Removable hard drive ssd for 16860-series logic analyzers (22 pages)
Summary of Contents for Keysight W4630 Series
Page 1
Keysight W4630 Series DDR4 BGA Interposer Installation Guide...
Page 2
Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR Keysight W4630 Series DDR4 DRAM BGA Interposers Installation Guide...
In This Guide This document provides installation information for the following Keysight products: • W4633A DDR4 x4/x8 BGA Command and Data Interposer • E5849A High Data Rate Single-ended ZIF Cable (for use with the W4633A interposer) Keysight W4630 Series DDR4 DRAM BGA Interposers Installation Guide...
Page 4
Keysight W4630 Series DDR4 DRAM BGA Interposers Installation Guide...
Connecting the E5849A Probe Cables to a Logic Analyzer Logic Analyzer Channel Mapping Signals not probed by the Logic Analyzer 3 Setting Up the Logic Analyzer for W4630 Series Probes Before You Start Load ing a Configuration File To save a configuration file...
Page 6
Guidelines for Shelf Life and Solder-ability of W4633A Interposers Cleaning of W4633A Interposer Gold Fingers Safety Information for the E5849A Cable Adapters To clean the instrument Safety Symbols Regulatory Notices WEEE Compliance China RoHS Index Keysight W4630 Series DDR4 DRAM BGA Interposers Installation Guide...
Keysight W4630 Series DDR4 DRAM BGA Interposers Installation Guide Introduction W4630 Series DDR4 BGA Interposer - Introduction / 8 Hardware and Software Requirements / 12 Mechanical Considerations / 13 DDR3 Probes / 17 This chapter introduces the hardware components that are needed for a W4630-series interposer setup.
DRAM would normally be soldered. The interposer is designed to be soldered to the PCB footprint for the DRAM on top of either the DDR4 riser included with the W4630 series BGA interposer or an optional Grypper socket (not included with the interposer). The DRAM being probed is then soldered to the top side of the interposer.
DDR4 78 ball DRAM. To know how to solder the riser to the interposer and PC board, refer to the topic “Soldering the W4633A Interposer and Riser" on page 21. Keysight W4630 Series DDR4 DRAM BGA Interposers Installation Guide...
E5849A Single-ended ZIF probe cable To know how to make connections between the W4633A interposer and E5849A probe cables, refer to the topic “Connecting the W4633A Interposer to E5849A Probe Cables" on page 23. Keysight W4630 Series DDR4 DRAM BGA Interposers Installation Guide...
U4201A 90-pin logic analyzer cable To know how to make connections between the U4201A cable and E5849A probe cable, refer to the topic “Connecting the E5849A Probe Cables to a Logic Analyzer" on page 27. Keysight W4630 Series DDR4 DRAM BGA Interposers Installation Guide...
Access to Number of Number of ZIF Number of Number of Logic Analyzer Mod ules Interposers Probes Cables Command, One W4633A Two E5849A Four U4201A One U4154A module Address, Control and Data Keysight W4630 Series DDR4 DRAM BGA Interposers Installation Guide...
Introduction Mechanical Considerations W4633A Interposer Dimensions The following figure shows the dimensions of a W4633A DDR4 DRAM BGA interposer. Figure 5 Dimensions of a W4633A interposer Keysight W4630 Series DDR4 DRAM BGA Interposers Installation Guide...
You must install the W4633A interposer on a riser (shipped with the interposer) or a grypper NOTE socket to provide clearance to surrounding DRAM and RCs on the bottom side of the interposer. Keysight W4630 Series DDR4 DRAM BGA Interposers Installation Guide...
Page 15
Introduction Figure 7 KOV of a W4633A interposer The following figure shows the KOV of an E5849A probe cable when connected to a W4633A interposer. Keysight W4630 Series DDR4 DRAM BGA Interposers Installation Guide...
Page 16
Introduction Figure 8 KOV of a W4633A interposer with an E5849A cable (SIDE VIEW) Keysight W4630 Series DDR4 DRAM BGA Interposers Installation Guide...
Keysight W4630 Series DDR4 DRAM BGA Interposers Installation Guide Setting up the W4630-Series Interposers W4633A Interposer Setup - Overview / 20 Soldering the W4633A Interposer and Riser / 21 Connecting the W4633A Interposer to E5849A Probe Cables / 23 Connecting the E5849A Probe Cables to a Logic Analyzer / 27...
Use a grounded wrist strap and other ESD control measures as appropriate. Do not open the vacuum sealed packs of the W4633A interposer until you are ready to install NOTE the interposer. Discard these packs once the package is opened. Keysight W4630 Series DDR4 DRAM BGA Interposers Installation Guide...
