Kangaroo Joey Quick Reference Manual

Kangaroo Joey Quick Reference Manual

Enteral feeding pump

Advertisement

Quick Links

QUICK
REFERENCE
GUIDE
FLUSHING MODE
OPERATION
Kangaroo
Joey
TM
Enteral Feeding Pump
Below are some of the probable causes of alarm conditions and how to correct them.
SYMPTOM PROBABLE CAUSE
Pump in Holding mode
Hold Error
for more than 10 minutes.
Occlusion in pump-patient
Flow Error
line (downstream occlusion).
Empty bag or occlusion
Feed Error
in bag-pump line
(upstream occlusion).
Pump Set
Empty bag or occlusion
Use >24 Hrs
in flush line.
Flow Error
Pump Set in use beyond
Use > 24 Hrs
recommended length of time.
Pump Set
Pump Set not properly
Dislodged
loaded to pump.
Battery Low
Battery charge too low.
Pump Set tubing not
properly loaded on rotor,
Rotor Error
or other unusual rotor
operating condition.
General error caused by
System Error
many factors. See Manual
for error list.
For additional operating instructions and information, see the operating manual
for the Kangaroo™ Joey Enteral Feeding Pump "Flushing Mode Operation."
© 2018 Cardinal Health. All Rights Reserved. CARDINAL HEALTH, the Cardinal
Health LOGO are trademarks of Cardinal Health and may be registered in the
US and/or in other countries. 01/2018 - H9818 - WF#2098506
15 Hampshire Street
800.962.9888
Mansfield, MA 02048
508.261.8000
CORRECTIONS
Press Continue to return to Holding screen, or press and hold the
Power button to turn off. Several options are available from the
Holding screen, such as Run and Adjust Settings.
Check for and correct the occlusion in the line between pump and
patient. Replace Pump Set if error can't be resolved. Check the valve
pocket in the pump set loading area for dampness or dirt.
Check for empty bag and refill, or check for and correct the occlusion
in the line between the pump and feed bag. Replace Pump Set if error
can't be resolved.
Check for empty bag and refill, or check for and correct the occlusion
in the line between the pump and flush bag. Replace Pump Set if error
can't be resolved.
Informational warning message blinks on upper left of screen. It's
recommended that the Pump Set be replaced.
Check black ring retainer (MISTIC) and valve; reload Pump Set and
restart pump to continue. Replace Pump Set if error can't be resolved.
Immediately plug pump into A/C outlet to recharge battery and to
provide power for pump.
Check that the tubing is loaded on the rotor and that it's not
damaged, torn, etc. Reload tubing or replace Pump Set.
Power down pump and attempt restart to clear error. Call customer
service, 800.962.9888, with screen code if error
can't be resolved.

Status LEDs

Red: Error
Yellow: Hold/Pause
Green: Normal Operation
Programming and Menu
Power Button
Selection Buttons

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Joey and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Kangaroo Joey

  • Page 1 For additional operating instructions and information, see the operating manual for the Kangaroo™ Joey Enteral Feeding Pump "Flushing Mode Operation." © 2018 Cardinal Health. All Rights Reserved. CARDINAL HEALTH, the Cardinal Health LOGO are trademarks of Cardinal Health and may be registered in the US and/or in other countries.
  • Page 2 Select Feed VTBD. Use the For additional operating instructions and information, see the operating manual for the Kangaroo™ Joey Enteral Feeding Pump, Flushing Mode Operation.

Table of Contents