Summary of Contents for BEAMEX MC5-IS-FF

  • Page 1 User Guide Fieldbus Option for FOUNDATION Fieldbus...
  • Page 2 We reserve rights to make modifications to this manual without any further notice. For more detailed technical data about MC5-IS Intrinsically Safe Cali- brator, please refer to MC5-IS User Guide or contact the manufactur- © Copyright 2009 BEAMEX OY AB Ristisuonraitti 10 68600 Pietarsaari FINLAND +358 - 10 - 5505000...
  • Page 3 Trademarks ® QCAL is a registered trademark owned by Beamex Oy Ab. FOUNDATION™ Fieldbus is a registered trademark of the Fieldbus Foundation. ® PROFIBUS is a registered trademark of PROFIBUS International. ® HART is a registered trademark of the HART Communication Foundation.
  • Page 4: Table Of Contents

      Viewing Device Descriptions in MC5-IS ............. 14   Device Description Sender ................. 15   Fieldbus Instruments and Beamex's Calibration Database Software ....16   Connecting to a FOUNDATION Fieldbus Instrument ......17   Method 1: External Supply ................18  ...
  • Page 5 Fieldbus Communication Settings ............34   Appendix 1, FOUNDATION Fieldbus Units Supported by MC5-IS ..35   Appendix 2, Useful FOUNDATION Fieldbus Codes ....... 38   Notes ......................41  ...
  • Page 6: General

    User Guide – MC5-IS Fieldbus Option for FOUNDATION Fieldbus General FI5-IS Intrinsically Safe Fieldbus Interface allows you to ca- librate Intrinsically Safe FOUNDATION Fieldbus instruments. All you need is a fieldbus compatible MC5-IS and a FI5-IS suited for your fieldbus type. The MC5-IS/FI5-IS for FOUNDATION Fieldbus features listed below are presented in more detail further on in this manual.
  • Page 7: What Is Fieldbus

    User Guide – MC5-IS Fieldbus Option for FOUNDATION Fieldbus What is Fieldbus Fieldbus is an industrial digital communications network specif- ically designed for process automation use. It is meant to re- place the existing standard 4 … 20 mA analog signal. Fieldbus technology has been around since late 80's, but de- veloping an international standard took a long time.
  • Page 8: Batteries

    User Guide – MC5-IS Fieldbus Option for FOUNDATION Fieldbus Batteries The rechargeable battery inside FI5-IS provides supply power to the communication between MC5-IS and FI5-IS. With fresh, fully charged batteries, FI5-IS may be used for ap- prox. 10 hours without recharging. To save batteries, discon- nect the fieldbus cables when FI5-IS is not used.
  • Page 9: Disposal Of Waste Electrical And Electronic Equipment

    User Guide – MC5-IS Fieldbus Option for FOUNDATION Fieldbus Disposal of Waste Electrical and Electronic Equipment Beamex is an environmentally conscious company developing products with a view to ensure that they are easy to recycle and do not introduce hazardous materials into the environ- ment.
  • Page 10: Safety

    User Guide – MC5-IS Fieldbus Option for FOUNDATION Fieldbus Safety Symbols Used The following symbols concerning electrical safety are used in FI5-IS. Caution! See manual for further information Ex Information and Approvals Please read the separate leaflet shipped with FI5-IS entitled (in English) FI5-IS Safety Information.
  • Page 11  +50 °C.. The battery shall not be charged on a hazardous location. Use only Beamex charger type BC09-IS for charging. The charger shall not be used in a hazardous location. The surface of FI5-IS contains light metal, so the user shall be aware of possible impact or friction sparks when using this de- vices.
  • Page 12 User Guide – MC5-IS Fieldbus Option for FOUNDATION Fieldbus Input Parameters Fieldbus terminals Maximum values, 24 V DC 250 mA 1.2 W 0 0 Non-FISCO Maximum values with 17.5 V DC 250 mA 1.2 W 0 0 FISCO, IEC 60079-27 MC5-IS connection 14 V DC 60 mA...
  • Page 13: Notes And Warnings

