AUTOMATIC SOMATROPIN PEN INJECTOR
®
HHT
PEN
®
®
®
®
USER'S MANUAL
®
20mg
HHT
PEN 20mg

Advertisement

Table of Contents
loading

Summary of Contents for Nemera HHT PEN

  • Page 1 0 1 9 7 AUTOMATIC SOMATROPIN PEN INJECTOR ® ® ® ® ® USER'S MANUAL ® 20mg PEN 20mg...
  • Page 2: Table Of Contents

    Table of contents before the first use construction of pen injector, pen needle and cartridge installation of cartridge and pen needle ® priming your HHT PEN 20mg injecting a dose removing a needle replacing a cartridge replacing a needle storage and cleaning information from the manufacturer...
  • Page 3: Before The First Use

    before the first use ® PEN 20mg is a reusable, intuitive automatic somatropin injector. It provides you or your child with a precise, safe and effortless way to administer somatropin doses recommended by the doctor, in the range of 0.07 mg (0,01mL) to 4.00 mg (0,6mL), with incremental dose of 0.07 mg (0,01mL).
  • Page 4 before the first use ® PEN 20mg can be used in the home environment by people from different cultural backgrounds, people of different ethnicity, with all levels of education. Use HHT ® PEN 20mg only with the somatropin recommended by your doctor. ®...
  • Page 5 ® construction of the pen injector HHT PEN 20mg window displaying pen body the end of the injection release button pen cap window of cartridge holder plunger selector window cartridge holder rubber cartridge tip dose selector inner needle shield needle somatropin pen needle cartridge...
  • Page 6: Installation Of Cartridge And Pen Needle

    installation of cartridge and pen needle Wash your hands and make sure that you have the following items: Step 1 Pull off ® PEN 20mg the cap somatropin cartridge new pen needle alcohol swab ® New somatropin cartridge or HHT PEN 20mg with cartridge attached should be taken out of the refrigerator 30...
  • Page 7 instalation of cartridge and pen needle Step 3 Press the plunger into the pen body. Step 4 Check the somatropin cartridge. Make sure that it is not cracked or broken, the somatropin solution is clear and transparent and check the medicine expiration date. Do not use the somatropin after the medicine expiration date.
  • Page 8 instalation of cartridge and pen needle Step 5 Insert somatropin cartridge into the cartridge holder, rubber tip to the front. Step 6 Attach the cartridge holder to the pen body by turning it according to direction shown on the drawing, until first resistance is met.
  • Page 9 instalation of cartridge and pen needle Step 7 Take a new sterile needle and remove the paper tab from the outer needle shield. For each injection use a new needle. Do not use needles if the paper cover if damaged. Reuse of the same needle or needle with a damaged unit pack (paper cover) is not in line with epidemiological recommendations.
  • Page 10 instalation of cartridge and pen needle Step 9 Remove the outer needle shield. Retain it for removal of the needle after the completion of the injection process. Step 10 Remove the inner needle shield as recommended by the manufacturer. To minimise the risk of unintended needle penetration and transmission of infectious diseases, never replace the inner needle cap once removed.
  • Page 11: Priming Your Hht ® Pen 20Mg

    ® priming your HHT PEN 20mg Step 11 Using the dose selector, choose 0.13 mg of somatropin. Turn the dose selector clockwise until the figure 0.13 appears in the selector window. Step 12 Take the pen with the needle pointing upwards. Using your thumb, slide the release button in the direction of the needle.
  • Page 12 injecting a dose Step Turn dose selector to your desired dose. Example: 0.13 mg shown in the drawing. If you dial too many units, you can correct the dose by turning the dose selector backwards. Step 14 Select the injection site (abdomen, thighs, buttocks) and wipe your skin with an alcohol swab.
  • Page 13: Injecting A Dose

    injecting a dose Step 15 Hold the release button until a green dot appears in the window displaying the end of the injection. If by accident you inject yourself a dose higher than If the green dot does not appear, it means that there prescribed, immediately contact your doctor.
  • Page 14: Removing A Needle

    removing a needle Step 17 Put the outer needle shield on the needle. Rotate the outer needle shield counterclockwise. Remove the used needle. Dispose of it in accordance with information provided by the manufacturer of the needle. Put the cap on the pen injector.
  • Page 15 Troubleshooting Troubleshooting Przyczyna: ROZWIĄZANIE: Problem: Problem: Possible cause: HOW TO ELIMINATE: Use a new needle with each injection. The crystallisation of The crystallisation of The needle is not replaced Używaj zawsze nowej igły Brak zmiany igły po somatropin at the somatropin at the with each injection.
  • Page 16 Troubleshooting Problem: HOW TO ELIMINATE: Possible cause: Despite a number of priming Check if the cartridge was properly Incorrectly installed procedures a drop of installed. There should be no space somatropin cartridge somatropin has not appeared between the plunger and the cartridge. on the end of the needle.
  • Page 17: Storage And Cleaning

    storage and cleaning Do not im- Always keep your HHT ® PEN 20mg and needles out of the reach of merse other people, especially children. HHT®PEN 20mg should be protected against major physical forces. ® 20mg ® PEN 20mg should always be stored and carried with the top cap on, without into water.
  • Page 18: Information From The Manufacturer

    In the events of serious incidents, this should be reported to the manufacturer and to the competent authority. Particular attention should be paid to the information preceded by this sign which will ensure safe and comfortable use of the device. Distributor/Importer: Manufacturer: Nemera Szczecin sp. z o.o. 3 Tytanowa Street 70-856 Szczecin, Poland plszc_info@nemera.net www.nemera.net...
  • Page 19 legend Indicates the medical device manufacturer (yellow color) Indicates that caution is necessary when operating the device or control close to where the symbol is placed or that the current situation needs operator awareness or operator action in order to avoid undesirable consequences Indicates the need for the user to consult the instructions for use Patient information website...
  • Page 20 notes...
  • Page 21 disposal of pen injector Remember to dispose Your pen injector 3 years after first use. Below note the date of first and last use of ® PEN 20mg. use on: Throw away after:...
  • Page 22 ® ead the instructions carefully before using HHT PEN 20mg Failure to follow of these instructions may have serious consequences to your health. Using the device in a different way from the process described in this manual is not recommended. AUTOMATIC SOMATROPIN PEN INJECTOR ®...