Chapter 3 -† Sentinel Control System
3.1.桠 Sentinel Control System
LCD
Display
1.†Fan Button
1 † Turns on and turns off the fan.
1 † Activate Standby mode
2.†Lamp Button
1 † Turns on and turns off the fluorescent lamps.
3.†Socket Button
1 † Turns on and turns off the electrical socket (retrofit kit).
1 † The maximum rating of all the outlets in the cabinet is 5 A. If there is overload, the fuse will
blow.
4.†UV/Gas Button
1 † Turns on and turns off the gas.
1 † Turns on and turns off the UV lamp.
1 † UV lamp can only be activated when the sash window is fully closed. Since the sash is
capable of filtering UV rays, users are protected from the harmful UV radiation.
5.†Up (▲) and Down (▼) Arrow Button
1 † Moves the menu options upwards and downwards.
1 † Increases and decreases corresponding value inside one of the menu options.
1 † Moves the sash window upward and downward.
6.†Set or Mute or Diagnostic Button
1 † To proceed to the next step, level or sequence inside the menu options.
1 † Mutes the fully opened sash and airfail alarm sound (during normal and quickstart mode).
1 † Enters diagnostic mode.
7.†Menu Button
When you are entering menu options, the alarm will sound to indicate that the microprocessor is
not monitoring the operation of the cabinet. No further warnings will be given.
1 † To enter and exit from the menu options.
Menu
Button
Set/Mute/
Visual
Diagnostic Button
Alarm
Figure 2.1. Sentinel Gold general parts.
Up
Fan
Socket
Button
Button
Button
Down
Lamp
Button
Button
UV/Gas
Button