M S- 9 5 S S e tt i ng R e por t A - 5 . S u r g e P r o t e c t i o n EKO INSTRUMENTS CO., LTD. - Sunshine duration sensor MS-95S - Instruction Manual Ver. 1 Pg. 1...
For warranty terms and conditions, please contact EKO Instruments or your distributor for further details. EKO guarantees that all products delivered to customers have been tested to ensure the instrument meets its published specifications. The warranty included in the conditions of delivery is valid only if the product has been installed and used according to the instructions provided in this operating manual.
2- 3. Abo ut th e Ins t ruc ti o n M an ua l Copy Rights Reserved by EKO INSTRUMENTS CO., LTD. Making copies of the whole or parts of this document without permission from EKO is prohibited.
Page 5
2- 5. CE Dec l a rati o n EKO INSTRUMENTS CO., LTD. - Sunshine duration sensor MS-95S - Instruction Manual Ver. 1 Pg. 4...
➢ When connecting this product to a measuring instrument, connect the shield wire of the output cable to the ground earth for the MS-95S. Noise may be included in the measurement data. ➢ Although this product is tested to meet EMC Directive compliance requirements, it may not fully satisfy...
This product supports communication through the RS485 Modbus RTU. ➢ It is recommended to use the optional EKO converter cable when connecting the MS-95S to a PC. ➢ Depending on the USB-RS485 converter type, an additional termination resistor (120Ω) and/or pull- up/pull-down resistor (680Ω) is required for proper communication.
Page 8
EKO developed the MS-95S as an alternative. With an optical sensor design built around a new fisheye lens that observes the sky in 180°, the MS-95S employs a unique method to distinguish and record direct beam radiation from total sky radiation.
Page 9
In addition, because of seasonal variations in solar altitude, some sensors require data correction throughout the year. In contrast, the MS-95S is equipped with a fisheye lens with a viewing angle of 180°, allowing it to collect solar radiation from the entire sky. As a result, the solar radiation entering the sensor is contained within a single field of view.
Page 10
With multiple smart interfaces, the MS-95S can be used in applications in various fields such as agriculture, industry, construction, medicine, and solar energy. The digital output Modbus RTU / SDI-12 can be used in multi- sensor networks for on-site monitoring of large measurement systems.
[M5] x4 Nuts [M5] x2 (1) Standard length is 10m. For different length of cables [e.g. to meet your application needs] please contact EKO or your local distributor. (2) Details about the setting report, see appendix A-4. Table 4-2. Accessories List...
3. Sun Screen, Body The MS-95S have a sun screen to prevent any excessive body temperature increase generated by direct sun light. Weather resistant metals are used to reduce the effects and degradation caused by heat radiation, rain, and windblown dusts.
Page 13
Durable materials are used for the cable and connector, and pin terminals are attached at the end of the cable for easy connection with a data logger terminal block. * If longer cables, round terminals or fork terminals are required, please contact EKO Instruments or your distributor. [Also see [7. Specification, 7-4. Accessories List] for optional items.] The connector should be protected by the self-fusing tape when the sunshine duration sensor is used in areas with a risk of salt damage.
The sun screen can be removed by loosening the knurling screw and sliding it towards the spirit level direction. *When carrying the MS-95S with a sun screen attached, always hold the instrument from the bottom. Holding only the sun screen may lead to dropping the sensor as it comes off from the sun screen.
Page 15
Fasten the sunshine duration sensor to the base with the 2 bolts [included] and put the sun screen back on the sunshine duration sensor. EKO INSTRUMENTS CO., LTD. - Sunshine duration sensor MS-95S - Instruction Manual Ver. 1 Pg. 14...
2. Wiring Procedure Connect the output cable to the MS-95S: Insert the output cable into the connector port on the back of the sensor unit, and twist clockwise to fasten it. Make sure to tighten the connector all the way.
Page 17
5- 4. Co nn ec ti o ns The MS-95S is based on a digital Smart processing interface providing advanced measurement functions with 4 different industrial signal outputs (Modbus RTU, configurable SDI-12, 4-20mA and 0-1V with external shunt resistor 100Ω).
