  • Page 1 User Guide Medisafe Plus 80 Medisafe Plus 311 Medisafe Plus 371 Powered by...
  • Page 2 Medisafe Plus Warranty Your new Medisafe Plus vaccine refrigerator is covered by a 24-month warranty on parts and labour. Avem Quirks Australia will repair or replace, at our discretion, any component found to be faulty in workmanship or operation for a period of 24 months from the time of installation.
  • Page 3: Table Of Contents

    Contents Acknowledgement, Contact and Warranty Information Getting Started Installation Basic Operation Stocking your Medisafe Plus Refrigerator Silencing the Alarm Light Activation Temperature Control System Min-Max Temperature Display Medisafe Plus Control and Monitoring System AQ Box Cloud Device Logging in Dashboard Overview User Settings Running and Printing Reports Troubleshooting...
  • Page 4: Getting Started

    Ensure the power cord is not damaged. If the cord becomes damaged, it must be replaced by the manufacturer, a certified service agent or qualified service professional. It is strongly recommended that any servicing of the unit be performed by Avem Quirks. Contact us on 1800 658 120 or servicedesk@avemquirks.com.au Installation •...
  • Page 5: Basic Operation

    Basic Operation-First use • As advised on previous page, the power failure alert “PF” should be acknowledged by pressing the red “SET” button once on the Min-Max temperature display. • It is recommended the Min/Max Thermometer is reset on first start up. Press and hold the red ‘SET’...
  • Page 6: Silencing The Alarm

    Press SET to mute the alarm Light Activation Medisafe Plus 80 model: The light can be controlled on the internal fan plate. Medisafe Plus 311 and 371 models: The light can be controlled from the toggle switch, located on the underside of the cover strip, above the door.
  • Page 7: Temperature Control System

    The only button you are required to press is the PRG/MUTE button. If the cabinet alarm sounds, you can press this to silence it. Any reprogramming of the controller must be done by a qualified Avem Quirks service technician. If an alarm has sounded, please allow 10-15 minutes for recovery.
  • Page 8: Min-Max Temperature Display

    Min-Max Temperature Display • The top left display is the minimum and maximum temperature recorder. This displays the current air temperature inside the cabinet, and also provides the minimum and maximum air temperatures reached inside the refrigerator since the last reset. •...
  • Page 9: Logging In

    AQ Box Cloud Temperature Monitoring Device All new Medisafe Plus vaccine refrigerators are pre-installed with the AQ Box Cloud Temperature monitor. Please complete the supplied Medisafe Plus ‘Cloud Monitoring Product Registration’ form and return it to medisafeplus@avemquirks.com.au to gain access to your portal and all reporting. •...
  • Page 10: User Settings

    User Settings Under the ‘USER SETTINGS’ tab is where you can manage passwords and notifications.
  • Page 11 Running Temperature Reports • Under REPORTS choose - “Device Temperature” • Select the Device in the dropdown box • Choose the desired ‘From’ date and time • Choose the desired ‘To’ date and time • Choose ‘Reading every (minutes) - this can be from 1-99 minutes.
  • Page 12: Troubleshooting

    Before turning on, check to see if it is safe! Check fuses from main power supply. If there is an electrical supply issue, please contact your electrician. If it is a Medisafe fridge issue, please contact Avem Quirks on 1800 658 120. Why did the fridge alarm go off...
  • Page 13 Can I obtain my own SIM card? No, you cannot. The SIM card placed inside the AQ Box has a serial number linked to an Avem Quirks software license. Replacing this with another SIM card that is not registered with us, will not allow you to view your data and the graphs via the cloud.
  • Page 14 Issue Solution Can I buy more GLYCEROL B.P Yes, you can acquire this at most local pharmacies. 100mL bottles of Glycerol B.P are in case of a mishap? available at most pharmacies for approximately $5.00. However, it is best to call our service team as this means the ‘putty/sealant compound’...
  • Page 15 Medisafe Plus refrigeration units come with a 24-month parts and labour warranty. The AQ Box temperature monitoring device remains the property of Avem Quirks at all times. As such, it is warranted for parts and labour during paid and active subscription periods.