MX Health MX86001 (AS-302-L) - Fingertip Pulse Oximeter Manual

MX Health MX86001 (AS-302-L) - Fingertip Pulse Oximeter Manual

Advertisement

MX Health MX86001 (AS-302-L) - Fingertip Pulse Oximeter Manual

Product Features

The product is easy to operate. It is small in dimensions, light in weight, convenient to carry around. Power consumption is low. Ambient light does not interfere with the performance of this device. Two AAA-batteries can give up to 12 hours of operation (not included). The battery voltage is indicated. The product is suitable for use in the following settings: Home, surgery, intensive care, medical centre, sports health facilities and the like.

Intended Use

The fingertip pulse oximeter is a portable non-invasive device intended for spot checking oxygen saturation of arterial haemoglobin SpO2 and the pulse rate of patients in hospitals, hospital-type facilities, and the home environments.
This device is suitable for use on patients of all skin tones and is not suitable for use on infants and neonates. No special training is required to use the device but careful reading of the instructions for use is a requirement. Adults should be able to understand and read this manual with ease as it is written in layman's terms.

Principal of Measurement

Arterial oxygen saturation is measured by a method called pulse oximetry. It is a continuous, non-invasive method based on the different spectra absorption of and oxyhaemoglobin. It measures how much light sent from the light source on the one side of the sensor is transmitted through the patient tissue (such as a finger), to a receiver on the other side. The two beams of different wavelength of light (660nm red and 895nm near infrared light) can be focused onto a human nail tip through clamping the finger-type sensor. The measured signal obtained by a photo sensitive element, will be shown on the oximeters' display through a process in the electronic circuit and micro-processor of the device.
warning SpO2 is a measurement of the functional oxygen saturation.
warning The values of SpO2 and PR are through average processing.
warning The light (the red light and the infrared light which is invisible and emitted from the device is harmful to the eyes. The user and maintenance personal must not stare at the light.
warning The device does not have a technical specifications manual. All relevant and important information is given in this instruction for use.

Saturation %

100 - 95 95 - 90 90 - 85
GOOD NOT GOOD SEEK MEDICAL HELP

If the SpO2 reading of the oximeter is 94 or below, contact an appropriately qualified Healthcare Professional to seek advice and /or help.

Operating Instructions

Carefully read this user manual before use.
MX86001(AS-302-L)
The LED digitron display oximeter cannot change the direction

  1. Press power switch key on the front panel to turn the pulse oximeter power on.
  2. Place your one finger (forefinger or middle finger) into the cavity of the pulse oximeter. If it is necessary to use another finger the ring finger, little finger or thumb may be used.
  3. Read the data display on the screen after 3-4 seconds have elapsed. The bar graph height indicates the strength of the pulse.
  4. The device will turn itself of automatically after 20 seconds of idle time.

The finger used must not have any pathological damage.
Do not use the fingertip pulse oximeter in a situation where an alarm is required. The device has no alarm for SpO2 and pulse.
Operation of the fingertip oximeter may be affected by the use of an electro surgical unit.

warning The fingertip pulse oximeter is not suitable for continuous monitoring of the same monitoring site of the same patient. The fingertip pulse oximeter is also not suitable for use on infants and neonates. Check the finger monitoring site every 2 hours of use. If the colour of the skin changes, move the device slightly. Some patients e.g. Patients with perfusion disorder or sensitive skin may require that the device reading site be changed more often. Continuous monitoring may result in unexpected change in skin condition such as sensitivity, redness, blistering or pressure necrosis.
warning If the finger used for monitoring is not inserted into the oximeter thoroughly it may result in incorrect readings.
warning Use medicinal alcohol to clean the finger used for monitoring and the oximeter rubber touching the finger. (The oximeter rubber is made of medical silicon. The medical silicon is non-toxic, has no harmful effects to the human skin and passed the required bio-compatibility tests.)
warning Do not use the device in an MRI or CT environment. It may cause erroneous measurements.
warning Do not use the device in close proximity of flammable anaesthetic mixed with air, oxygen or nitrous oxide.
warning Portable and mobile RF communication equipment affects medical electrical equipment.
warning The materials that come into contact with the patient's skin contain medical silicon and ABS plastic enclosures. These materials have all passed the ISO 10993-5 tests for in-vitro cytotoxicity and ISO 10993-10 tests for irritation and delayed-type hypersensitivity.
warning The temperature of the device may increase because of use. If the temperature of the device never exceeds 41°C, no harm will come to the patient.
warning It is recommended that the device be used indoors and in an environment were there is no dust and strong light.
warningThe product must be stored between -20°C and +55°C and <93% humidity.
warningDo not open, take apart or dismantle the product. There are no serviceable parts in this device. Should any problems arise in using this device, contact the manufacturer.
warning Should there be a suspicion that the readings from the device is incorrect, check the patient's vital signs by an alternative method first, and then check the oximeter. Inaccurate measurements may be because of any of the following factors:

