Introduction
Rossmax Fingertip Pulse Oximeter SB210 is used to measure arterial oxygen saturation (% SpO2) of hemoglobin and pulse rate, an important indicator of your respiratory function. It is non-invasive device intended for spot-check of adult and pediatric whose age is over 3 at home, hospital and clinics.
Attention: Consult the accompanying documents. Please read this manual carefully before use. Please be sure to keep this manual.
ACT (Artery Check Technology)
ACT processes the SpO2 signal and determines the elasticity of blood vessel based on the derived wave form. It further classifies the vascular status into 6 levels and presents the result in an intuitive graphical interface.
![]() | Artery and blood circulation in excellent condition |
![]() | Artery and blood circulation in good to average condition |
![]() | Artery and blood circulation in average condition |
![]() | Artery and blood circulation in below average condition |
![]() | Artery and blood circulation in poor condition |
![]() | Artery and blood circulation in critical condition |
Note: the classification of artery and blood circulation condition is for reference only, Please consult with your physician for further advice.
Error code for your reference
SENSOR ERROR: ![]() | Sensor cannot be detected, return the device to your local distributor or service centre. |
MEASURE ERROR: ![]() | Signals cannot be detected, turn the device off and measure again. |
AC ERROR: ![]() | Weak signal for artery check, turn the device off and measure again. |
Name/Functions of each part
![Rossmax - SB210 - Parts Parts]()
- OLED display
- Pulse strength
- Pulse search icon
- Artery check icon
- Alarm icon
- Battery indicator
- Power On/Off Button
- SpO2 icon
- Pulse icon
- Battery cover
Installing Batteries
![Rossmax - SB210 - Installing Batteries Installing Batteries]()
- Use thumb to slide battery cover out
- Insert or replace 2 "AAA" sized batteries down with the correct electrical polarity.
You need to replace the batteries when
- LOW BATTERY appears on display
- The function button is pressed and nothing appears on display
![]()
Batteries may leak or explode if used or disposed of improperly. Remove batteries if the device will be stored for long time. Do not use different types or brands of batteries at the same time. Do not mix fully charged and partially charged batteries at the same time.
Attaching the lanyard
![Rossmax - SB210 - Attaching the lanyard Attaching the lanyard]()
- Insert the narrow end of the lanyard through the holder.
- Draw the other end of the lanyard through the loop at the narrow end and tighten.
How to use
![Rossmax - SB210 - How to use - Step 1 How to use - Step 1]()
![Rossmax - SB210 - How to use - Step 2 How to use - Step 2]()
- Open the clip; press the Power On/Off button as
. Bluetooth turns on automatically after power on. - Information of software version appears. Insert one finger(left hand middle finger is recommended), nail side up, into the finger opening of the pulse oximeter as
.
Note: If no finger insert, the device will auto shut off after 30 seconds - The display shows
, pulse oximeter begins its measurement.
Note: Make sure the finger is lying flat. Do not shake and keep body steady during measurement. - Your SpO2 and pulse rate values will appear on the screen after few seconds and artery check result will appear on screen after 30 - 60 seconds as
.
Note: If artery check result cannot be detected, "
" will appear on the screen. - Remove the finger, the screen will show artery check level as
. - Press button shortly to switch the display to the 3 parameters (artery check, SpO2 and pulse rate) as
.
Note:
- The SpO2 sensor and photoelectric receiving tube should be arranged in a way with the subject's arteriole in a position there between.
- Make sure the optical path is free from any optical obstacles like rubberized fabric.
- Excessive ambient light may affect the measuring result. It includes fluorescent lamp, dual ruby light, infrared heater, direct sunlight and etc.
- Strenuous action of the subject or extreme electrosurgical interference may also affect the accuracy.
- The device will reflect the visual and audio signal when the measurement of SpO2 is lower than the default value of 90%.
Data Transfer via Bluetooth®
Pairing the Pulse Oximeter with your Smartphone.
![play.google.com]()
![www.apple.com]()
![]()
- Bluetooth® turns on automatically after device power on.
- Download and install the free APP onto your smartphone.
- To pair this device with your smartphone for data transmission.
Cautionary Notes
- This device is to be operated by trained personnel only.
- The device has intended only for spot-checking, but not medical result evaluation.
- This device is designed to determine the percentage of arterial oxygen saturation of functional hemoglobin. Factors that may degrade pulse oximeter performance or affect the accuracy of the measurement include the following:
- Do not apply the pulse oximeter on the same arm as a blood pressure cuff, arterial catheter or infusion line(s)
- Excessive light, such as sunlight or direct home lighting.
- Not steady at the site of application (e.g. trembling)
- Moisture in the device
- Improperly applied device
- Finger is too large or too small to fit into the device.
- Poor pulse quality
- Venous pulsations
- Anemia or low hemoglobin concentrations.
