Troubleshooting
9
Problem:
Despite a number of priming
procedures a drop of
teriparatide has not appeared
on the end of the needle.
Problem:
After the injection the green
dot did not appear in the the
window displaying the end
of injection
Possible cause:
Incorrectly installed
teriparatide cartridge.
Clogged pen needle.
Possible
cause:
There was not enough
teriparatide in the cartridge
to inject the full set dose.
HOW TO
ELIMINATE:
Check if the cartridge was properly
installed. There should be no space
between the plunger and the cartridge.
Replace the needle with a new one.
HOW TO ELIMINATE:
Note the number of units remaining, as it
corresponds
to
the
number
teriparatide which you must inject to complete
the injection. Attach the outer needle shield and
remove the needle. Replace the old cartridge with
a new one, accordingly to Section 7. Prepare
the pen injector to use accordingly to the
actions described in steps 7 to 10. Perform
the
priming
dose
injection. Complete
interrupted dose delivery by turning the dose
selector until the figure, which you have seen in
the selector window, appears. Follow steps
through 15.
mcg
of
of
the
14