also serve to verify physical access. Refer to Appendix A for specific site requirements (structural, environmental, etc.). Figure1-2 shows a layout suggestion for a typical Agilent TS-5400 Series IIB System. Allow 1 meter (3 feet) of space behind the system for service.
  • Page 13: System Power Requirements

    System Power Requirements This section describes the electrical power requirements of the Agilent TS-5400 Series IIB Test Systems using the Agilent E1135C Power Distribution Units (PDUs). It also describes how to connect AC mains power to the system. Figure1-3 shows the location of the power distribution unit in your system.
  • Page 14: Power Recommendations

    The power service must have a mains disconnect installed adjacent to the system to quickly remove power in case of emergency (see "Mains Disconnect Requirements" for more information). The Agilent E1135C PDU, used with newer TS-5400 Systems, contains a mains disconnect. •...
  • Page 15: Mains Disconnect Requirements

    Mains Disconnect Requirements The Agilent 1135C PDU contains a lockable mains disconnect switch. If may also want to install an external mains disconnect circuit breaker. Be sure the circuit breaker is adjacent to the equipment for easy access. Usual mounting is an approved enclosure on a floor-mounted pedestal. Disconnect Circuit Breaker Requirements The disconnect circuit breaker must be: •...
  • Page 16 • Make certain that the wire used for the neutral is the same size as the hot leads or larger. • Do not rely on conduit for the ground; always use at least one wire for the ground-the more strands the better.
  • Page 17: Mains Wiring And Sizes

    VXI Mainframe. For systems without a VXI Mainframe, subtract approximately 500W of power. This comes to approximately 4.2 Amps of current for a line voltage of 120V and 2.4 Amps for a line voltage of 208V. Table 1-1. Agilent TS-5400 System Power Options 05B, 05G 05K Power Phase...
  • Page 18 Table 1-2. Agilent TS-5400 System Power Options 0ED, 05C, 05H, 0E6 Power Phase Power Type Option Country Voltage Three-Phase Delta Japan 200V Brazil, France, S. Korea, Philippines, Taiwan 220V Brazil, France, Philippines 230V United States 240V Source to System Distance in...
  • Page 19 69-76 (225-250) 50 Amps / 1 Pole 16 (6) 76-91 (250-300) 50 Amps / 1 Pole 16 (6) * Includes Safety Grounding Conductor Table 1-5. Agilent TS-5400 System Power Options AWX, 0EB, 0EJ, 0EC Power Phase Power Type Option Country Voltage...
  • Page 20 Table 1-6. Agilent TS-5400 System Power Options AWV, AWZ Power Phase Power Type Option Country Voltage Single-Phase Wye with Neutral United States 120V/208V Brazil, France, Italy, Mexico, Philippines 127/220V Source to System Distance in Full Load Circuit Breaker Wire Size...
  • Page 21: Installing The Keyboard, Mouse And Printer

    Chapter 2 Installing the Keyboard, Mouse and Printer This chapter describes how to install the keyboard shelf, mouse tray and printer tray onto the Test Rack, how to connect the keyboard and mouse, and how to connect an optional strip printer. Tools Required # 2 phillips screwdriver.
  • Page 22: Installation Procedure

    Installation Procedure 1. Locate the keyboard and mouse extension cables hanging inside the Test Rack. Temporarily pull the extension cables through the slot in the keyboard tray and attach the keyboard shelf to the test rack using four 10-32 screws. C ap tive N u ts K eybo ard an d M o u se...
  • Page 23 Note The following figures assume a right-hand mouse and left-hand printer arrangement. You can reverse this orientation if necessary 2. Attach the printer tray to the keyboard shelf. Make sure that the two hooks engage the shelf and the bottom of the tray is inserted into the tab under the keyboard shelf. Printer Tray Figure 2-2.
  • Page 24 3. Attach the mouse tray to the keyboard shelf. Make sure that the two hooks engage the shelf and the bottom of the tray is inserted into the tab under the keyboard shelf. Mouse Tray Figure 2-3. Attach Mouse Tray 24 Installing the Keyboard, Mouse and Printer Chapter 2...
  • Page 25 4. Install the plastic palm rest by slipping it into the slots in the keyboard shelf. Palm Rest Slot Figure 2-4. Install Palm Rest Chapter 2 Installing the Keyboard, Mouse and Printer...
  • Page 26 Figure 2-5. Connect Keyboard and Mouse 6. If you have an optional Strip Printer (not provided by Agilent Technologies) connect its serial cable the one of the PC’s serial ports. Power (ac mains) is provided at the PDU. You can route the ac power cable through the holes underneath the Test System Interface to the back of the system rack.
  • Page 27: Connecting Power And Getting Started

    T10 Torx driver for removing covers (access plates) • Wire strippers to connect cables to the Agilent E1135 PDU Connecting AC Mains Power to the Agilent TS-5400 System For all power options, a customer-supplied power cord/plug assembly must be wired to the system's Agilent E1135C Power Distribution Unit (PDU) by an electrician.
  • Page 28: Procedure