Page 18
The MS-95S has two analog outputs (4-20mA and 0-1V). The 4-20mA output is default, the 0-1V(0-10mA) output can be enabled through the ‘Hibi’ software which can be downloaded from the EKO website (MS-95S product page). See appendix A-1 to change the current output settings. The current output can be converted into a voltage output when a shunt resistor is used in series.
Page 19
The current output can be converted into a voltage output when a shunt resistor is used in series. The default output is 4-20mA. The 0-10mA (0-1V) output is disabled. The settings can be changed when the MS-95S is connected to a PC through the Hibi software, see appendix A-1 for instructions.
Page 20
2. Digital Output (Modbus RTU / SDI-12) The MS-95S can work in two digital output modes, respectively Modbus RTU and SDI-12. Modbus RTU output is set to default, SDI-12 output can be selected through the Hibi software. PC Connection (Modbus RTU) ➢...
Page 21
RS485 configuration standards as shown in figure 5-7. The master represents the data-logging device and slaves represent devices such as the MS-95S or other serial devices in the same network. Connect the communication wires of the slave to the Modbus RTU communication input of the master.
Page 22
(see appendix A-5). ➢ Using Modbus RTU and SDI-12 in a network at the same time may cause communication problems. EKO INSTRUMENTS CO., LTD. - Sunshine duration sensor MS-95S - Instruction Manual Ver. 1 Pg. 21...
Page 23
Figure 5-9. How to connect the MS-95S (SDI-12) The MS-95S can be configured within a multi sensor SDI-12 communication network. Up to 20 units of S- series sensors can be connected and be assigned with a unique address (*). For a multi sensor network the sensors need to be configured according to the recommended SDI-12 configuration standards as shown in figure 5-10.
Page 24
(*) When 0-1V output is selected, a separate shunt resistor is required. Figure 5-11. Connector pin number of the MS-95S Each number corresponds to the number in Table 5-5. EKO INSTRUMENTS CO., LTD. - Sunshine duration sensor MS-95S - Instruction Manual Ver. 1 Pg. 23...
In this case, Appendix A-2 or A-3 communication format must be supported and programming is required. If only one MS-95S is connected, Hibi software can be installed on a PC for measurement setup and data acquisition. With the optional EKO converter cable, operation is possible simply by connecting the PC and MS-95S via USB.
Page 26
2. Sunshine Duration Sensor Default Settings The MS-95S is based on a digital Smart processing interface providing advanced measurement functions with different industrial signal outputs (Modbus RTU, configurable SDI-12,). Where to find information regarding the MS-95S defaults settings and important parameters: Table 5-7.
6- 1. Main te na nc e By using the MS-95S, accurate results can be obtained if the glass dome and the condition of the instrument are maintained properly. Regular maintenance can extend the lifetime of the sunshine duration sensors. However, environmental conditions can have a deteriorating effect on the materials.
Page 28
Table 6-2. Maintenance Items 2 (Advanced remote check can be done on the MS-95S) Maintenance Frequency How To Effect Item Check the sunshine duration data and When a large difference occurs, Data Validity Weekly compare to previous days or adjacent operating problems or installation the sunshine duration sensors.
6- 3. Tro ubl es h ooti ng Read the following in case of any sensor trouble. If any questions should remain, please contact EKO for further technical support. Table 6-3. In-field troubleshooting Failure Action Make sure that the sensor is properly connected, and type of power supply and voltage There is no output values are appropriate.
Temperature/ humidity alarm/ tilt angle WMO compliant Sensor setting can be changed by connecting the sensor to a PC. Use the EKO Convertor Cable (option) and download the free configuration software “Hibi” from the EKO website. When the instrument is used in the ambient temperature exceeding the accuracy assurance temperature range, the measurement error may increase.
