Rayto RT-7600S User Manual

Rayto RT-7600S User Manual

Auto hematology analyzer
Table of Contents

Advertisement

RT-7600S
Auto Hematology Analyzer
User's Manual
Copyright: Rayto Life and Analytical Sciences Co., Ltd.
Address: C&D/4F, 7
th
Xinghua Industrial Bldg, Nanhai Rd, Nanshan,
Shenzhen 518067, P.R. China
Rev: 1.3e

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the RT-7600S and is the answer not in the manual?

Questions and answers

Cyrus Anceno
February 16, 2025

The display is only white in color..no other details seen

User image 67b1cd0d988f6

Summary of Contents for Rayto RT-7600S

  • Page 1 RT-7600S Auto Hematology Analyzer User’s Manual Copyright: Rayto Life and Analytical Sciences Co., Ltd. Address: C&D/4F, 7 Xinghua Industrial Bldg, Nanhai Rd, Nanshan, Shenzhen 518067, P.R. China Rev: 1.3e...
  • Page 2 In order to get best test result, before doing clinical test, you shall be familiar with this analyzer and its performance. This user’s manual is the using guidline of RT-7600S Auto Hematology Analyzer from Rayto, including installation, daily test, QC and daily maintenance.
  • Page 3 RT-7600S Auto Hematology Analyzer User’s Manual abnormal running of instrument, the test results may be not reliable, also it can break the parts of analyzer even endanger the health of operator. Main graphical representations used on this instrument Temperature limitation...
  • Page 4: Warnings And Security Attentions

    If the smoke, peculiar smell and sound appears while using, the power supply must be immediately cut off. Meanwhile, the inspection apply shall be immediately presented towards the distributors or agents of Rayto company. If the instrument continues to be used under this circumstance, the misfire, electric shock or casualties of personnel might be caused.
  • Page 5 RT-7600S Auto Hematology Analyzer User’s Manual If the hands or skins are touched with the reagents, they shall be immediately  washed to be clean using a large amount of clear water.  The waste products such as used test tubes, other instruments and consumables, etc shall be appropriately handled as the medical waste products or infectious waste products.
  • Page 6: Table Of Contents

    RT-7600S Auto Hematology Analyzer User’s Manual Contents How to use this manual................2 Main graphical representations used on this instrument....... 3 Warnings and security attentions............... 4 1 Instrument Introduction................9 1.1 Product introduction....................9 1.1.1 Product name: Auto Hematology Analyzer..............9 1.1.2 Model: RT-7600S......................9 1.1.3 Test items..........................
  • Page 7 RT-7600S Auto Hematology Analyzer User’s Manual 4 Quality Control..................20 4.1 L-J QC........................20 4.1.1 L-J QC setting........................20 4.1.2 L-J QC run.......................... 21 4.1.3 L-J QC list........................... 21 4.1.4 L-J QC chart........................22 4.2 X-B QC........................23 4.2.1 X-B QC setting........................24 4.2.2 X-B QC list..........................
  • Page 8 RT-7600S Auto Hematology Analyzer User’s Manual 10.1.1 Startup and shutdown process..................45 10.1.2 Automatic rinse........................ 45 10.1.3 Clean instrument surface....................45 10.2 Maintenance program..................45 10.2.1 Back flush......................... 45 10.2.2 High voltage pulse......................45 10.2.3 Drain chambers....................... 45 10.2.4 Drain pipeline........................45 10.2.5 Remove Blockage......................
  • Page 9: Instrument Introduction

    RT-7600S Auto Hematology Analyzer User’s Manual 1 Instrument Introduction 1.1 Product introduction 1.1.1 Product name: Auto Hematology Analyzer 1.1.2 Model: RT-7600S 1.1.3 Test items This analyzer is using impedance principle (colorimetry for hemoglobin measurement) to categorize and count blood cells in blood. Test items are shown in Table 1-1.
  • Page 10: Front View

    RT-7600S Auto Hematology Analyzer User’s Manual 1.3.1 Front view Fig.1-1 Front view ① Display screen: Display the software interface ② Indicator light: yellow at startup, turn red when starting sample test, turn yellow after test. ③ Feed key: built-in printer feeds paper outward ④...
  • Page 11 RT-7600S Auto Hematology Analyzer User’s Manual RS-232 serial port: to connect with data receiving devices Network interface Power interface: to connect with external power supply Power switch: switch instrument power 10) Grounding hole: used for instrument grounding 11) Cleanser port...
  • Page 12: Installation