    Procedure The steps on the following pages show in detail, how to determine the power cord needed and how to connect it to the Agilent TS-5400 System. WARNING All steps are to be done by a qualified, licensed Electrician with AC POWER SHUT OFF.
  • Page 29: Determine The Power Cord Type Needed

    In addition to supplying a power cord, you must supply a power cord strain relief and a power plug that mates with your style of power receptacle (see Table 3-1). Table 3-1. TS-5400 Recommended Power Cord Wire Sizes Full Power Cord Wire Size/...
  • Page 30 * Use a larger wire size if the voltage drop between the AC source and system is greater than 2% **Primary countries using the Agilent TS-5400 are shown. If your country is not shown, contact your Agilent Sales and Service office to determine which power option you should use.
  • Page 31: Locate The Agilent E1135C Pdu

    Locate the Agilent E1135C PDU The Agilent 1135C is located in the back of the test stand. The figure below shows a typical location. Agilent 1135C PDU Chapter 3 Connecting Power and Getting Started...
  • Page 32: Set Pdu Mains Disconnect To 0 (Off), Attach Strain Relief To The Pdu's Access Panel

    Set PDU Mains Disconnect to 0 (Off), Attach Strain Relief to the PDU’s Access Panel Note The power cord should slide freely through the strain relief--DO NOT tighten the strain relief at this point. Typical Strain Relief (not supplied) Access Panel Power Cord (not supplied) 32 Connecting Power and Getting Started...
  • Page 33: Connect Power Cord To Pdu

    Connect Power Cord to PDU The figures on the following pages show the wiring configurations for the various power options. Stripping Wires and Torquing Terminals To ensure that all wires are safely connected in the PDU, strip the wires and torque the terminal screws as shown in below.
  • Page 34 220-415V Single-Phase Wye with Neutral Agilent Options: 05B, 05G, 05K Mains Disconnect 10A 10A Connect Mains Here L1 L2 Jumper Note For phase voltage > 240V, loads on the output block are 15mm (0.6 in.) connected L1 to N, and L2 to N. Phase voltage is defined as the voltage potential between phases, L1 - L2.
  • Page 35 208 or 220V Three-Phase Wye Agilent Options: 0EF, 05D Mains Disconnect 10A 10A Connect Mains Here L1 L2 L3 220-240V Single-Phase Earthed Agilent Options: 05M, 05F, 05J Mains Disconnect 10A 10A Connect Mains Here Jumper Chapter 3 Connecting Power and Getting Started...
  • Page 36 200V, or 220-240V Single-Phase Non-Earthed Agilent Options: 0EB, 0EJ, 0EC, AWX, 0EA, 0EN Mains Disconnect 10A 10A Connect Mains Here Jumper 120, 127, 220V Single-Phase Wye with Neutral Agilent Options: AWX, AWZ Mains Disconnect 10A 10A Connect Mains Here L1 L2 Jumper 36 Connecting Power and Getting Started Chapter 3...
  • Page 37: Replace Access Panel, Tighten Strain Relief Onto Power Cord

    Replace Access Panel, Tighten Strain Relief onto Power Cord A. Replace Access Panel B. Tighten Strain Relief onto Power Cord Chapter 3 Connecting Power and Getting Started...
  • Page 38: Connect Safety Ground

    Connect Safety Ground If the Agilent TS-5400 system is hardwired to the AC mains and includes a permanent earth ground, no other safety grounding is necessary. However, if the Agilent TS-5400 is connected to the AC mains by means of a plug/socket connection, a permanent earth ground must be supplied to reduce the risk of electric shock.
  • Page 39: Attach Power Plug (Not Supplied)

    Attach Power Plug (not supplied) The figure below shows a typical power plug containing strain relief. Strain Relief Typical Power Plug (not supplied) Chapter 3 Connecting Power and Getting Started...
  • Page 40: Plug-In Power Cord, Set Pdu Mains Disconnect To 1 (On)

    Plug-In Power Cord, Set PDU Mains Disconnect to 1 (On) Turn On the System Move the switch on the top front of the system rack to 1 (On) to turn on the system. 40 Connecting Power and Getting Started Chapter 3...
  • Page 41: Selecting Agilent Testexec Sl

    Start | Programs | Agilent TestExec SL 5.1 | TestExec SL 5.1 Getting Started Examples The TS-5400 Online Help contains getting started examples for E8780/8786 systems. To access the getting started examples, from TestExec SL, click: Tools | TS-5400 Online Help The help file opens up to a selection of getting started examples.
  • Page 42 Notes: 42 Connecting Power and Getting Started Chapter 3...
  • Page 43: Index

    Index Mains Wiring and Sizes Mouse, Installing Agilent TS-5400 System drawing plan mains disconnect requirements mains wiring and sizes Power Recommendations power recommendations Printer, Installing power requirement typical Agilent TS-5500 System Safety Ground Information connecting AC mains power Site Configuration for AC Power...