Rossmax SA210 - Handheld Pulse Oximeter Manual

Rossmax SA210 - Handheld Pulse Oximeter Manual

Advertisement

Rossmax SA210 - Handheld Pulse Oximeter Manual

Intoduction

Rossmax Handheld Pulse Oximeter is used to measure oxygen saturation in blood (SpO2) and pulse rate, also to issue warnings immediately. It is a non-invasive device intended for spot-check of adults, child, and infants with corresponding probe applied at home, hospital and clinics. The probe contains a dual light source and a photodetector.


Consult the accompanying documents. Please read this manual carefully before use. Please be sure to keep this manual.

Name/ Functions of each part

Name/ Functions of each part

No. Item Purpose
1 Pulse strength
2 Probe icon
3 LCD backlight button For light on/ off
4 Sound button

For Heartbeat sound

loud volume
lower volume
silence

5 Time Shows the time
6 SpO2 %SpO2 icon High SpO2
Low SpO2
7 Pulse Rate High Pulse Rate
Low Pulse Rate
8 UP/DOWN button For adjust/view all setting/ memory
memory icon
9 Set button For view/ adjust default setting
10 Alert button Alert off / Alert pause
11 Power button Power on/ off
12 Battery compartment For install or replace batteries
Full batteries
out of batteries
13 it's for data transmission.

Battery installation

  1. Use thumb to slide battery cover out.
  2. Insert or replace 4 "AA" batteries according to the (+/-) polarity.


Need to replace when the batteries icon " " is blinking on display/ while pressed the function button and nothing appears on display.

Batteries may leak or explode if used or disposed of improperly. Remove batteries if the device will be stored for long time. Do not use different types or brand of batteries at the same time.

Battery installation

Probe connection

Rossmax PA100/ PC100 or compatible probe is used. (Please install carefully.)


It may damage the efficiency of the device if not apply with a Rossmax compatible probe.

Probe connection

How to measure

  1. Press the Power On button for 1 second, when the device activates, the beep sound will last for 2 second.
    warning Note: After the device activates, the software version will pop up directly. For first SA210 All symbols time activates, it will directly jump to time setting. (Please refer to setting instruction.)
  2. Information of software version appears; insert one finger to probe, nail side up, SA210 All symbols for self-test.
    warning Notes: The device will turn itself off automatically after 1 minute idling with two beep sounded.
  3. The pulse strength shows " ", pulse oximeter begins its measurement.
    warning Notes: The heartbeat is sounded though the buzzer. If need to become silence mode, press the sound button and the LCD screen will have the sound icon shown. If need the heartbeat sound, press the sound button to exit.
  4. Your SpO2 and pulse rate values will appears on the screen after few seconds.

warning Notes:

  1. Don't remove your finger until the measurement is completed.
  2. Any other problems or unrecognized icon, please refer to trouble shooting.

Preview Default Setting

Preview Default Setting

  1. Press the button 1 second to preview default setting value.
  2. Press to exit preview default mode or wait for 3 second for automatically.

How to Change Default Setting

Press the button 2 second to enter the set up mode and change default. For jump to next setting, press after finishing. To exit, press or wait for 10 second.

SPO2 mode

SPO2 mode

  • To setup Max SpO2 rate, press to enter, and it shows as picture and press button for adjustment. To set up Min SpO2 rate, press , and it shows as picture and same as max setting.

Pulse Rate mode

Pulse Rate mode

  • To setup Max Pulse rate, press to enter, and it shows as picture and press button for adjustment. To set up Min Pulse rate, press , and it shows as picture and same as max setting.

Memory interval mode

Memory interval mode

  1. Optional of SA210: 5 / 10 / 30 / 60 minutes.
  2. While the memory mode, the memory interval mode appears on the screen. To change the default, press to change (5)-(10)-(30)-(60) or (60)- (30)-(10)-(5) minutes.

Sound volume control

Sound volume control

  • For the control of sound volume, for loud volume, for lower volume. To change the sounds volume, Press to change --> .

Time Setup

Time Setup

  • To set up the time, it starts from year as picture, and press to adjust. Following the same frequency for month, date, hour to minutes.


Press or wait for 10 seconds to exit, the setup value will be reserved. Please check the setting value before use.

Memories reading and delete

Enter the memory mode with the or to read the SpO2 and pulse rate in memory mode.
 Memories reading and delete - Part 1

warning Note: When entering the memory reading mode, the will appear on the screen.

Press for two second during the memory reading mode, it will enter the memory delete mode. The screen show and means not to delete any data, while the screen withand, it means to delete data.


The deleted data could not be restored.

 Memories reading and delete - Part 2

Audio Signals

No Name LCD Display Sound Control
1 Power on


All symbols


software version


start measure

Beep for 2 second Press for 1 second
2 Pulse search The icon flickers X X
3 Max./ Min SpO2 and Pulse rate warning orand yellow backlight flickers Beep-beep sounded repeatedly Default
Mute for temporary, will beep-beep sounded after two minutes. mode
Mute before power off mode
4 Probe failure alert shows on the screen and yellow back light flicker Beep-beep sounded repeatedly X
5 Low battery alert shows on the screen and yellow backlight flickers. Beep-beep-beep sounded repeatedly for 1 minute and power off/ x
6 Automatic Off X Beep-beep sounded and
then power off
After pulse is undetectable for around 1 minutes.
7 Unable measure Blood saturation & pulse rate appears (- -)
and yellow backlight flickers.
Beep-beep sounded
repeatedly.
X
8 Data Updated Period Less than 2 seconds. X X
9 Alert Condition Delay Less than 8 seconds for SpO2. Less than 16 seconds for pulse rate. X X
10 Alert Signal Delay Less than 1 second for SpO2 and pulse rate X X

Specification

Specification

Troubleshooting

Symptoms Check points Corrections
SpO2 or Pulse rate cannot displayed The icon "- -" shows on the screen Place the finger properly and try again
This icon shows on the screen means probe dysfunction Be sure"Rossmax" probe is connected to the device correctly.
Applied finger improperly Place the finger properly and try again
SpO2 or Pulse rate are not displayed stably Finger is shaking or body is moving Keep body steady
Applied finger improperly Place the finger properly and try again
No display when the bottom is pressed Batteries run down Replace with new batteries
Batteries inserted incorrectly Re-insert batteries
The display disappears
suddenly
The device will auto power off when it gets no signal Normal
Low battery Replace with new batteries

warning Note: If the unit does not work, return it to your dealer. Under no circumstance should you disassemble and repair the unit by yourself.

Documents / Resources

Download manual

Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.

Download Rossmax SA210 - Handheld Pulse Oximeter Manual

Advertisement

Need help?

Need help?

Do you have a question about the SA210 and is the answer not in the manual?

Questions and answers

Table of Contents