Download Print this page
Next NEXT600-S User Manual

Next NEXT600-S User Manual

Peristaltic pump

Advertisement

Quick Links

peristaltic pump User Manual
NEXT600-S

Advertisement

loading

Summary of Contents for Next NEXT600-S

  • Page 1 User Manual NEXT600-S...
  • Page 2 : NOTED Before operating, please read this User Manual carefully. Warning: Due to the wear, may cause the tube cracks. If it happened, the liquid will overflow from the tube, then may damage the human body and equipment, so please often check and timely replace the tube! Please ensure plug the power cord directly into a wall outlet, and to avoid the use of extension cords...
  • Page 3 TABLE OF CONTENTS Product Introductions Operation Panel Menu Descriptions Operating instructions Technical Parameters Suitable Pump Heads & Tubes External Interface Functions & Features Flow Rate Descriptions Pump Tubes Descriptions Pump Tubes Parameters Maintenance, Warranty & Service...
  • Page 4 ◆Multi channel transmission is widely used in bulk liquid interface E built-in fuse F power socket --NEXT600-S Digital distribution type pump Velocity range Note: the use of external control and see page 8 1 . 3 ~ 1 2 0 0 0 m l / m i n...
  • Page 5 The peristaltic pump is composed of three parts: driver, pump head, pump assembly is simple, easy to understand. This manual will provide various parameters of peristaltic pump value and operation method for you. Peristaltic pump operating panel NEXT600-S NEXT600-S 4 color LED display Speed, velocity Speed / full speed knob...
  • Page 6 Operating instructions ◆ Start-stop direction control Reversible left - stop - right switch place in the "left", the pump running at setting speed counterclockwise; place in the "right", the pump running at setting speed clockwise; place in "stop" the pump stop running. Speed ​...
  • Page 7 C3-0 R s 4 8 5 E N A B L E L E V E L ( D E F A U L T ) C 3 - 1 RS485 DISALE Co-0 P U L S E (DEFAULT ) C o - 1 EXIT 4 8 5 E N A B L E...
  • Page 8 Application one: speed mode transmission Transmission fluid to rotate speed of 80rpm/min. A206 Dg(10) ⑴Press the button to enter the (display of choice). A207 KZ25 ⑵Press the knob to enter (and rotate speed display mode). ⑶Press the knob to determine the current model for the speed display mode. A208 Yz35 ⑷To rotate and press the knob (return to the main interface).
  • Page 9 Part of the pump head and pump Optional pump head, pump type T U B E SUITABLE PUMP HEAD & FLOWRATE REFERENCE M O D E DRIVE MODE NEXT600-S Note: the use of the pump head to see the pump head vser manual.
  • Page 10 Pump Head Tube Corresponding To The Table The pursuit of excellence The pursuit of excellence The creation of customer value The creation of customer value PUMP HEAD MODE TUBE NO TUBE MODE PUMP HEAD MODE TUBE NO TUBE MODE - 1 3 - - 1 3 - DEFAULT DEFAULT...
  • Page 11 The pursuit of excellence The pursuit of excellence The creation of customer value The creation of customer value - 1 5 - 1 5 # - 8 6 - 8 6 # D E F A U L T D E F A U L T - 2 4 - 2 4 # - 9 0 -...
  • Page 12 DRIVE EXTROCONTROL INTERFACE (DB-15) DESCRIPTION 15-pin connector on the back of the machine is done by an external signal to control the machine interface. First prepare a DB15 connector (hole), and several different colors of the signal cable to the 15-pin female connector to open, according to the following sequence diagram welding wire signal lines of different colors, and the signal lines 15-pin connector clip line fixed together, if necessary, can be reinforced plastic injection, and finally installed on the shell...
  • Page 13 Import Hose And Its Characteristics C h e m i c a l ◆ resistance long time Adsorption of Low Resistant to strong Strong heat resistance Corrosion resistance acid (such c o n c e n t r a t e d In accordance with America standard s u l f u r i c a c i d...
  • Page 14 Hose Parameters Micro flow tube Tube Hose section Wall thickness Inside diameter Continuous Pressure Interval The basic flow tube Tube Hose section Wall thickness Inside diameter Continuous Pressure Interval Industrial hose Tube Hose section Wall thickness Inside diameter Continuous Pressure Interval Accessories Optional FK-1C...
  • Page 15 Trouble shooting F a u l t D i s p l a y R e a s o n M e t h o d Motherboard insurance Change It Open the pump, The fan and display are blown fuse normal work ,to control the start /stop , the pump not work.