    Avoid short circuiting the batteries. The short circuit cur- rent may cause burns to you, damage to the device or even fire. The batteries may be replaced by Beamex or a Beamex authorized service only. Rechargeable batteries may vent small amounts of gas during recharge.
  • Page 14 User Guide – MC5-IS Fieldbus Option for FOUNDATION Fieldbus Fieldbus Related Warnings When calibrating an instrument that is part of a control system (live segment), first make sure, the control loop the instrument is part of is set to manual. Also remember to check other dependencies between the instrument to be calibrated and other instruments.
  • Page 15: Service

    User Guide – MC5-IS Fieldbus Option for FOUNDATION Fieldbus Service FI5-IS may only be serviced by Beamex or a Beamex author- ized service. Contact: BEAMEX OY AB Ristisuonraitti 10 FIN-68600 Pietarsaari FINLAND E-mail: service@beamex.com Internet: http://www.beamex.com Opening of the casing of FI5-IS is not allowed due to ATEX safety reasons.
  • Page 16: Terminology

    In the user interfaces and manuals of fieldbus instruments and fieldbus control system software, the second phase is called "Calibration", "Adjustment" and/or "Trimming". So, to avoid confusion Beamex uses the term "Verifying Calibration" for the comparison phases in fieldbus environment. General...
  • Page 17: Mc5-Is, Fi5-Is And Foundation Fieldbus

    MC5-IS, FI5-IS and FOUNDATION Fieldbus A FOUNDATION Fieldbus segment always has a master de- vice called Link Active Scheduler (LAS). Beamex's MC5-IS and FI5-IS are together seen as a guest device (visitor). Con- necting FI5-IS only to the fieldbus does not "awaken" LAS, but once MC5-IS is communicating with FI5-IS, the LAS should in- dicate that a guest device is part of a fieldbus segment.
  • Page 18 User Guide – MC5-IS Fieldbus Option for FOUNDATION Fieldbus 2. Verifying Calibration, As Found Perform the Verifying Calibration using MC5-IS/FI5-IS for FOUNDATION Fieldbus as presented in chapters further on. Verifying Calibration of fieldbus instruments is basically done as the calibration of analog and HART® instruments. The mi- nor differences that exist are due to the instrument's fully digi- tal output.
  • Page 19: Device Descriptions

    Therefore Beamex publishes special device descriptions for MC5-IS and the instruments connected to it. Please contact Beamex if you need device descriptions for new instruments. For contact info, see the first pages of this manual. The following chapters describe how available device descrip- tions can be viewed in MC5-IS and also how to add new de- vice descriptions to MC5-IS.
  • Page 20: Device Description Sender

    User Guide – MC5-IS Fieldbus Option for FOUNDATION Fieldbus Device Description Sender Beamex offers a tool for sending device descriptions to MC5- IS. This can however be done in safe are only. Running the Device Description Sender software opens a win- dow with a list of device description data it contains.
  • Page 21: Fieldbus Instruments And Beamex's Calibration Database Software

    User Guide – MC5-IS Fieldbus Option for FOUNDATION Fieldbus Fieldbus Instruments and Beamex's Calibration Database Software CMX Calibration Management Software V2, revision 2.1 and later support fieldbus instruments. Older calibration software (QM6, QD3 etc.) do not support fieldbus instruments. General...
  • Page 22: Connecting To A Foundation Fieldbus Instrument

    Beamex recommends that instruments should first be removed from a live segment before attempting to communicate with it using MC5-IS and FI5-IS. For advice on how to remove an in- strument from a live segment, see chapter Calibration Proce- dure of a Fieldbus Instrument, General on page 12.
  • Page 23: Method 1: External Supply

    User Guide – MC5-IS Fieldbus Option for FOUNDATION Fieldbus Method 1: External Supply This is a method where the instrument is not part of the process or a factory fieldbus segment. The instrument com- municates with MC5-IS as a standalone instrument. A sepa- rate power supply provides the loop power for the fieldbus "stub".
  • Page 24: Method 2: Connecting Mc5-Is To A Fieldbus Segment

    Please read chapter Notes and Warnings on page 8 before attempting to connect using this method. Beamex cannot be held responsible for any damages caused by connect- ing MC5-IS to a live factory fieldbus segment.
  • Page 25: Fieldbus And Mc5-Is' Basic Mode

