Eilersen 5024G Options Manual

Remote display
Hide thumbs Also See for 5024G:

Advertisement

Quick Links

5024G
5024G LCD weighing terminal
Remote display: 5024 remote display using remote display option
Software:
StdLim.140630.6v3r
Doc. no.:
StdLim-140630-6v3-OG-RemDisp-eng
Date:
2022-11-28
Rev.:
6v3
Options Guide
Remote Display
Contact:
Eilersen Electric A/S
Kokkedal Industripark 4
DK-2980 Kokkedal
Denmark
www.eilersen.com
info@eilersen.com
Tel: +45 49 180 100
Fax: +45 49 180 200

Advertisement

Table of Contents
loading

Summary of Contents for Eilersen 5024G

  • Page 1 5024G Options Guide Remote Display 5024G LCD weighing terminal Remote display: 5024 remote display using remote display option Contact: Software: StdLim.140630.6v3r Eilersen Electric A/S Doc. no.: StdLim-140630-6v3-OG-RemDisp-eng Kokkedal Industripark 4 Date: 2022-11-28 DK-2980 Kokkedal Rev.: Denmark www.eilersen.com info@eilersen.com Tel: +45 49 180 100...
  • Page 2: Table Of Contents

    Introduction ..........................4 Installation of 5024G remote display ..................4 How to ............................4 – Enable and configure remote display on 5024G ..........4 Enable or change Display Type ................. 5 Set/Read Display Address ................. 5 Set/Read Endian ....................5 Read Status .......................
  • Page 3 Revision History ........................15 Contact ............................15 WWW.EILERSEN.COM...
  • Page 4: Introduction

    Eilersen Electric when used for connecting a 5024G remote display. With the software version stated on the front page installed in the 5024G main terminal, and with the Remote Display option enabled, as well as a on RS485 externally connected 5024G remote display from Eilersen Electric, the system can on the remote display show weight indications (gross/net weight) and possibly perform zeroing of these from its keyboard.
  • Page 5: Enable Or Change Display Type

    REMOTE DISPLAY menu select the DISPLAY TYPE menu item to get the selection list with possible dis- play types. Possible display types are 5024 display. If no remote display is enabled the 5024G display will show “-------” / ”None”.
  • Page 6: Monitor External Display Data On 5024G

    – Monitor external display data on 5024G Below the REMOTE DISPLAY DATA screen is shown when a 5024G remote display has been enabled and chosen on the 5024G main terminal. This screen will show the data sent to the external device (OUT) and the data received from the external device (IN).
  • Page 7: Operating 5024 Remote Display

    • If an error is present the weight reading will indicate error, like UL, OL, - 0080 – etc. • If an error in communication with 5024G main terminal occurs the weight reading is replaced by “-------”. Status indications: •...
  • Page 8: Operate The Menu System

    Parameters cannot be changed and actions not performed when the system is powered on. Before such operations are possible the parameters must be unlocked: • Press to invoke the menu system. • Press to highlight the SERVICE MODE menu item. • Press to select the SETUP screen. WWW.EILERSEN.COM...
  • Page 9: Check The Software Version

    The weighing terminal is ready and enters the NORMAL screen. Select the SYSTEM INFO screen to read the software version (software name, date and revi- sion): The SETUP screen also shows the software version (software name, date and revision): WWW.EILERSEN.COM...
  • Page 10: Trouble Shooting

    5024G Check the 5024G remote display is powered correctly. remote display. Re- Check parameters in REMOTE DISPLAY screen on the 5024G main terminal mote display shows “-- are configured correctly: -----“. - Check ”Device Type” parameter matches 5024G remote display (5024).
  • Page 11: Appendices

    Appendices Appendix A – 5024G Main Display: Screens overview The 5024G main terminal has the following screens, which are selected using the menu sys- tem. With the Remote Display option enabled an REMOTE DISPLAY screen and a DATA screen is added: NORMAL ─┬─...
  • Page 12: Appendix C - 5024G Main Display: Electrical Connection

    RS485 connection. 5024G Main Display: RS485 communication connector The 3 pin RS485 serial communication connector (J4) on the 5024G main terminal can be used for RS485 communication with external equipment. This connector has the following connections:...
  • Page 13: Appendix D - 5024G Remote Display: Electrical Connection

    (J4) is used; thus no connection of load cells, digital I/O signals or analog output. 5024G Remote Display: Power Connection The 5024G remote display is powered by +24VDC which is connected to either J1 or J6 (J1 and J6 are internally short-circuited).
  • Page 14: 5024G Remote Display: Rs485 Communication

    5024G Remote Display: RS485 communication The 3 pin RS485 serial communication connector (J4) on the 5024G remote display can be used for RS485 communication with the 5024G main terminal. This connector has the follow- ing pin-out: J4 pin Function Connection...
  • Page 15 Date Author Rev. Update 2022-11-28 Initial document created and adapted. (based on StdLim-201216-1v0-OG-RemDisp-eng) Contact With further questions or improvement suggestions please contact us: Eilersen Electric A/S Kokkedal Industripark 4 DK-2980 Kokkedal Denmark www.eilersen.com info@eilersen.com Tel: +45 49 180 100 Fax: +45 49 180 200 WWW.EILERSEN.COM...