Page 31
Value at 20mA continuous output 300mW 640mW Shunt resistance 100Ω 0-1V Output (Approx. 30mA) (Approx. 27mA) Value at 10mA continuous output (*) Using external shunt resistor 100Ω EKO INSTRUMENTS CO., LTD. - Sunshine duration sensor MS-95S - Instruction Manual Ver. 1 Pg. 30...
Page 32
C. Levelling Screw Height 16 mm ⌀ 96 mm D. Width [including sun screen/Cover] E. Overall Height [approx.] 101mm Figure 7-1. External Dimensions (MS-95S) EKO INSTRUMENTS CO., LTD. - Sunshine duration sensor MS-95S - Instruction Manual Ver. 1 Pg. 31...
EKO Converter Cable 95S and allows to connect to PC via USB terminal. Cable Length: 5m The standard cable length for the MS-95S is 10m. EKO INSTRUMENTS CO., LTD. - Sunshine duration sensor MS-95S - Instruction Manual Ver. 1 Pg. 32...
* Connect the power supply GND and RS485GND Figure A1-1. Hardware preparation If you use the optional ‘EKO Converter cable’, you can install it simply by inserting the cable connector into the main unit. Power for the MS-95S can be supplied via USB. (See Figure A3-2).
Page 35
After the software starts, the ‘Welcome’ window appears. Figure A1-3. Welcome window Automatic connection When using the ‘Auto-connect’ function to connect MS-95S with the ‘Hibi’ software, select the COM port that the RS485/USB converter cable is connected to. Figure A1-4. Com port select Click ‘Select’...
Page 36
Check that the power supply of the sensor unit is ‘OFF’. Then, click the ‘CONNECT’ button. If the COM port is not recognized, press the refresh button and try again. Power on the MS-95S within 10 seconds once the ‘Connecting your sensor’ message appears. Figure A1-6. Connecting screen When the connection is completed, the window automatically switches to the ‘Dashboard’...
Page 37
Make sure the connection settings are correct. To perform the ‘Auto-connection’ again, turn off the power of the MS-95S, select ‘COM port’, and then click ‘CONNECT’, then Power ON to the MS-95S. For manual connections, click ‘view settings’ at the bottom of ‘CONNECTION FAILED’ to go to the ‘Settings’...
Page 38
- ‘Save’ button: Save the settings. - ‘Read’ button: Read out the settings. - ‘Output’ button: Output the settings to CSV file Figure A1-12. Apply changes screen EKO INSTRUMENTS CO., LTD. - Sunshine duration sensor MS-95S - Instruction Manual Ver. 1 Pg. 37...
Page 39
100 when ‘00’) SDI-12 address 0 to 9 / A to Z / a to z Connection status Connected / Disconnect Figure A1-14. Communication settings screen EKO INSTRUMENTS CO., LTD. - Sunshine duration sensor MS-95S - Instruction Manual Ver. 1 Pg. 38...
Page 40
Firmware version Firmware version of the MS-95S (cannot be changed) Hardware version Hardware version of the MS-95S (cannot be changed) Figure A1-15. Sensor information settings EKO INSTRUMENTS CO., LTD. - Sunshine duration sensor MS-95S - Instruction Manual Ver. 1 Pg. 39...
Page 41
4mA±0.4mA:Sunshine No Signal 4-20mA 20mA±1.6mA:Sunshine Yes Signal output 0-1V 0V±0.1V:Sunshine No Signal (100Ω shunt resistor 1V±0.1V:Sunshine Yes Signal required) Figure A1-16. Analog output settings screen EKO INSTRUMENTS CO., LTD. - Sunshine duration sensor MS-95S - Instruction Manual Ver. 1 Pg. 40...
Page 42
Clicking ‘ZERO ADJUSTMENT’ allows you to adjust Tilt (X axis) and Roll (Y axis) of Tilting items using the present tilt angle as 0°. Figure A1-18. Sensor position setting EKO INSTRUMENTS CO., LTD. - Sunshine duration sensor MS-95S - Instruction Manual Ver. 1 Pg. 41...
