DEI PVX-4000-2kV Operation Manual

DEI PVX-4000-2kV Operation Manual

Pulsed voltage source

Advertisement

Quick Links

Scientific Instruments GmbH
SI
... the better technology
Roemerstr. 67 | 82205 Gilching/Germany | Phone: +49 8105 77940 | E-mail: info@si-gmbh.de | Web: www.si-gmbh.de
PVX-4000-2kV
Pulsed Voltage Source
Operation Manual
Directed Energy, Inc.
1609 Oakridge Dr., Suite 100, Fort Collins, CO 80525
(970) 493-1901
Manual Document 7650-0022 Rev. A05
© Copyright 2017 IXYS Colorado. All rights reserved.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PVX-4000-2kV and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for DEI PVX-4000-2kV

  • Page 1 Scientific Instruments GmbH ... the better technology Roemerstr. 67 | 82205 Gilching/Germany | Phone: +49 8105 77940 | E-mail: info@si-gmbh.de | Web: www.si-gmbh.de PVX-4000-2kV Pulsed Voltage Source Operation Manual Directed Energy, Inc. 1609 Oakridge Dr., Suite 100, Fort Collins, CO 80525 (970) 493-1901 Manual Document 7650-0022 Rev.
  • Page 3: Table Of Contents

    Contents Contents  ................................ 3  Safety ................................ 5  Introduction .............................. 6    Description ............................ 6    Front Panel Features  ......................... 7    Rear Panel Features .......................... 9    Accessories Included  ........................ 10  Operating Considerations ........................... 11    Output Cabling .......................... 11    Pulse Voltages .......................... 11    Trigger Method .......................... 11  Local Operation  ............................ 12    Local Mode  ............................ 12    Setup ............................... 12    Power Up  ............................ 12    Set the Output Voltage ........................ 13   ...
  • Page 4  Fault Codes  ............................ 19  Warranty and Service  .......................... 20   Warranty ............................ 20   Factory Service and Support ...................... 21  Declaration of Conformity .......................... 22 ...
  • Page 5: Safety

     Always refer to the Safe Operating Area (SOA) of the PVX-4000-2kV system, located in the datasheet.  Do not open the cover of the PVX-4000-2kV. There are no user-serviceable parts inside. Opening the cover exposes you to shock and voids the factory warranty.
  • Page 6: Introduction

    50 Ω. The sync output pulse is synchronized to the leading edge of the output voltage pulse and is active with the external trigger. The PVX-4000-2kV includes a pulse generator and high voltage DC power supplies, providing you with a standalone test system that can be programmed to perform device testing using the front panel LCD controls or a computer interface (RS-232 and USB).
  • Page 7: Front Panel Features

     Front Panel Features     Power LED The Power LED is lit when there is support power applied to the PVX-4000-2kV. Fault LED The Fault LED is lit when there is a fault/error sensed within the system. Enable LED...
  • Page 8 Enable Input The Enable BNC input accepts either an external dry-contact closure or the factory- supplied shorting BNC plug (PCA-9410).  Closed contacts or use of the shorting plug allows the output to be enabled. Less than 1000 Ω will allow the system to be enabled. ...
  • Page 9: Rear Panel Features

    The voltage output connector, a Winchester Kings 10 kV, 1065-2 plug, is the output connection for system-generated voltage pulses. DO NOT TOUCH any part of this output or cable while the PVX-4000-2kV is powered up and enabled. Please review the Safety section.
  • Page 10: Accessories Included

    Pulsed Output Cable The output cable (PCA-9400) is used to connect the PVX-4000-2kV to the load and is two meters in length. The factory supplied output cable has 1.57 pF / 1.0 cm of cable. This needs to be calculated into the total load value.
  • Page 11: Operating Considerations

    1.57 pF / 1.0 cm of cable. This needs to be calculated into the total load value.  Pulse Voltages The PVX-4000-2kV pulser is rated at a maximum pulse output voltage of ±2000 V DC. Proper precautions should be taken by the user to ensure that the maximum voltage is not exceeded.
  • Page 12: Local Operation

