TABLE OF CONTETENTS INTRODUCTION........................2 General Description Indications for Use Contraindications Measurement Principle Safety Information LCD Display Signal Monitor Components List BEFORE YOU START......................9 The Choice of Power Supply Installing and Replacing the Batteries Setting Date, Time and Measurement Unit Setting a user ID Pair a smart device with the monitor MEASUREMENT........................17 Tie the Cuff...
The design provides you with 3 years of reliable service. Readings taken by the L-BM02 are equivalent to those obtained by a trained observer using the cuff and stethoscope auscultation method. This manual contains important safety and care information, and provides step by step instructions for using the product.
INTRODUCTION • Safety Information The symbols below might be in the user manual, labeling or other component. They are the requirement of standard and using. Symbol for “THE OPERATION Symbol for “TYPE BF APPLIED GUIDE MUST BE READ” PARTS” Symbol for “ENVIRONMENT Symbol for “COMPLIES WITH PROTECTION - Electrical waste MDD 93/42/EEC REQUIREMENTS”...
Page 6
INTRODUCTION CAUTION * This device is intended for adult use in homes only. * The device is not suitable for use on neonatal patients, pregnant women,patients with implanted, electronical devices, patients with pre-eclampsia, premature ventricular beats, atrial fibrillation, peripheral, arterial disease and patients undergoing intravascular therapy or arterio-venous shunt or people who received a mastectomy.
Page 7
INTRODUCTION CAUTION * This device is intended for adult use in homes only. * The device is not suitable for use on neonatal patients, pregnant women,patients with implanted, electronical devices, patients with pre-eclampsia, premature ventricular beats, atrial fibrillation, peripheral, arterial disease and patients undergoing intravascular therapy or arterio-venous shunt or people who received a mastectomy.
Page 8
* If you have any problems with this device, such as setting up, maintaining or using, please contact the SERVICE PERSONNEL of Lifebox. Don’t open or repair the device by yourself in the event of malfunctions. The device must only be serviced, repaired and opened by individuals at authorized sales/service centers.
Page 9
INTRODUCTION • Display and Symbols SYMBOL DESCRIPTION EXPLANATION Systolic blood pressure High pressure result Diastolic blood pressure Low pressure result PUL/min Pulse display Pulse in beats per minute Measurement Unit of the blood pressure Measurement Unit of the blood pressure mmHg (1mmHg=0.133kPa) User ID...
BEFORE YOU START • The Choice of Power Supply Battery powered mode: 6V DC 4× AA batteries AC adapter powered mode: 5V AC adapter (Please only use the AC adapter adaptor model). Please unplug the adaptor to depart fromthe using utility power. USB cable CAUTION In order to get the best effect and protect your monitor, please use...
BEFORE YOU START • Setting Date, Time and Measurement Unit It is important to set the clock before using your blood pressure monitor, so that a time stamp can be assigned to each record that is stored in the memory. (The setting range of the year :2020—2099 time format:24H) When the monitor is off, Hold press “SET”button,it will display [YEAR] enter...
Page 13
BEFORE YOU START When you get the right year, press “SET” to set down and turn to next step. Repeat steps 2 and 3 to set the [MONTH] and [DAY]. Repeat steps 2 and 3 to set the [HOUR] and [MINUTE].
Page 14
BEFORE YOU START Repeat steps 2 and 3 to set the [UNIT]. After the unit is set,the LCD will display “dOnE”first,then display all the settings you have done and then it will turn off.
BEFORE YOU START • Setting a user ID 1. When the monitor is off, long press “MEM” button and then the user ID will shows. Press “MEM” button to switch the user ID between user 1, user 2 and user Guest mode. 2.
Page 16
The operating system must be Android 8.0 or more. CAUTION Interference may occur in the vicinity of equipment marked with the following symbol . And L-BM02 may interfering vicinity electrical equipment. Sensitive people, including pregnant women, including pre-eclamptic, patients and those patients with implanted, electrical devices should avoid using this product whenever possible.
Page 17
BEFORE YOU START • Tie the cuff Remove all jewelry, such as watches and bracelets from your left arm. Note: If your doctor has diagnosed you with poor circulation in your left arm, use your right arm. Roll or push up your sleeve to expose the skin.
Page 18
BEFORE YOU START Start the Measurement • Before you start the measurement,Download the Lifebox Health app from APP Store or Google Play,and turn on the Bluetooth. Install the APP, and register an account. Then set your personal information (Gender, Birthday, Height, Weight, Name and so on).
MEASUREMENT 2.After the measurement was finished, the symbol will start blinking, and the data will start transmitting. ( Please connect the app during the transmission ) 3. If the data transmits successful, the symbol symblo will disappear, and then the monitor will turn off.
