Page 1
22/06/2022 YAV90CIN Note the guidelines in this document are specific to the YAV90CIN . The other components in the system might not meet the same safety ratings. Refer to the documentation of each component in the system to determine...
ELECTROMAGNETIC COMPATIBILITY GUIDELINES The product YAV90CIN to which this manual relates meet the provisions of the EU Directives 2014/30/CE which declares the compliance of electromagnetic compatibility (EMC) and directive 2014/35/CE of low voltages electronic devices.
DESCRIPTION The YAV90CIN module is designed to cover the need to control the available power supplies and loads in a test system. All relay contacts have a 10 mOhm shunt resistor allowing current measurement without circuit interruption; shunt voltage must be measured with an external device.
PINOUT (FRONT) Module function Receiver Pos. # I.T.A. Module P/N 510 108 126 YAV90CIN: 12 Power relays I.T.A. contact P/N 610 110 108 I.T.A. Patchcord 720 102 101 Description Description Description CAN L +24VDC CAN H Out K12 Shunt Current Sense K12 (-)
CONNECTOR PINOUT (REAR) X1 Connector (Power and CAN) Pin Number Signal Description Ground connection +24V Board power supply CAN L CAN-Bus Low signal CAN H CAN-Bus High signal X1 pinout description Plug type: • Manufacturer: Weidmuller • Model: BLZF 3.50/04/180...
LED BEHAVIOUR LED Name LED Behaviour Definition of Behaviour STATUS Board is not powered on, power supply is not reaching the microcontroller, or firmware of microcontroller is not present or is not running due to an error. Solid green Board is powered on, and firmware is running properly. LED behaviour SWITCH CONFIGURATION Switch...
10. HARDWARE SPECIFICATIONS Operation parameters Property Min. Typ. Max. Units • Control Voltage • Operating current @24VDC and 25ºC • kbps CAN Bus operation speed • degC Ambient operating temperature Operation specifications Switching specifications Property Min. Typ. Max. Units • Rated current •...
11. OPERATION Through the PHI6 Explorer soft panel the user can interface with the YAV90CIN board manually. This tool is a very useful for table-top debugging of test-systems or system maintenance on-site. Activate relays By clicking a relay in the panel, the relay closes and its status in the panel is updated.
Page 11
Config tags To operate with Phi6 driver each relay must be given a tag name. By clicking this button, the panel will change its background color from grey to yellow, and tag name can be modified.
To start using the product just remove vacuum bag carefully to avoid damage. To install the YAV90CIN simply insert the module on a VPC receiver, vertical or horizontally, according to the Receiver arrangement. Once it is completely inserted, fix YAV module with two 4-40 UNC screws on each side of connector.