• When possible use a PCA rework vendor, to attach the riser, interposer and memory components, such as: Circuit Technologies Center. More useful information can be found at: http://www.circuitrework.com. Keysight W4630 Series DDR4 DRAM BGA Interposers Installation Guide...
When the W4633A interposer is soldered to a riser, flatness must be maintained on the order of 3.5 mils or less across the BGA footprint to maximize successful soldering to the interposer. Keysight W4630 Series DDR4 DRAM BGA Interposers Installation Guide...
The right wing of Cable 2 connects to P3-Address/Command wing of the interposer while the left wing of Cable 2 has not been used and left unconnected. Figure 10 W4633A Interposer and E5849A Cable Connections Keysight W4630 Series DDR4 DRAM BGA Interposers Installation Guide...
Page 24
To attach an E5849A ZIF connector to a flex wing of the W4633A interposer, perform the following three steps. Angle the flex wing of the interposer into an E5849A ZIF connector. GND towards door closure. Align the E5849A ZIF connector tabs with interposer’s wing notches. Keysight W4630 Series DDR4 DRAM BGA Interposers Installation Guide...
When you connect the E5849A probe cables to a U4154A Logic Analyzer as per the diagram in Figure 12, the logic analyzer channels are mapped to signals as per the table displayed below. Keysight W4630 Series DDR4 DRAM BGA Interposers Installation Guide...
U4154A Logic Analyzer Pods Signals not probed by the Logic Analyzer The following signals are omitted from the Logic Analyzer connections for the W4633A interposer. Interposer Signal Name W4633A VREFCA, TEN, ZQ Keysight W4630 Series DDR4 DRAM BGA Interposers Installation Guide...
Keysight W4630 Series DDR4 DRAM BGA Interposers Installation Guide Setting Up the Logic Analyzer for W4630 Series Probes Before You Start / 30 Loading a Configuration File / 31...
Setting Up the Logic Analyzer for W4630 Series Probes Before You Start Ensure that all the software components listed in the topic “Hardware and Software Requirements" on page 12 are installed on the host computer and the required software licenses are also obtained and installed.
Because of these dependencies, there is no single logic analyzer configuration file setup for the W4630-series interposers. A set of DDR4 x4 and x8 configuration files is provided with the Keysight B4621B Decoder software. To obtain this set of configuration files, you must install the B4621B software (licensing of this software is not required for obtaining these configurations).
Page 32
Setting Up the Logic Analyzer for W4630 Series Probes Keysight W4630 Series DDR4 DRAM BGA Interposers Installation Guide...
Keysight W4630 Series DDR4 DRAM BGA Interposers Installation Guide Characteristics, Regulatory, Safety and Storage Information Operating Characteristics / 34 Storage, Inspection, Baking, and Cleaning Guidelines / 35 Safety Information for the E5849A Cable Adapters / 36 Regulatory Notices / 37...
Up to 50% noncondensing. Avoid sudden, extreme temperature changes which could cause condensation on the circuit board. For indoor use only. Characteristics Description To interposer Memory bus signals from target system From interposer High-density connectors for Keysight U4154A AXIe-based logic analyzer modules Keysight W4630 Series DDR4 DRAM BGA Interposers Installation Guide...
If the test passes, proceed with the assembly (reflow) of interposer. • If delamination occurs, the interposer cannot be used. Cleaning of W4633A Interposer Gold Fingers • Use Isopropyl alcohol to clean W4633A interposer contacts. • Never use abrasive cleaning materials. Keysight W4630 Series DDR4 DRAM BGA Interposers Installation Guide...
"Caution" or "Warning" risk of danger marked on product. See "Safety Notices" on page 2 and refer to this manual for a description of the specific danger. Hazardous voltage symbol. Earth terminal symbol: Used to indicate a circuit common connected to grounded chassis. Keysight W4630 Series DDR4 DRAM BGA Interposers Installation Guide...
Product Category: With reference to the equipment types in the WEEE Directive Annex I, this product is classed as a "Monitoring and Control Instrumentation" product. Do not d ispose in domestic household waste. To return unwanted products, contact your local Keysight office, or see "www.keysight.com"...
Page 38
Characteristics, Regulatory, Safety and Storage Information Keysight W4630 Series DDR4 DRAM BGA Interposers Installation Guide...
Page 39
U4201A, 11, interposer, inputs to, interposer, outputs from, introduction, W4633A, W4633A interposer setup, WEEE compliance, keep-out volume, Keysight W4630 Series DDR4 DRAM BGA Interposers Installation Guide...
Page 40
Index Keysight W4630 Series DDR4 DRAM BGA Interposers Installation Guide...