    Note: Product packing feature is as per packing list, in case of inconformity, please contact seller. 2.2 Installation environment In order to ensure instrument work normally, please choose working place compliant with following condition to place RT-7600S Auto Hematology Analyzer:  No direct sunlight; ...
  • Page 13: Reagent

    RT-7600S Auto Hematology Analyzer User’s Manual 2.4 Reagent The instrument needs to use Lyse, cleanser, diluents for measurement and maintenance. To ensure accuracy of test result, please use reagents accompanied with instrument. 2.4.1 Connection of Lyse and cleanser From reagent packing carton, take out lyse and cleanser respectively, open the bottle lids, and place them in instrument reagent booth horizontally.
  • Page 14: Sample Test

    Check whether built-in or external printer paper is sufficiently installed in right position. 3.2 Startup Switch on the RT-7600S, indicator light is on, system begins to check component status, transfer parameter needed by front end, run blank tests, as shown in figure below: Fig.3-1 Startup program...
  • Page 15: Blood Sample Collection

    RT-7600S Auto Hematology Analyzer User’s Manual Blank test only shows test results of WBC, RBC, HGB, HCT and PLT. The operator can run blank test at any time required. Press “Aspiration Key” to start blank test. During test, the indicator light turns red, status column on top of screen will display test status.
  • Page 16: Blood Sample Analysis

    RT-7600S Auto Hematology Analyzer User’s Manual prolonged storage, or inferior mixing will influence test result accuracy. 3.4 Blood sample analysis Prior to sample analysis, it is recommended to take one QC test of instrument, see detail operation in Chapter 4 “Quality Control”.
  • Page 17: Blood Sample Counting

    RT-7600S Auto Hematology Analyzer User’s Manual Fig.3-4 System soft keyboard When soft keyboard is used, the cursor has to be moved into input box, then click to input. “Caps” is to toggle between uppercase and lowercase, click “Close” to close soft keyboard.
  • Page 18: Parameter Prompt Message

    RT-7600S Auto Hematology Analyzer User’s Manual Fig.3-6 Blood sample counting screen If “Print instantly” setting is selected, the instrument will print analysis result automatically after test ends. If ”Transmit instantly” setting is selected, the instrument will transfer analysis result to the computer by RS-232 serial port after test ends.
  • Page 19: Histogram Prompt Message

    RT-7600S Auto Hematology Analyzer User’s Manual L-: indicates that parameter test result is lower than minimum of predetermined pathogenic value. indicates that parameter test result is less or equal to maximum pathogenic value, greater than maximum of normal value. indicates that parameter test result is greater or equal to minimum of pathogenic value, less than minimum of normal value.
  • Page 20: Print

    RT-7600S Auto Hematology Analyzer User’s Manual Fig.3-8 Adjust confirmation Click “OK” button to save adjust result, click “Cancel” button to cancel this adjustment result. 3.4.6 Print Before sample test, please set up printer and print format in system setting module, see detail in Chapter 7 “System Setting”.
  • Page 21: Quality Control

    RT-7600S provides three QC methods, L-J QC, X-B QC and X-R QC. 4.1 L-J QC In main screen, click “QC” button, choose 20 QC files of “L-J QC”. The system can control 20 parameters at a time.
  • Page 22: L-J Qc Run

    Send: use instrument RS-232 serial port to transfer QC data of QC file currently selected to designated receiver. Prior to transfer, please use RS-232 serial port cable to connect RT-7600S to equipment ready to receive data, start up and set up receiver software parameters. For setting of instrument transferring terminal parameter, please refer to Chapter 7 “System Setting”.
  • Page 23: L-J Qc Chart

    RT-7600S Auto Hematology Analyzer User’s Manual Fig.4-4 L-J QC list The instrument will display 20-parameters QC result, if some parameters are not controlled, the corresponding column shall be blank. Send function can transfer part or all of the L-J QC data to external equipment. Use mouse to click list column, select one group of QC data.
  • Page 24: X-B Qc

    RT-7600S Auto Hematology Analyzer User’s Manual Fig.4-6 L-J QC chart  Lot No.: lot Number of QC sample corresponding to QC file.  Expiry: validity period of QC sample.  Time: test time of datum corresponding to QC point. ...
  • Page 25: X-B Qc Setting

    Send: use instrument RS-232 serial port to transfer X-B QC setting data to designated receiver. Before sending, please use RS-232 serial port cable to connect RT-7600S to equipment to receive data, start up and set up receiver software parameters. For instrument transfer terminal parameter setting, please refer to Chapter 7 “System...
  • Page 26: X-B Qc Chart