- Cardiogreen and other intravascular dyes
- Carboxyhemoglobin
- Methemoglobin
- Dysfunctional hemoglobin
- Artificial nails or fingernail polish
- On fingers with anatomical changes, oedemas, scars or burns.
- Using the device for long periods may cause pain for people with circulatory disorders. Reposition the device at least once every 4 hours to allow the patient's skin to breath and to check patient's condition regularly.
- Do not use the device near flammable or explosive gas mixtures.
- Do not use the device during an MRI or CT scan, be used no closer than 30 cm (12 inches) to any part of the SpO2 device, including cables specified by the manufacturer.
- The device may not work when circulation is reduced. Warm or rub the finger, or re-position the device.
- This device is a precision electronic instrument and must be repaired by qualified technical professionals. Field repair of the device is not possible. Do no attempt to open the case or repair the electronics. Opening the case may damage the device and void the warranty.
- Do not overextend the device's spring.
- A functional tester cannot be used to access the accuracy of a pulse oximeter monitor.
- Do not self-diagnose or self-medicate on the basis of the measurements without consulting your doctor. In particular, do not start taking any new medication or change the type and/ or dosage of any existing medication without prior approval.
- Do not look directly inside the housing during the measurement. The red light and the invisible infra-red light in the pulse oximeter are harmful to your eyes.
- This device is not intended for use by people (including children) with restricted physical, sensory or mental skills or a lack of experience and/or a lack of knowledge, unless they are supervised by a person who has responsibility for their safety or they receive instructions from this person on how to use the device. Children should be supervised around the device to ensure they do not play with it.
- Neither of the displays for the pulse wave and pulse bar allows the strength of the pulse or circulation to be evaluated at the measurement site. Rather, they are exclusively used to display the current signal variation at the measurement site and do not enable reliable diagnostics for the pulse.
- The maximum temperature of sensors which the user will touch might reach 43°C when operating in the 40°C environment.
- The safety for user use in high-temperature environment is measuring for 10 minutes, and turn it off for 20 minutes before measure again.
- The oximeter is calibrated in the factory before sale. There is no need to calibrate during its life cycle.
Cleaning
- Please clean the surface of the device before using. Wipe the device with medical alcohol (70% isopropyl alcohol) first, and then let it dry in air or clean it by dry clean fabric. When cleaning the device with water, the water temperature should be lower than 60°C
- Using the medical alcohol to disinfect the product after use, prevent from cross infection for next time use.
- The best storage environment of the device is -25°C to 70°C ambient temperature and not higher than 90% relative humidity.
Note:
- Do not sterilize, autoclave or immerse this device in liquid. Do not pour or spray any liquids onto the device.
- Do not use caustic or abrasive cleaning agents, or any cleaning agent containing ammonium chloride or isopropyl alcohol.
Troubleshooting
Symptoms | Check points | Corrections |
SpO2 or pulse rate cannot displayed | Applied finger improperly. | Place the finger properly and try again. |
SpO2 is too low to detect | Try again; go to consult with your physician if you are sure the device works well. |
SpO2 or pulse rate are not displayed stably | Applied finger improperly. | Place the finger properly and try again. |
Finger is shaking or body is moving. | keep body steady |
No display when button is pressed | Batteries run down | Replace with new batteries |
Batteries not inserted correctly. | Re-insert batteries |
The display disappears suddenly | The device will auto power off when it gets no signal. | Normal |
Low battery | Replace with new batteries |
Note: If the unit does not work, return it to your dealer. Under no circumstance should you disassemble and repair the unit by yourself.
Specification
SpO2 |
Measuring range | 35%~99%, (the resolution is 1%). |
Accuracy | 70%~99%: ±2%, Below 35~69%: unspecified. |
Optical Sensor | The wavelength of red LED is 660 nm and Infrared LED is 905/880 nm with maximum optical output power of 4 mW/sr. |
Pulse |
Measuring range | 30bpm~250bpm (the resolution is 1 bpm) |
Accuracy | ±3bpm |
Power source | AAA × 2 (Alkaline) |
Battery life | Continually for 16 hours with two alkaline batteries |
Operating Condition | Temperature: 5°C~40°C (41°F ~ 104°F), Relative Humidity: 15-95% (non condensing), Atmospheric pressure: 700hPa ~ 1060hPa, Attitude: -1,280 to 12,000 feet (-390m to 3,658m) |
Storage / Transportation Condition | Temperature: -25°C~+70°C (-13°F ~ 158°F), Relative humidity: 15-90%(non condensing), Atmospheric pressure: 700hPa ~ 1060hPa, Attitude: -1,280 to 12,000 feet (-390m to 3,658m) |
Dimensions | 63.5(L) × 34W) × 35(H) mm |
Weight | About 37g (without the batteries) |
Standards | IEC60601-1-2, Class B, IEC60601-1, Type BF, ISO80601-2-61, IEC60601-1-11 |
![]() | Type BF applied parts |
IP Classification | IP22: Protection against harmful ingress of water and particulate matter |