  • Ambient light radiation.
  • Patient movement.
  • Diagnostic tests.
  • Low perfusion.
  • Electromagnetic interference such as that caused by a nearby cell phone.
  • Electrical equipment.
  • Nail polish on fingernail.
  • Artificial nails.
  • Excess dirt/damage on fingers/nails.
  • Skin pigmentation (like Henna, as darker pigmentation can give an over estimated SpO2 reading.
  • warningshock hazard
    Arterial blood is to low to measure as can be caused by shock, anaemia, low temperature or use a vasoconstrictor.
  • Heavy smokers may have a high CO reading causing a high haemoglobin CO reading.
  • Patients with severe jaundice will have high bilirubin which affects CO which in turn causes high SpO2.

warning When the pulse graph display is not normal MX86001(AS-302-L), the displayed SpO2 or pulse rate reading may be incorrect.

Product readiness

Once the device has: The batteries correctly installed, and the display is lit; and the device silicone has been wiped with alcohol; and the device is fitted to a suitable finger, then the device is ready for use.


If the device does not give the desired results with the second attempt to get a proper reading, a different device should be used and the attending healthcare professional should be informed of the difficulty experienced with the device.

Product Accessories

  • 1 x Lanyard
  • 1 x Instruction Manual

Battery Installation

Operating Instructions - Battery Installation

  1. Open the battery compartment cover
  2. Install 2 x AAA batteries into battery compartment to match positive and negative indicators. Slide the battery door cover horizontally along the arrow as shown in the picture.

NOTE:
The device will automatically start, and the luminous tube lights will light up if the batteries are installed correctly. If the pulse oximeter will not be used for an extended period, remove the batteries.

warning CAUTIONS
The battery power is displayed in the middle of the screen. When the battery sign is displayed as a blank space, replace the batteries.

The batteries must be installed according to the correct polarity indicators otherwise the device may be damaged.
warning When disposing this medical device, or its batteries, follow local government procedures and laws.
warning Keep the oximeter and its accessories out of the reach of children. Small items such as the battery door, the batteries and lanyard pose choking hazard to children.

Using the Lanyard

Thread the thin end of the lanyard through the loop which is use for the lanyard. Thread the thick end through the thin end and pull tightly. Do not hang the lanyard from the device electrical wire.

Maintenance

Cleaning:
Use medical alcohol to clean the silicon on the inside of the oximeter that touches the finger. Use a soft cloth dampened with 70% isopropyl alcohol. Clean the finger to be used, using the same alcohol before and after each test.
Do not pour or spray any liquids onto the oximeter and do not allow any liquid to enter any of the openings in the device. Allow the oximeter to dry thoroughly before re-use.
This device requires no routine calibration or maintenance other than the replacement of the batteries.

warning Disinfecting:
Damage may be caused by disinfecting the device. It is recommended that the device only be disinfected on the recommendation of a doctor. The device should be cleaned before disinfecting.

warning Maintenance:
The device should be powered up for at least three hours every three months to prevent water vapour damp.
The device should not be stored in a moist environment if it is not going to be used for a long time. Should there be a hospital maintenance plan for medical devices, this plan may be adhered to.
Clean the surfaces of the fingertip pulse oximeter every time before it is used to diagnose a patient. Do not use any caustic or abrasive detergents. Store this device in a dry environment. Extreme moisture may affect the lifetime and may cause damage to the device. The expected lifetime of the device is 5 years if it is used for 15 measurements every day with a duration of 10 minutes per measurement. When the product reaches the end of its lifetime, dispose of the device according to local government recycling and disposal regulations.

warning Should one of the following occur, please contact the service representative or manufacturer:

  • The device display is not normal or there is no display at all.
  • If new batteries inserted and the device does not power on.
  • The sensor of the oximeter is damaged and causes the oximeter to function properly.
  • The sensor is damaged (not the light), indicating that the sensor is damaged. Do not continue the use of the oximeter and contact the device manager.