    RT-7600S Auto Hematology Analyzer User’s Manual Fig.4-8 X-B QC list The instrument displays 20-parameter QC result, if some parameters are not controlled, the corresponding column shall be blank. Upper limit of QC data columns of X-B QC list shown at a time is 200, if data are more than 200 columns, they can be viewed by switching using “...
  • Page 27: X-R Qc

    Send: use instrument RS-232 serial port to transfer QC data of currently selected QC file to designated receiver. Before sending, please connect RT-7600S to equipment to receive data with RS-232 serial port cable, start up and set up receiver software...
  • Page 28: X-R Qc Run

    RT-7600S Auto Hematology Analyzer User’s Manual parameters. For instrument transfer terminal parameter setting, please refer to Chapter 7 “System Setting”.  Help: to display system help.  Exit: return to main screen. 4.3.2 X-R QC run After QC parameter setting of selected QC file, QC analysis of this QC file can begin. In main screen, select QC file, go to QC run screen, as shown below: Fig.4-11 X-R QC run...
  • Page 29: X-R Qc Chart

    RT-7600S Auto Hematology Analyzer User’s Manual Fig.4-12 X-R QC list The instrument will display 20-parameter QC result. Click “Send” button, some or all QC results can be transferred to external receiver. Operation method is the same as data transfer in “4.1.3 L-J QC list”.
  • Page 30 RT-7600S Auto Hematology Analyzer User’s Manual  Seq#.: sequence number of current QC point Each screen displays two-parameter QC data distribution. QC chart of each parameter is composed of X chart and R chart, reflecting QC data mean and range fluctuation respectively.
  • Page 31: Calibration

    Commercial calibration sample can be used to calibrate this instrument. All mathematic calculations related with calibration can be done by instrument automatically, calibration coefficient is automatically saved. RT-7600S has three sets of calibration coefficients, anti-coagulating whole blood, anti-coagulating peripheral blood and pre-diluted peripheral blood;...
  • Page 32: Auto-Calibration

    RT-7600S Auto Hematology Analyzer User’s Manual Fig.5-1 Manual calibration Manual calibration procedure is as follows: In Sample Test menu, use calibration sample to test several times (at least three). Record the tested data. Calculate new calibration coefficient Calculate new calibration coefficient as per formula below: ...
  • Page 33 RT-7600S Auto Hematology Analyzer User’s Manual Fig.5-2 Auto-calibration At first, calibration sample is to be set up, click “Setting” button, open auto-calibration setting menu, as shown below: Fig.5-3 Auto-calibration setting  Lot No.: lot number of calibration sample, input and save, on next logon, you can select from dropdown list, check or modify data of this lot no.
  • Page 34 RT-7600S Auto Hematology Analyzer User’s Manual follows: Have calibration sample ready, press Aspiration key to start test. After test, calibration result will be shown in current test number column. Repeat test, the instrument allows 20 runs at most. Statistic results and new calibration coefficient will be shown in table below frame after each test automatically.
  • Page 35: History Data

    RT-7600S Auto Hematology Analyzer User’s Manual 6 History Data After test of each sample, the system automatically saves test data; RT-7600S can save parameter test values and histograms of at most 35000 samples. User can check parameter test data and histograms of all samples, print, delete, modify and retrieve sample data of designated condition.
  • Page 36 RT-7600S Auto Hematology Analyzer User’s Manual Click “Cancel” button, the system will not select any record.  Send It provides data transfer function, data selecting methods: a) use select function. b) Use mouse to click list column, select one history data; using keyboard Ctrl key or Shift key, you can select multiple columns.
  • Page 37 RT-7600S Auto Hematology Analyzer User’s Manual  Detail “Detail” function is used to check detail data message of a certain sample, select one record in history list, click “Detail” button, and go to Sample Detail Message Review menu as follows: Fig.6-5 Sample history data...
  • Page 38: System Setting

    RT-7600S Auto Hematology Analyzer User’s Manual 7 System Setting System setting is used to set up system parameters, in main screen, click “Setting”, and go to System Setting menu as follows: Fig.7-1 System setting 7.1 General setting  Serial number: machine serial number, used to identify one instrument.
  • Page 39: Other

    RT-7600S Auto Hematology Analyzer User’s Manual sample number is above this value, the instrument will run cleaning procedure one time. If it is set zero, it will not execute automatic cleaning. Pre-diluted test mode inform: select the box, before pre-diluted peripheral blood test, inform message will popup.
  • Page 40: Normal Limits

