Page 2
If you have any feedback on our products or services, we would like to hear from you. Please send all feedback to: Manufacturer: SIA Biosan Ratsupites iela 7 k-2, Riga, LV-1067, Latvia Phone: +371 674 261 37 Fax: +371 674 281 01 https://biosan.lv...
EU Declaration of conformity .................. 33 About this edition of user instructions The current edition of the user instructions applies to the following models: Model Version Labaqua Trace, water purification system V.5A01 Labaqua HPLC, water purification system V.5A02 Labaqua Bio, water purification system V.5A03...
Safety precautions Symbols used in the manual Caution! Make sure you have fully read and understood the present instructions before using the equipment. Please pay special attention to sections marked by this symbol. Caution! Do not service the unit with an operating UV compartment. Otherwise, operator can receive dangerous levels of UV emission.
Page 5
Electrical safety • Use a power line regulator if the local mains network is susceptible to fluctuations exceeding 10% of the nominal values. • Connect only to the mains with voltage corresponding to that on the serial number label. • When replacing fuses, re-check its placements.
General information Unit overview. Biosan Labaqua is Ultrapure water purification system which requires external water tank that use tap water as feed water (tap water system). There are three models available: Labaqua Trace, Labaqua HPLC and Labaqua Bio. Water purification system Biosan Labaqua produces pure water that complies with ISO 3696 Grade II and Grade I water requirements.
Page 7
1. Power switch (on the back panel) 2. Water access connection (on the back panel). 3. Control panel and display. 4. Filter and clean water dispenser. 5. Pre-filter set, deionization and polish- ing modules (behind side door) Figure 2. Labaqua Trace scheme Edition 5.02...
Page 8
Figure 3. Labaqua HPLC / Bio scheme Operation principle. 3.2.1 The hydraulic diagram of the Biosan Labqua water purification system can be seen in figures 2 and 3. 3.2.2 The input valve controls intake of feed water from the access. The first cleaning stage is the pre-filter set that uses activated carbon to remove particles, free chlorine, or- ganics and colloids.
There must be at least one filter on the feedwater access pipe. Filter is available in the local plumber shops or by ordering at Biosan. Edition 5.02...
Page 10
Figure 4. Filter on the water access pipe. Caution! Please note that if the pre-filter clogs prematurely, such malfunction will be covered by warranty only of the service department receives photo- graphic proof of the particle filter installed in the feedwater access pipe. 4.3.3 Feedwater connection port has to be ½"...
Caution! Due to the high weight of the unit, its unpacking and installing must be carried out by two persons. Complete set. Package contents: 5.2.1 Standard set: Water purificatiom system, Labaqua Trace/HPLC/Bio ........1 pce. Storage tank Pro ....................1 pce. Pre-filter cartridges ................... 2 pcs. Polishing module ....................1 pce.
Page 12
5.3.2 Tube connection. Push the tube into the fitting to the stop (fig. 6/a) and tug back to ensure the clamping (fig. 6/b). Figure 6. Connecting tubes 5.3.3 Disconnecting tubes. Ensure that the system is not under pressure. Press the clamp sleeve on the fitting (fig.
Page 13
Bottom Figure 9. Rear panel. 1. Unit info and serial number. 2. Connector for the Flow Point 3 water dispenser. 3. Service socket. 4. Tank water level sensor socket. 5. Mains’ cable socket and fuse socket. 6. Power switch. 7. Connector for the Flow Point 3 water dispenser. 8. Feedwater inlet fitting. 9.
Page 14
Order of filter and module installation. Open the side door (fig. 1/5) and install cleaning modules in the following order: the deionization module and the polishing module first (in sockets on fig. 8/1 and 8/2), then the pre-filters in sockets on figure 8/3.
Page 15
5.7.3 Prepare two straight angle module connecting fittings (fig. 13/a) and insert them in the module fittings to the stop, so that the free ends are rotated towards the side of the module with guide rail (fig. 13/b). Check the connecting fittings for a tight fit. Figure 13.
Page 16
Pre-filter cartridge installation. Turn and remove the protective cap from both of the cartridges (fig. 15/a). Insert a cartridge in the socket (fig. 1/3 and 15/b) and lock it in place by turning cartridge a quarter turn to the right ( fig. 15/c) Check the cartridge for a tight fit.
Page 17
5.10 Connecting the storage tank to the system. Position the tank on an even horizontal surface (see 4.6). Locate two pairs of REC IN and REC OUT fittings: one pair on the unit (fig. 9/9 and 9/10) and one pair on the tank (fig. 10/6 and 10/7) 5.10.1 Connect the REC IN fitting on the rear panel of the unit (fig.