    User Guide – MC5-IS Fieldbus Option for FOUNDATION Fieldbus Fieldbus and MC5-IS' Basic Mode Connecting to a FOUNDATION Fieldbus Instrument To initiate communica- 2 2 . 1 2 . 2 0 0 6 1 2 : 4 6 tion with a fieldbus in- R T D - t e m p e r a t u r e Q u a n t i t y E T : R T D T e m p .
  • Page 26 If MC5-IS does not have required Device Descriptions, full communication cannot be initiated. In this case a window opens telling what kind of Device Description is needed. Contact Beamex to receive instructions on how to correct the problem. Fieldbus and MC5-IS' Basic Mode...
  • Page 27 User Guide – MC5-IS Fieldbus Option for FOUNDATION Fieldbus After required instrument 0 2 . 0 4 . 2 0 0 7 1 3 : 4 2 data is read, MC5-IS re- R T D - t e m p e r a t u r e D i s c o n n e c t E T : R T D T e m p .
  • Page 28: Fieldbus Window Setup Menu

    User Guide – MC5-IS Fieldbus Option for FOUNDATION Fieldbus Fieldbus Window Setup Menu A window reserved for fieldbus communication has its own menu. All menu items are presented in the following subchap- ters. Disconnect Device This option disconnects the fieldbus instrument from MC5- IS/FI5-IS.
  • Page 29: Transducer Block Information

    User Guide – MC5-IS Fieldbus Option for FOUNDATION Fieldbus Transducer Block Information To open the Transducer 1 4 . 0 9 . 2 0 0 6 1 3 : 2 2 Block Information win- T R A N S D U C E R B L O C K dow, press S T _ R E V T A G _ D E S C...
  • Page 30: Mode Block

    User Guide – MC5-IS Fieldbus Option for FOUNDATION Fieldbus Mode Block Mode Block is an important Transducer Block parameter while calibrating instruments. That's why it is directly available via MC5-IS' menu. Use this to set the instrument in either OOS - Out Of Service or Auto - Automatic mode.
  • Page 31: Add A Fieldbus Device To Database

    User Guide – MC5-IS Fieldbus Option for FOUNDATION Fieldbus Add a Fieldbus Device to Database This is a semi-automatic 0 2 . 0 4 . 2 0 0 7 1 3 : 5 5 tool for adding a fieldbus R T D - t e m p e r a t u r e D i s c o n n e c t E T : R T D T e m p .
  • Page 32 User Guide – MC5-IS Fieldbus Option for FOUNDATION Fieldbus When the instrument is 1 5 . 0 9 . 2 0 0 6 0 8 : 4 2 added to MC5-IS, a dis- I N S T R U M E N T T T 1 2 2 - A P play like the adjacent pic- 1 1 2 - T T - 0 0 3 .
  • Page 33: Unit

    User Guide – MC5-IS Fieldbus Option for FOUNDATION Fieldbus Unit To change the unit used in MC5-IS' display, press Menu Window 2 Setup , if needed (opens the second page of the menu) and Display Unit Notes. In Basic Mode, the unit used in MC5-IS' display need not be the same as in the instrument itself.
  • Page 34: Performing A Verifying Calibration

    User Guide – MC5-IS Fieldbus Option for FOUNDATION Fieldbus Performing a Verifying Calibration Selecting the Instrument To be able to calibrate a fieldbus instrument, it has to be added to MC5-IS' instrument database. Adding the instruments can be done using the utility described in chapter Add a Fieldbus Device to Database on page 26.
  • Page 35: Doing The Verifying Calibration

    User Guide – MC5-IS Fieldbus Option for FOUNDATION Fieldbus Doing the Verifying Calibration Doing a Verifying Cali- 2 2 . 1 2 . 2 0 0 6 1 3 : 3 0 bration for a fieldbus in- I n p u t R T D T e m p e r a t u r e [ E T : S i m u l ] strument does not differ 7 5 .
  • Page 36: Calibrating (Trimming, Adjusting) A Fieldbus Instrument

