Swan Analytical Instruments AMI Option USB Interface V1 Installation Instruction page 8

Usb interface

Advertisement

12. Press <Enter> again.
The Firmware update starts.
This method updates the firmware of the AMI instrument with all files with the exten-
sion <.a37> or <.hex> that are located in the root directory of the USB stick.
After the update has been finished:
13. Navigate to <Installation>/<Miscellaneous>/<Set defaults>.
14. Press [Enter] and select <completely> .
15. Press [Enter], [Exit] and confirm with <Yes>.
The AMI transmitter restarts and all settings are set to the default values.
16. Enter the customer-specific settings according to the notes taken.
USB Interface installation
108498 - 05/16