Page 18
5.10.4 Connect the water level sensor cable (fig 19/a) to the socket on the tank (fig. 19/b & 10/1) and to the socket on the unit (fig. 19/c & 9/4). Figure 19. Water level data cable 5.11 Installation of dispenser filter. 5.11.1 For Trace and HPLC models.
Page 19
5.11.2 For Bio model. Unpack the dispenser ultrafilter (fig. 21/a) and the yellow sealing ring (fig. 21/b). Screw the filter in the slot below the display so that the ring stays on top of the thread of the filter (fig. 21/c) in the slot as shown on figure 21/d. Blue bell cap must cover the dispenser at all times when the water is not being dispensed.
Operation Open feedwater access (see 4.4). Check the power cable for damages and connect the plug to a grounded mains socket. Turn the power switch (fig. 9/6) to I (on). After several second of loading, the display shows (figure 22): •...
Page 21
6.5.1 Diagnostic screen (figure 23). Press the Mode key twice in quick succession. Display will show the diagnostic screen that provide information and status for all sensors and components, as well as the tank water level. Figure 23. Diagnostics mode. 1.
Page 22
Menu of the unit (figure 24). Additional settings for the unit are available by press- ing the Menu key (fig. 22/2). To navigate the menu, use arrow keys (fig. 22/9), to select an item, press the ОК key (fig. 22/10), to return to the previous level or to exit the menu –...
Page 23
6.9.2 Parameters (figure 27). • Measurement units (figure 28). Choose between μS / cm or MOhm * cm. • Prevent dispensing poor quality (ON/OFF). Water dispense will not start until the wa- ter quality exceeds threshold. • Recirculation (figure 29). Setup the recirculation schedule. First, the Recirculation pe- riod (figure 30): every 10, 25 or 50 minutes.
Page 24
• Temperature compensation (figure 33). Turn on and off the DI and RO water temper- ature compensation that affects conductivity measurement. Note. Measurement units, time and date must be set before using temperature compensation. • Night mode (ON/OFF). Turn off the booster pump and oxidation UV lamps between 21:00 and 7:00.
Page 25
6.10 Dispensing Grade I water. After rinsing (see 6.6) and filling the tank (see 6.7), remove the bell cap and place the vessel for purified water. Press the PUMP / Grade I key (fig. 22/4). Unit performs a 5 to 10 s internal rinsing and starts dispensing purified water.
+15°C to +30°C (59 to 86 °F) in a non-condensing atmosphere and relative humidity between 20% and 80%. Biosan is committed to a continuous programme of improvement and reserves the right to alter design and specifications of the equipment without additional notice.
V.5A02 BS-070104-A02 Labaqua Bio, water purification system V.5A03 BS-070106-A02 To inquire about or order optional or replacement parts, contact Biosan or your local Biosan representative. 8.2.1 Optional parts Description Catalogue number External pre-filter set (1 μm filter, carbon) with manometer BS-070104-KK External pre-filter set (1 μm filter, carbon, polyphosphate) with manometer...
9.1.1 If the unit is disabled (e.g., no water dispensing, no reaction to key presses, etc.) or requires maintenance, follow the troubleshooting table below. If the problem persists or is not listed, disconnect the unit from the mains and contact Biosan or your local Biosan representative.
Page 29
9.2.2 To disinfect the exterior plastic and metal parts, use 75% ethanol or DNA/RNA re- moving solution (e.g., Biosan PDS-250). After disinfecting, wipe the surfaces dry. 9.2.3 No parts of the unit are autoclavable. Recirculator disinfection.
Page 30
Service schedule. Use only the components recommended by the manufacturer. Replace components according to the table below: Table 5. Component replacement period Replacement part Replacement period Note Stage changes between When the filters are clogged or every 6 months. Pre-filter cartridges Filling tank and Low pres- Replace both simultaneously.
Page 31
9.5.3 Polishing module. To replace, use the tool on figure 5. Put the unit in the OFF stage by pressing the Run key; Open the side door; Insert the tool under the right-angle fitting (fig. 34/a) and press down (fig. 34/b); Rotate the fittings inside (fig.
To obtain the claim form, visit Technical support page on our website at link below. 11.5 Extended warranty. Contact your local Biosan representative or our service depart- ment through the Technical support section on our website at the link below.
12. EU Declaration of conformity 12.1 Water purification systems Labaqua Trace, Labaqua HPLC and Labaqua Bio are in conformity with the following relevant Union legislations: LVS EN 61010-1:2011 Safety requirements for electrical equipment for LVD 2014/35/EU measurement, control, and laboratory use. General requirements.