Waterlase Fractional Handpiece
Instructions for Use

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Waterlase and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Biolase Waterlase

  • Page 1 Waterlase Fractional Handpiece Instructions for Use...
  • Page 2 Section 2: Contraindications, Warnings, and Precautions Contraindications, Warnings, and Precautions related to the use of the Waterlase iPlus laser are provided in the Waterlase iPlus User Manual. Clinician must know and adhere to state guidelines and regulations when purchasing and using this accessory. Check with your state and/or local government for required permits or other applicable laws.
  • Page 3: Section 3: Overview

    Section 3: Overview Fractional Handpiece and Disposable Applicator Specifications Maximum Up to 9 W at 15 Hz Spot spacing: 0.9 mm & 1.0mm* in Y direction (*based on movement speed of 1.5 cm/second) power Setting: Total energy/pulse: 100 - 400 mJ Fluence per spot: 20 - 82 J/cm2 Energy per spot:...
  • Page 4 HANDPIECE HOLDER The Handpiece Holder Attachment is designed to fit over the built-in Waterlase iPlus Handpiece Holder on the laser console and, once in place, will support the Fractional handpiece as well as all other Waterlase iPlus handpieces. A Slide the attachment from the back toward the front onto the existing Handpiece Holder on the laser console.
  • Page 5: Section 4: Cleaning And Disinfection

    The Handpiece To clean and disinfect the Waterlase Fractional Handpiece, wipe the entire outer surface up to the laser optics housing using a cotton gauze and an appropriate disinfecting solution, such as Cavicide™, or a similar quaternary ammonium compound product (containing 20% alcohol or less), and follow the manufacturer’s instructions.
  • Page 6 Section 5: Set-Up Refer to the Waterlase iPlus User Manual for detailed instructions on system set-up, installation and operation. CONNECTING THE FRACTIONAL HANDPIECE TO THE FIBER OPTIC CABLE • If switching from a Gold or Turbo Handpiece to the Fractional Handpiece, always purge the Handpiece before disconnecting it from the Fiber Optic Cable.
  • Page 7 Figure 5A the Fiber Optic Cable. D. To unplug the auxiliary air hose from the Waterlase iPlus laser, press down on the release tab at the top of the of the auxiliary air output port and gently pull on the connector to detach.
  • Page 8: Section 6: Operation

    Refer to the Waterlase iPlus User Manual for instructions on how to turn on and activate the Waterlase iPlus All-Tissue Laser. Make sure the Handpiece is properly connected to the Fiber Optic Cable and the auxiliary air hose is connected to the Waterlase iPlus.
  • Page 9: Section 7: Clinical Applications

    Section 7: Clinical Applications The Waterlase Fractional Handpiece generates 10 microbeams in a single line per laser pulse. Treatment is always in contact mode, i.e., the Applicator attached to the distal end of the Fractional Handpiece is applied directly to the skin while the laser is activated.
  • Page 10 It is also possible to switch from H mode (narrow coagulation zone) to S mode (shallow penetration in tissue), as needed, based on the patient’s skin type. NOTE: Water, when combined with air, creates a spray that cools the treatment site and facilitates the effects of the laser energy applied. Frequency cannot be modified, and remains fixed.
  • Page 11: Section 8: Warranty And Liability

    Limited Liability BIOLASE, Inc. will not be liable for incidental, consequential, indirect or special damages of any kind including, but not limited to, damages for loss of revenue, loss of business or business opportunity or other similar financial loss arising out of or in connection with the performance, use or interrupted use of the BIOLASE laser system(s) or any BIOLASE materials.
  • Page 12 27042 Towne Centre Drive, Suite 270 M T Promedt Consulting GmbH Foothill Ranch, CA 92610 USA Altenhofstrasse 80 Tele: +1 949.361.1200 D-66386 St. Ingbert/Germany Toll Free: 833.BIOLASE +49 6894 581020 Rx Only www.mt-procons.com Service: 800.321.6717 Fax: +1 949.273.6687 biolase.com P/N: 5201615 Rev F...