Scpi Command Summary - Boonton SGX1000 Series Programming Reference Manual

Rf signal generator
Hide thumbs Also See for SGX1000 Series:
Table of Contents

Advertisement

SYSTem:COMMunicate:LAN:ADDRess
Description:
Set or return the IP address for the Ethernet port.
Syntax:
SYSTem:COMMunicate:LAN:ADDRess <character data>
Argument:
<character data> = instrument IP address in nnn.nnn.nnn.nnn (dot decimal) format
Valid Modes:
Any for queries. DHCP/AutoIP must be disabled (OFF) to set the instrument IP address.
SYSTem:COMMunicate:LAN:SMASk
Description:
Set or return the subnet mask for the Ethernet port.
Syntax:
SYSTem:COMMunicate:LAN:SMASk <character data>
Argument:
<character data> = subnet mask in nnn.nnn.nnn.nnn (dot decimal) format
Valid Modes:
Any for queries. DHCP/AutoIP must be disabled (OFF) to set the subnet mask.
SYSTem:COMMunicate:LAN:DHCP
Description:
Set or return the state of DHCP/AutoIP system for the Ethernet port.
If DHCP/AutoIP is enabled (1 | ON), the instrument will attempt to obtain its IP Address,
Subnet Mask, and Default Gateway from a DHCP (dynamic host configuration protocol)
server on the network. If no DHCP server is found, the instrument will select its own IP
Address, Subnet Mask, and Default Gateway values using the "AutoIP" protocol.
If DHCP/AutoIP is disabled (0 | OFF), the instrument will use the IP Address, Subnet
Mask, and Default Gateway values that have been entered by the user.
Syntax:
SYSTem:COMMunicate:LAN:DHCP[:STATe] <Boolean>
Argument:
<Boolean> = {0, 1, OFF, ON}
Valid Modes:
Any
1.5.6

SCPI Command Summary

Table 1-1 SCPI COMMAND SUMMARY
*RST
*IDN?
DISPlay:BACKlight:BRIGhtness
[OUTput:]STATe
[SOURce:]FREQuency
[SOURce:]FREQuency:MAXimum?
[SOURce:]FREQuency:MINimum?
[SOURce:]POWer
[SOURce:]POWer:MAXimum?
[SOURce:]PHASe:MINimum?
[SOURce:]ROSCillator:EXTernal:FREQuency
[SOURce:]ROSCillator:SOURce]
[SOURce:]MODulation:MODe
[SOURce:]FREQuency:STARt
[SOURce:]FREQuency:STOP
[SOURce:]SWEep:TRIGger:SOURce
10
Recall factory default settings.
Identification Query
Set/return backlight brightness.
Set/return RF Output state ON/OFF.
Set/return frequency in Hz.
Query maximum frequency in Hz
Query minimum frequency in Hz
Set/return power in dBm.
Query maximum power in dBm.
Query minimum power in dBm.
Set/return external reference frequency in Hz.
Set/return reference oscillator source.
Set/return operating mode.
Set/return the sweep start frequency in Hz.
Set/return the sweep stop frequency in Hz.
Set/return the sweep trigger.