    RT-7600S Auto Hematology Analyzer User’s Manual administrator logon is as follows: Fig.7-4 Other setting Click “Log off” button, the user identity will be changed to common user. 7.2.2 Normal limits Click “Normal Limits” button, open setting menu as follows: Fig.7-5 Normal value setting...
  • Page 41: Panic Limits

    RT-7600S Auto Hematology Analyzer User’s Manual 7.2.3 Panic limits When parameter value exceeds normal value range, it does not definitely indicate that the subject is patient, when it falls within panic limits, the conclusion has to be made on specific case.
  • Page 42: Communication

    RT-7600S Auto Hematology Analyzer User’s Manual Select appropriate unit, and click “OK” to save and exit, click “Cancel”, the Units will not be changed. 7.2.5 Communication Communication setting is used to set up each parameter value for serial port transfer interface, to check or modify these settings, click “Communication”...
  • Page 43: Doctor Information

    RT-7600S Auto Hematology Analyzer User’s Manual Append: click “Append” button to add one row of blank column, input department name in department column, click “Append” button, the added department message can be saved. Modify: Select the column you want to modify, input new department name, click “Modify”...
  • Page 44: System Information

    RT-7600S Auto Hematology Analyzer User’s Manual 8 System Information System message provides several types of message about system; user can check system status through system message function module. 8.1 System status On system message menu of main screen, click “System Info” key, the click “system status”, go to system status menu as follows: Fig.8-1 system status...
  • Page 45: Statistic Information

    RT-7600S Auto Hematology Analyzer User’s Manual 8.3 Statistic information Statistic information includes instrument running time after startup and total instrument running time, in system message menu of main screen, click “Statistic”, it will go to statistic information menu as below:...
  • Page 46: Shutdown

    RT-7600S Auto Hematology Analyzer User’s Manual 9 Shutdown After work every day, you must run shutdown program before turning off instrument power. During shutdown, the instrument will perform routine maintenance and rinse test tubing. In main screen, click “Shut down” button, the system will pop up confirmation dialog box as below: Fig.9-1 Shut down confirmation...
  • Page 47: Service

    RT-7600S Auto Hematology Analyzer User’s Manual 10 Service RT-7600S Fully-automatic Hematology Analyzer is a clinical precision analyzer; in order to maintain good instrument condition, get reliable test result, and reduce instrument fault rate, the instrument shall be maintained routinely, this chapter introduces related maintenance operation.
  • Page 48: Prime

    While component is moving, carefully watch whether component movement is smooth and uniform, if there is still anomaly after repeated tests, please contact Rayto customer service department. Click “Valves” button, open solenoid valve test screen as shown below:...
  • Page 49: Debug

    Is for the engineer to test and debug the instrument. 10.2.11 Egineering Just for the engineer to sets up system parameters. Warning: engineers not trained by Rayto Company are refrained from using this function, otherwise test result accuracy will be affected, or even normal test can not be performed.
  • Page 50: Stop Use

    RT-7600S Auto Hematology Analyzer User’s Manual 10.2.13 Stop use If the instrument will not be used for more than two weeks or to be packed for transportation, please conduct in following steps: In “Service” menu, select “Stop Use”, complete operations according to the message given by system, when screen displays it’s ready to shut down, shut off...
  • Page 51: Troubleshooting

    RT-7600S Auto Hematology Analyzer User’s Manual 11 Troubleshooting This chapter describes common faults of instrument and their solutions, if failing to eliminate faults as per tips in this chapter or needing more detailed data, please contact Rayto customer service department. Fault...
  • Page 52: Substance Or Element In Product

    RT-7600S Auto Hematology Analyzer User’s Manual AppendixXI Name and concentration of toxic and hazardous substance or element in product Toxic / hazardous substance or element Component name Hexavalent Polybrominated Lead Mercury Cadmium Polybrominated chromium diphenyl ethers (Pb) (Hg) (Cd) biphenyls (PBB)
  • Page 53: Appendixii List Of Product Tubes

    RT-7600S Auto Hematology Analyzer User’s Manual AppendixII List of product tubes Tube OD(mm) Length(mm) Description ID (mm) (mm) (mm) (mm) Tube76- 1 1.30 3.00 Connect the sampling needle and piston 3. Tube 76-2 1.52 3.22 Connect the temperature sensor and piston 1.