    User Guide – MC5-IS Fieldbus Option for FOUNDATION Fieldbus Calibrating (Trimming, Adjusting) a Fieldbus Instrument To start Calibrating (Trimming, Adjusting) a Fieldbus Instru- ment in Calibration Mode, press Menu Start Fieldbus Adjustment while viewing the calibration windows. The opened 2 2 . 1 2 . 2 0 0 6 1 3 : 3 0 menu has I n p u t...
  • Page 37 User Guide – MC5-IS Fieldbus Option for FOUNDATION Fieldbus 2. If some of the options in the Fieldbus Instrument Ad- justment menu's items are greyed (refer to the picture on the previous page), the selected output for that par- ticular instrument cannot be trimmed. Change the output Select Output using menu option Trim Method...
  • Page 38 User Guide – MC5-IS Fieldbus Option for FOUNDATION Fieldbus Trim Upper 100% Range 7. Then select the option from the Fieldbus Instrument Adjustment menu. Other- wise, the trimming procedure is similar to phases 5 and In this case, the corrected value is sent to Transducer Block parameter, CAL_POINT_HI (or similar, depending on the instrument at hand).
  • Page 39: Fieldbus Communication Settings

    User Guide – MC5-IS Fieldbus Option for FOUNDATION Fieldbus Fieldbus Communication Settings You may view/edit field- 0 2 . 0 4 . 2 0 0 7 1 3 : 2 0 bus settings in Basic R T D - t e m p e r a t u r e S e t t i n g s E T : R T D T e m p .
  • Page 40: Appendix 1, Foundation Fieldbus Units Supported By Mc5-Is

    User Guide – MC5-IS Fieldbus Option for FOUNDATION Fieldbus Appendix 1, FOUNDATION Fieldbus Units Supported by MC5-IS Note. Certain instruments may not support all codes listed here. When in doubt, please consult the instrument's manual. Temperature Units Unit Code in Fieldbus Displayed Unit in Displayed Unit in Instruments...
  • Page 41 User Guide – MC5-IS Fieldbus Option for FOUNDATION Fieldbus Pressure units, continued Unit Code in Fieldbus Displayed Unit in Displayed Unit in Instruments Fieldbus Instruments 1143 psig psig 1144 gf/cm gf/cm kgf/cm kgf/cm 1145 1146 O(4°C) O @ 4°C 1147 O(68°F) O @ 68°F 1148...
  • Page 42 User Guide – MC5-IS Fieldbus Option for FOUNDATION Fieldbus Pressure units, continued Unit Code in Fieldbus Displayed Unit in Displayed Unit in Instruments Fieldbus Instruments inHga 1577 inHg 1578 inHgg inHg 1579 inHga(0°C) inHg @ 0°C 1580 inHgg(0°C) inHg @ 0°C 1581 mmHga mmHg...
  • Page 43: Appendix 2, Useful Foundation Fieldbus Codes

    User Guide – MC5-IS Fieldbus Option for FOUNDATION Fieldbus Appendix 2, Useful FOUNDATION Fieldbus Codes Sensor Types (Variable SENSOR_TYPE) Sensor Type, Code Description Flow sensor unknown Coriolis (Gyroscopic) Electromagnetic   Nuclear magnetic resonance Positive displacement Refraction Taggin Ultrasonic (Doppler) Ultrasonic (time of travel) Vortex Target...
  • Page 44 User Guide – MC5-IS Fieldbus Option for FOUNDATION Fieldbus Sensor Types, continued Sensor Type, Code Description T/C Type B (IEC 584-1, and NIST 175) T/C Type C (NIST 175) T/C Type E (IEC 584-1, and NIST 175) T/C Type J (IEC 584-1, and NIST 175) T/C Type K (IEC 584-1, and NIST 175) T/C Type N (IEC 584-1, and NIST 175) T/C Type R (IEC 584-1, and NIST 175)
  • Page 45 User Guide – MC5-IS Fieldbus Option for FOUNDATION Fieldbus Calibration Methods (Variable SENSOR_CAL_METHOD) Calibration Method, Code Description volumetric static weigh dynamic weigh factory trim standard calibration user trim standard calibration factory trim special calibration user trim special calibration 240-254 Reserved (Manufacturer Specific) other Appendix 2, Useful FOUNDATION Fieldbus Codes...
  • Page 46: Notes

    © Copyright 2009 Beamex Inc BEAMEX OY AB 2152 Northwest Parkway Ristisuonraitti 10 Suite A FIN-68600 PIETARSAARI Marietta, GA 30067 FINLAND U.S.A. Phone +358 - 10 - 5505000 Phone 800 888-9892, +358 - 10 - 5505404 +1-770-951-1927 E-mail sales@beamex.com +1-770-951-1928 service@beamex.com...