Pinion Products Corporation Inc. ____________________________ “ESD and Ionization Instruments for Industry ” 32086 Share 28 Road INSTRUCTION MANUAL – COMBINATION RESISTIVITY METER, MODEL SRM/RTG IMPORTANT!! READ THIS MANUAL BEFORE OPERATING THIS UNIT. Page 1 of 5 SRM/RTG Manual 14 December 2007...
Page 2
For this reason, great pressure is not needed to perform a measurement. Each probe contact of the SRM/RTG applies a pressure of 40 lbs/in^2, due to the 2 pound weight of the SRM/RTG, which is sufficient. The three-point probe design enables the unit to be balanced on irregular and curved surfaces.
Page 3
Turn off the SRM/RTG when it is not being used, and charge ONLY with the supplied charger when the Low Battery indicator is illuminated, or at the end of the day’s use. Charge for a minimum of 12 hours. Do NOT leave the charger connected continuously.
Page 4
The voltage should equal 30V +/- 2 V at each probe. Connect a programmable current source to the center probe contact and ground of the SRM/RTG. Apply the currents in the following table and check that the appropriate display segment is illuminated.
Page 5
Page 5 of 5 SRM/RTG Manual 14 December 2007 Revision 2.0...