     Setup 1. Press the Enable button to disable the voltage output of the system. 2. The PVX-4000-2kV should be OFF for at least ten minutes when connecting or disconnecting the voltage output connector. 3. Connect the output to the load. Verify that the load polarity is correct. If necessary, review the output cable for polarity specifications.
  • Page 13: Set The Output Voltage

    1. Power up the system with the power switch on the rear panel of the PVX-4000- 2kV. The Power LED lights and the Fault LED blinks. The instrument boots up in about 3 seconds. 2. If there are no errors or faults, the Fault LED turns off. If there are errors or faults, the Fault LED lights without blinking after the system has booted.
  • Page 14: Set The Pulse Width

    1. Rotate the encoder knob to the Frequency screen. 2. Press the encoder knob to change the value of the frequency. 3. Rotate the encoder knob to change the value between 40 Hz to 30,000 Hz. 4. Press the encoder knob to change the multiplier value of the frequency. 5.
  • Page 15: Power Down

    2. Enable the PVX-4000-2kV unit by pressing the enable button. The green enable LED turns ON. Typically, the user must press the enable button for up to 2 seconds. 3. The PVX-4000-2kV is now sending high voltage pulses through the output connection.
  • Page 16: Remote Operation

    Home screen of the front panel. Any communication on the USB port will disable the RS-232 port. The RS-232 port will be disabled until the PVX-4000-2kV is power cycled, at which time the RS-232 is active by default.
  • Page 17: Command List

    In the examples, the command or query is in black; the response is in blue italics. *IDN? Abbreviation for identity; returns the manufacturer’s name, unit model number, serial number, firmware revision, and manufacture date. Example: *IDN? DEI Scientific, PVX-4000-2kV, 170151, D01P03, 00 OUTPUT? Returns the state of the output (Enabled or Disabled). Example: OUTPUT? Enabled OUTPUT:ENABLE Enables the output.
  • Page 18 PULSE:WIDTH? Returns the pulse width of the output pulse in nano-seconds. Example: PULSE:WIDTH? 10,000 ns PULSE:WIDTH nn Sets the pulse width of the output pulse in nano-seconds. Example: PULSE:WIDTH 10000 10,000 ns PULSE:FREQ? Returns the trigger frequency in hertz. Example: PULSE:FREQ? 6000 Hz PULSE:FREQ n...
  • Page 19: Fault Codes

    SAVE:n Saves the current configuration to non-volatile memory. The value of n must be between 1 and 5. Example: SAVE:3 SAVE:3 RECALL:n Recalls saved configurations. The value of n must be between 1 and 5. Example: RECALL:2 RECALL:2 FAULT? Returns the fault codes. If no fault the response will be 0. Example: FAULT? FAULT:CLEAR Clears the fault codes.
  • Page 20: Warranty And Service

    DEI’s sole obligation, and buyer’s sole remedies, under this agreement shall be limited to a refund of the purchase price, or at DEI’s sole discretion, to the repair or replacement of products in kind that prove, to DEI’s satisfaction, to be defective, when returned to the DEI factory, transportation prepaid by the buyer, within the warranty period.
  • Page 21: Factory Service And Support

     Factory Service and Support For more information about your instrument or for an operation problem, please contact the factory: Directed Energy, Inc. 1609 Oakridge Dr., Suite 100 Fort Collins, Colorado 80525 (970) 493-1901, ext. 101 sales@ixyscolorado.com techsupport@ixyscolorado.com http://ixyscolorado.com/ SI Scientific Instruments GmbH +49 8105 77940 www.si-gmbh.de Roemerstr.
  • Page 22: Declaration Of Conformity

    Declaration of Conformity Page 22 of 24  For more information contact IXYS Colorado: 970.493.1901 or sales@ixyscolorado.com Document #7650‐0022 Rev A05. © Copyright 2017 IXYS Colorado. All rights reserved.