TECHNICAL ACCURACY

A calibration tested cannot be used to assess the accuracy of a pulse oximeter. Clinical testing is used to establish the SpO2 accuracy. The measured arterial haemoglobin saturation value (SpO2) of the sensors is compared to arterial haemoglobin oxygen (SaO2) value, determined from blood samples with a laboratory CO-oximeter. The accuracy of the sensors in comparison to the CO-oximeter samples measured over the SpO2 range of 70% to 99%. Accuracy data is calculated using the route – mean – squared (Arms value) for all subjects, as per ISO 80601-2-61: 2011, medical electrical equipment particular requirements for basic safety and essential performance of pulse oximeter equipment.
A functional tester is used to measure how accurately the fingertip pulse oximeter reproduces the specified calibration curve and the PR accuracy. The model of functional tester is the Index2 FLUKE simulator and the version is v3.00.

TECHNOLOGY

Display Type:

  • LED digitron display MX86001(AS-302-L) SpO2
  • Measurement range: 36% - 99%
  • Resolution: 1%
  • Precision: ±2% (70% - 99%). Less than 70% no definition

Pulse rate

  • Measurement range: 30bpm – 250 bpm
  • Resolution: 1 bpm
  • Precision: 1% or 1bpm choose the bigger one between the two on error.

Low perfusion error

  • SpO2 and pulse rate can be shown correctly when pulse-filling ration is 0.2%
  • Resistance to ambient light interference ability.
  • The deviation between the SpO, value measured in the condition of indoor day lighting or interior lighting and that of darkroom is less than ±1%.

Power supply

  • d.c 3.0V (2 x AAA alkaline batteries)

Working current

  • 20mA – 130mA

Battery life

  • Not less than 12 hours

Safety type

  • Interior battery: Type BF

Light sensor

  • Red light (wavelength 660nm – 666nm 7mW)
  • Infrared light (wavelength 890nm – 904nm 5.5mW)

Date update cycle

  • Not more than 12s

Degrees of protection

  • IP21

NOTE: The information about wavelength range can be especially useful to clinicians.

DIMENSIONS

Model Length Width Height Weight (incl batteries)
MX86001
(AS-302-L)
58mm 35mm 32mm 50g

Environment Requirements

Operations temperature: 5°C – 40°C
Ambient humidity: ≤80% at Atmospheric pressure: 86 106kPa
Storage, transportation condition
Operation temperature: (-20 – 55) °C
Ambient humidity: ≤93% (no condensation)
Atmosphere pressure: (50 -106) kPa

Problems Possible Cause Corrective Action
SpO 2 or pulse rate is incorrect Insert the finger incorrect position Measure again after reinserting your finger
SpO 2 or Pulse rate is erratic
  1. Finger might not be in the correct position
  2. Finger trembling or patient motion
  1. Reinsert the finger
  2. Patient must remain still
The oximeter does not turn on
  1. Batteries weak or batteries not installed
  2. Batteries incorrectly installed
  3. Damaged oximeter
  1. Replace the batteries
  1. Ensure the batteries are installed correctly
  2. Contact the manufacturer
The screen suddenly shuts down
  1. The oximeter is powered off automatically when no signal of the finger is detected for a long time
  2. Inadequate power of the batteries
  1. Normal
  2. Replace the batteries
The oximeter blackout The oximeter probe failure The device is faulty. Contact the manufacturer.
MX86001(AS-302-L) displays bar graph with no SpO2 and pulse rate The signal of SpO2 source is suspect or incorrect Use a different finger and measure again. The device is faulty. Contact the manufacturer

Symbol Definitions

Symbol Definition Symbol Definition
Type BF Equipment No SpO2 Alarm
Attention! Consult accompanying documents Use by date
SpO%2 Oxygen saturation CE MARK
PR BPM Pulse rate (BPM) Manufacturers information
IP 21 Drip proof Follow operating instructions
SN Serial No. Authorized representative in the European Community
Symbol for disposed waste electrical and electronic equipment (Follow local government ordinance and recycling instructions regarding disposal of batteries.

Equipment response time

As shown in the following figure.
Response time of normal average is 10.4s

Equipment response time - Graphic diagram

Manufactured by Medinox London
Solar House, 282 Chase Road,
London N14 6NZ.
www.medinox.co.uk

Australian Sponsor:
Medinox Australia (Pty) Ltd
Unit 15, 265 - 271 Pennant Hills Road,
Thornleigh, 2120 Tel: 1800724903
www.medinox.com.au

South African Sponsor:
Medinox South Africa (Pty) Ltd
2 Sunrock Close, Sunnyrock, Germiston,
Gauteng, 1401.
www.medinox.co.za

Documents / Resources

References

Download manual

Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.

Download MX Health MX86001 (AS-302-L) - Fingertip Pulse Oximeter Manual

Advertisement

Need help?

Need help?

Do you have a question about the MX86001 and is the answer not in the manual?

Questions and answers

Table of Contents