Rossmax Handheld Pulse Oximeter is used to measure oxygen saturation in blood (SpO2) and pulse rate, also to issue warnings immediately. It is a non-invasive device intended for spot-check of adults, child, and infants with corresponding probe applied at home, hospital and clinics. The probe contains a dual light source and a photodetector.
Consult the accompanying documents. Please read this manual carefully before use. Please be sure to keep this manual.
No. | Item | Purpose |
1 | Pulse strength | ![]() |
2 | Probe icon | ![]() |
3 | LCD backlight button | For light on/ off |
4 | Sound button | For Heartbeat sound
|
5 | Time | Shows the time |
6 | SpO2 %SpO2 icon | ![]() ![]() |
7 | Pulse Rate ![]() | ![]() ![]() |
8 | UP/DOWN button | For adjust/view all setting/ memory ![]() |
9 | Set button | For view/ adjust default setting |
10 | Alert button | ![]() ![]() |
11 | Power button | Power on/ off |
12 | Battery compartment | For install or replace batteries ![]() ![]() |
13 | ![]() | it's for data transmission. |
Need to replace when the batteries icon " " is blinking on display/ while pressed the function button and nothing appears on display.
Batteries may leak or explode if used or disposed of improperly. Remove batteries if the device will be stored for long time. Do not use different types or brand of batteries at the same time.
Rossmax PA100/ PC100 or compatible probe is used. (Please install carefully.)
It may damage the efficiency of the device if not apply with a Rossmax compatible probe.
Notes:
Press the button 2 second to enter the set up mode and change default. For jump to next setting, press
after finishing. To exit, press
or wait for 10 second.
Press or wait for 10 seconds to exit, the setup value will be reserved. Please check the setting value before use.
Enter the memory mode with the or to read the SpO2 and pulse rate in memory mode.
Note: When entering the memory reading mode, the
will appear on the screen.
Press for two second during the memory reading mode, it will enter the memory delete mode. The screen show
and
means not to delete any data, while the screen with
and
, it means to delete data.
The deleted data could not be restored.
No | Name | LCD Display | Sound | Control |
1 | Power on | | Beep for 2 second | Press ![]() |
2 | Pulse search | ![]() ![]() | X | X |
3 | Max./ Min SpO2 and Pulse rate warning | ![]() ![]() | Beep-beep sounded repeatedly | Default |
Mute for temporary, will beep-beep sounded after two minutes. | ![]() | |||
Mute before power off | ![]() | |||
4 | Probe failure alert | ![]() | Beep-beep sounded repeatedly | X |
5 | Low battery alert | ![]() | Beep-beep-beep sounded repeatedly for 1 minute and power off/ | x |
6 | Automatic Off | X | Beep-beep sounded and then power off | After pulse is undetectable for around 1 minutes. |
7 | Unable measure | Blood saturation & pulse rate appears (- -) and yellow backlight flickers. | Beep-beep sounded repeatedly. | X |
8 | Data Updated Period | Less than 2 seconds. | X | X |
9 | Alert Condition Delay | Less than 8 seconds for SpO2. Less than 16 seconds for pulse rate. | X | X |
10 | Alert Signal Delay | Less than 1 second for SpO2 and pulse rate | X | X |
Symptoms | Check points | Corrections |
SpO2 or Pulse rate cannot displayed | The icon "- -" shows on the screen | Place the finger properly and try again |
This icon ![]() | Be sure"Rossmax" probe is connected to the device correctly. | |
Applied finger improperly | Place the finger properly and try again | |
SpO2 or Pulse rate are not displayed stably | Finger is shaking or body is moving | Keep body steady |
Applied finger improperly | Place the finger properly and try again | |
No display when the ![]() | Batteries run down | Replace with new batteries |
Batteries inserted incorrectly | Re-insert batteries | |
The display disappears suddenly | The device will auto power off when it gets no signal | Normal |
Low battery | Replace with new batteries |
Note: If the unit does not work, return it to your dealer. Under no circumstance should you disassemble and repair the unit by yourself.