DATA MANAGEMENT • Recall the Records 1. When the monitor is off, please hold press the “MEM”,it will display the latest record first when the records are less than three groups. When there are three or more than three groups ,it will display the average value of the latest three records first.
DATA MANAGEMENT • Delete the Records If you did not get the correct measurement, you can delete all results for the selected user by following steps below. 1.Hold pressing “MEM” for 3 seconds when the monitor is in the memory recall mode ,the flash display “dEL ALL”...
INFORMATION FOR USER • Tips for Measurement Measurements may be inaccurate if taken in the following circumstances. Within 1 hour Immediate measurement after dinner or drinking after tea, coffee, smoking Within 20 minutes after taking a bath When talking or moving your fingers In a very cold environment When you want to discharge urine...
Page 23
INFORMATION FOR USER • Maintenance In order to get the best performance, please follow the instructions below. Put in a dry place and Avoid touching water, clean it with a dry cloth in case. avoid the sunshine Avoid intense shaking Avoid dusty and unstable and collisions temperature environment...
ABOUT BLOOD PRESSURE • What are systolic pressure and diastolic pressure? When ventricles contract and pump blood out of the heart, the blood pressure reaches its maximum value in the cycle, which is called systolic pressure. When the ventricles relax, the blood pressure reaches its minimum value in the cycle, which is called diastolic pressure.
ABOUT BLOOD PRESSURE Blood Level Optimal Normal High-normal Mild Moderate Severe Pressure (mmHg) <120 120-129 130-139 140-159 160-179 ≥180 <80 80-84 85-89 90-99 100-109 ≥110 CAUTION Only a physician can tell your normal BP range. Please contact a physician if your measuring result falls out of the range.
ABOUT BLOOD PRESSURE • Why does my blood pressure fluctuate throughout the day? 1. Individual blood pressure varies multiple times everyday. It is also affected by the way you tie your cuff and your measurement position, so please take the measurement under the same conditions.
Page 27
INFORMATION FOR USER This section includes a list of error messages and frequently asked questions for problems you may encounter with your bloodpressure monitor. If the products not operating as you think it should, check here before arranging for servicing. PROBLEM SYMPTOM CHECK THIS...
Input: AC 100-240V 50/60Hz 0.2A Max Output: 6V 1000mA • Contact Information For more information about our products, please visit www.lifeboxasia.com/ Imported by: LIFEBOX ASIA.Co.,Ltd Company: LIFEBOX ASIA.Co.,Ltd Address: 26, Soi Sukhumvit 62, Phra Khanong Tai, Phra Khanong, Bangkok Thailand 10260...
COMPLIED STANDARDS LIST • Complied Standards List EN ISO 14971:2012 / ISO 14971:2007 Medical devices - Risk management Application of risk management to medical devices EN ISO 15223-1:2016 / ISO 15223-1:2016 Medical devices. Symbols to be used with medical device labels, labelling and Labeling information to be supplied.
Warning: Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the equipment L-BM02 including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
Page 32
TROUBLESHOOTING Table 2 Guidance and manufacturer’s declaration – electromagnetic Immunity IEC 60601-1-2 Immunity Test Compliance level Test level Electrostatic ±8 kV contact ±8 kV contact discharge (ESD) ±2 kV, ±4 kV, ±8 kV, ±15 kV air ±2 kV, ±4 kV, ±8 kV, ±15 kV air IEC 61000-4-2 ±2 kV for power supply lines ±2 kV for power supply lines...
Page 33
TROUBLESHOOTING Table 3 Guidance and manufacturer’s declaration - electromagnetic Immunity Compliance Radiated RF Test Band Service Modulation Distance Maximum 60601-1-2 level Frequency (MHz) IEC61000-4-3 Power Test Level (V/m) (MHz) (Test (V/m) specifications 380-390 Pulse TETRA modulation ENCLOSURE 18 Hz PORT IMMUNITY to 430-470 GMRS 460,...
Page 50
การวั ด • เริ ่ ม การวั ด Before you start the measurement,Download the Lifebox Health app from APP Store or Google Play,and turn on the Bluetooth. Install the APP, and register an account. Then set your personal information (Gender, Birthday, Height, Weight, Name and so on).
รายการมาตรฐานที ่ ป ฏิ บ ั ต ิ ต าม • Complied Standards List EN ISO 14971:2012 / ISO 14971:2007 Medical devices - Risk management Application of risk management to medical devices EN ISO 15223-1:2016 / ISO 15223-1:2016 Medical devices. Symbols to be used with medical device labels, labelling and Labeling information to be supplied.
Need help?