iET HRRS Series User And Service Manual

iET HRRS Series User And Service Manual

5kv and 10kv high resistance, 5kv and 10kv decade substituter

Advertisement

Quick Links

¨
HRRS 5kV and 10kV im/August 2021
IET LABS, INC.
1.888.610.7664
PRECISION INSTRUMENTS FOR TEST AND MEASUREMENT
HRRS 5kV AND 10kV SERIES
High Resistance, 5kV and 10kV
Decade Substituter
User and Service Manual
Copyright © 2021 IET Labs, Inc.
www.calcert.com
15
10
20
5
25
0
30
¨
sales@calcert.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the HRRS Series and is the answer not in the manual?

Questions and answers

Summary of Contents for iET HRRS Series

  • Page 1 PRECISION INSTRUMENTS FOR TEST AND MEASUREMENT HRRS 5kV AND 10kV SERIES High Resistance, 5kV and 10kV Decade Substituter User and Service Manual Copyright © 2021 IET Labs, Inc. HRRS 5kV and 10kV im/August 2021 IET LABS, INC. 1.888.610.7664 www.calcert.com sales@calcert.com...
  • Page 2 ¨ ¨ PRECISION INSTRUMENTS FOR TEST AND MEASUREMENT IET LABS, INC. 1.888.610.7664 www.calcert.com sales@calcert.com...
  • Page 3 IET specifi cations. If within one year after original shipment, it is found not to meet this standard, it will be repaired or, at the option of IET, replaced at no charge when returned to IET. Changes in this product not approved by IET or application of voltages or currents greater than those allowed by the specifi...
  • Page 4 Such noncompliance would also violate safety standards of design, manufacture, and intended use of the instrument. IET Labs assumes no liability for the customer’s failure to comply with these precautions. The HRRS complies with EN 60477.
  • Page 5 To avoid the danger of introducing additional hazards, do not install substitute parts or perform unauthorized modifi cations to the instrument. Return the instrument to an IET Labs for service and repair to ensure that safety features are maintained in operational condition.
  • Page 6 HRRS 5kV and 10kV Series Safety Symbols General defi nitions of safety symbols used on the instrument or in manuals are listed below. Caution symbol: the product is marked with this symbol when it is necessary for the user to refer to the instruction manual.
  • Page 7: Table Of Contents

    HRRS 5kV and 10kV Series Table of Contents Chapter 1 Introduction 1.1 Overview ....................... 1 Chapter 2 Specifi cations 2.1 Specifi cations ......................2 2.2 Ordering Information .................... 3 Chapter 3 Operation 3.1 Initial Inspection and Setup .................. 5 3.2 Dial Setting ......................5 3.3 Environmental Conditions ..................
  • Page 8 HRRS 5kV and 10kV Series This page is intentionally left blank. 1.888.610.7664 www.calcert.com sales@calcert.com...
  • Page 9: Chapter 1 Introduction

    HRRS 5kV and 10kV Series Chapter 1 INTRODUCTION 1.1 Overview The High Resistance, 5kV and 10 kV Decade To ensure low leakage, the connection terminals are insulated with Kel-F high-resistance, non-moisture- Substituters (Figure 1-1 and Figure 1-2) is a family absorbing material.
  • Page 10: Chapter 2 Specifi Cations

    HRRS 5kV and 10kV Series Chapter 2 SPECIFICATIONS For convenience to the user, the pertinent specifi cations are given in an OPERATING GUIDE affi xed to the case of the instrument. Figure 2-1 shows a typical example. 2.1 Specifi cations Accuracy ** HRRS-5kV HRRS-10kV...
  • Page 11: Ordering Information

    HRRS 5kV and 10kV Series 2.2 Ordering Information Number of Number of Model Total resistance Resolution Model Total resistance Resolution decades decades HRRS-*-7-10-** 111.111 1 MΩ 10 Ω HRRS-*-3-10k-** 11.1 MΩ 10 kΩ HRRS-*-7-100-** 1.111 111 GΩ 100 Ω HRRS-*-3-100k-** 111 MΩ...
  • Page 12 HRRS 5kV and 10kV Series Specifi cations 1.888.610.7664 www.calcert.com sales@calcert.com...
  • Page 13: Chapter 3 Operation

    HRRS 5kV and 10kV Series Chapter 3 OPERATION 3.1 Initial Inspection and Setup and a setting of 10-10-10-l0-10.10 W becomes: This instrument was carefully inspected before ship- 1 0 0 0 0 0 . 0 ment. It should be in proper electrical and mechanical 1 0 0 0 0 .
  • Page 14: Storage

    HRRS-WT can easily fi t inside. A faraday cage can also easily and inexpensively be build using copper mesh over a frame. IET Labs uses a similar faraday cage during calibration of the HRRS series.
  • Page 15 HRRS 5kV and 10kV Series This page is intentionally left blank. Operation 1.888.610.7664 www.calcert.com sales@calcert.com...
  • Page 16: Chapter 4 Maintenance

    (12) months. This procedure may be carried out by the user if a calibration capability is available, by IET Labs, or by a certifi ed calibration laboratory. Maintenance 1.888.610.7664 www.calcert.com...
  • Page 17: Calibration Procedure

    HRRS 5kV and 10kV Series 4.4 Schematic WARNING Connect the GND terminal to earth Figure 4-1 shows a schematic of an HRRS-5kV/10kV or other suitable ground in order to unit. maintain the case at a safe voltage. W h e n e v e r h a z a rd o u s vo l t a g e s (>45 v) are used, take all measures to avoid accidental contact with any live components.
  • Page 18: Replaceable Parts

    HRRS 5kV and 10kV Series 4.5 Replaceable Parts Model Ref IET part number Description BP-1000-RD Binding Post, Red BP-1000-BK Binding Post, Black BP-1000-BL Binding Post, Blue BP-1000-GN Binding Post, Green HARS-X-4300-KNB Knob Assembly Not Shown HARS-X-3100 Foot Not Shown HRRS-*-SW-10-** 10 Ω/step Decade Switch Assembly...