Page 43
<Save file to> Enter the name of the setting file storage folder. Figure A1-19. Data files setting 6) Measurement (Dashboard) - Measurement data of irradiance, temperature, humidity, and tilt angle are read from MS-95S every 1 second and displayed. Figure A1-20. Dashboard screen <Sunshine Duration>...
Page 44
(Start time < End time) <LOG DATA buttons> Click to start logging measurement data. When ‘STOP’ is clicked while measurement data is being logged, logging stops. EKO INSTRUMENTS CO., LTD. - Sunshine duration sensor MS-95S - Instruction Manual Ver. 1 Pg. 43...
Page 45
Contact (Contact) When you click the ‘Contact’ button, the contact page on the [EKO Instruments] website appears in the browser. Figure A1-22. EKO Instruments website contact page Sensor internal humidity abnormality alert Hibi has a function to monitor the relative internal humidity of the sunshine duration sensor at all times. When the internal relative humidity exceeds the standard value due to external failure or deterioration of drying agent over time, Hibi judges it to be abnormal and displays an alert.
Page 47
Table A2-4. Data format to use. Format Abstract Unsigned 16bit Integer Signed 16bit Integer Unsigned 32bit Integer Signed 32bit Integer IEEE754 32bit floating point format ASCII characters string EKO INSTRUMENTS CO., LTD. - Sunshine duration sensor MS-95S - Instruction Manual Ver. 1 Pg. 46...
Page 48
· The contents of registers No. 0 to 49 are switched according to the register type setting (No.103). · Registers No. 0 to 99 are read-only. · Registers after No.100 can be read or written to. EKO INSTRUMENTS CO., LTD. - Sunshine duration sensor MS-95S - Instruction Manual Ver. 1 Pg. 47...
Page 49
X-axis component of the tilt angle X-axis tilt angle Unit: ° Y-axis component of the tilt angle Y-axis tilt angle Unit: ° 12 to 49 are reserve EKO INSTRUMENTS CO., LTD. - Sunshine duration sensor MS-95S - Instruction Manual Ver. 1 Pg. 48...
Page 50
A2-9. Register details from No.50 to 99 · A fixed value of zero (U16) has been written to an address that is not used. · The ‘EKO’ character is written as the company name registration in register No. 96 and 97. Table A2-13. Register No.50-95 50 to 95 are reserve Table A2-14.
Page 51
Sensor name (10,11) Sensor name (12,13) Sensor name (14,15) 174 to 189 are reserve Cal. date Calibration date in YYYYMMDD format 192 to 199 are reserve EKO INSTRUMENTS CO., LTD. - Sunshine duration sensor MS-95S - Instruction Manual Ver. 1 Pg. 50...
Page 52
· Do not write to any Coil other than Reboot/Save. · The readout is always zero. A2-12. Register details Table A2-16. Coil register Address Parameter Type Reboot Save EKO INSTRUMENTS CO., LTD. - Sunshine duration sensor MS-95S - Instruction Manual Ver. 1 Pg. 51...
Page 53
14: SDI-12 version (represents version 1.4) EKOINST: Company name (8 characters) MS-95S: Sensor model name (6 characters) V30: Sensor version (3 characters) 20000001: Serial number (8 characters) EKO INSTRUMENTS CO., LTD. - Sunshine duration sensor MS-95S - Instruction Manual Ver. 1 Pg. 52...
Page 54
Read the calibration date of this device. YYYY: AD, MM: month, DD: day. <CR><LF> Variable Identify Measurement Commands and Identify Measurement Parameter METADATA depending on Commands are supported. request data EKO INSTRUMENTS CO., LTD. - Sunshine duration sensor MS-95S - Instruction Manual Ver. 1 Pg. 53...
However, the type, rating, and connection method of the SPD to be used vary depending on the manufacturer, so please read the instruction manual of each manufacturer carefully. EKO INSTRUMENTS CO., LTD. - Sunshine duration sensor MS-95S - Instruction Manual Ver. 1 Pg. 55...