PcCR 1417 ACL4 System
Automatic Cassette Loading
Version
Installation Manual
Catalog number: AT095047-00
January, 2003
Ver. 01

Advertisement

Table of Contents
loading

Summary of Contents for OREX PcCR 1417 ACL4 System

  • Page 1 PcCR 1417 ACL4 System Automatic Cassette Loading Version Installation Manual Catalog number: AT095047-00 January, 2003 Ver. 01...
  • Page 2 The company’s mission is to become a leading provider of compact personal CR systems. Following the successful penetration of the global dental market, the company changed its name from Digident to OREX and expanded its markets and product lines to serve the dental, medical, military, industrial and veterinary fields.
  • Page 3 Safety Summary LIFTING HAZARD The PcCR 1417 scanner weighs 40 Kg (88lb). • Do not try to lift the scanner by yourself. • Always seek assistance from another person. • Lifting equipment that is too heavy may result in serious injury to personnel and/or damage to equipment.
  • Page 4: Table Of Contents

    Table of Contents Terms and Conventions..............1-1 Packaging..................2-1 Site Preparation................3-1 General ....................3-1 Computer Requirements..............3-1 Space Requirements ................3-1 Electrical Requirements ..............3-2 Environmental Requirements............. 3-3 3.5.1 Temperature ................3-3 3.5.2 Light ..................3-3 Installation ...................4-1 General ....................4-1 System Configuration On-Site ............4-2 Equipment Required For Complete Installation.........
  • Page 5 5.2.1 Main Screen................5-1 5.2.2 Setup Anatomical ..............5-3 5.2.3 Setup Image................5-7 5.2.4 Setup Configuration ............... 5-9 5.2.5 Setup Operation Update ............5-13 5.2.6 Setup Diagnostics..............5-14 5.2.7 Setup Automatic Tests ............5-16 5.2.8 Setup Advance..............5-18 5.2.9 Setup About................5-20 5.2.10 Image Diagnostics ..............
  • Page 6 List of Figures Figure 1: Closed Box.......................2-1 Figure 2: Open Box .........................2-1 Figure 3: PcCr 1417 Dimensions ....................3-2 Figure 4: System Configuration ....................4-2 Figure 5: Dongle Parallel Plug ....................4-3 Figure 6: CVLicense Window Before GOP Licensing............4-15 Figure 7: Setting the GOP License Key ................4-16 Figure 8: CVLicense Window After GOP Licensing ............4-17 Figure 9: Scanner Interface, Main Screen................5-1 Figure 10: Setup Menu, Anatomical tab .................5-3...
  • Page 7: Terms And Conventions

    1. Terms and Conventions The following abbreviations are used throughout this manual: Term Meaning Field Service Engineer Photo-Stimulate Phosphor Universal Serial Bus Terms and Conventions...
  • Page 9: Packaging

    The PcCR 1417 is shipped in a custom designed box that protects the scanner. If the PcCR 1417 needs to be returned to Orex or one of its dealers, the scanner must be repacked in its original box, and must include the CD and the USB cable.
  • Page 11: Site Preparation

    3. Site Preparation 3.1 General Inform the client about the site preparation requirements prior to installing the PcCR 1417. Without these requirements being met, you cannot complete the installation, and the client’s ability to use the PcCR 1417 properly can not be guaranteed.
  • Page 12: Electrical Requirements

    50 cm 80 cm 100 cm Figure 3: PcCr 1417 Dimensions 3.4 Electrical Requirements Voltage: 90 V–250 V AC Frequency: 50Hz–60Hz Fuses: Two fuses are located next to the On/Off switch at the back of the scanner. Each fuse protects one of the power supplies.
  • Page 13: Environmental Requirements

    3.5 Environmental Requirements 3.5.1 Temperature Maintain the room temperature between 18ºC and 30ºC, normal lab conditions; relative humidity 80%, non-condensing. 3.5.2 Light The PcCR 1417 is sealed and protected from light penetration. However, it is recommended to keep the PcCR 1417 in standard room light, away from the window.
  • Page 15: Installation

    The FSE who installs the PcCR 1417 shall follow the steps described herein. Upon completion of the installation, fill out the Installation Report (Appendix B). 4.1 General On-site installation of the PcCR 1417 shall be carried out by FSEs authorized by Orex, and will follow the instructions below: who do the following: 1.
  • Page 16: System Configuration On-Site

    PC – Make sure that the PC you are using meets the requirements described in Section 3.2, Computer Requirements Scanner interface – Software supplied by Orex with every system GOP – Enhancements for the scanner interface software for the purpose of image...
  • Page 17: Equipment Required For Complete Installation

    4.3 Equipment Required For Complete Installation 1. PC (refer to “computer requirements” for details) 2. Cables: Power cable (not included with the PcCR 1417), USB cable (included with the PcCR 1417) 3. “Scanner interface” CD 4. “Or-Acquire” CD 5. Dongle (parallel plug) for licensing GOP filters and Or-Acquire Figure 5: Dongle Parallel Plug Installation...
  • Page 18: Installation Process

    4.4 Installation process The installation process includes the following: 4.4.1 Connecting the cables 4.4.2 Installing drivers for PcCR 1417 4.4.3 Installing “scanner interface” 4.4.4 Installing Or-Acquire 4.4.6 Installing the Dongle (license for GOP) 4.4.4 Licensing the GOP filters 4.4.1 Connecting the cables 1.
  • Page 19 3. Choose “Search for a suitable…”; then click “Next” 4. Leave only the “specify a location” check-box selected; then click “Next” Installation...
  • Page 20 5. Insert the “scanner interface” CD in to the CD- Rom drive. Click the “Browse” button. 6. A “Locate file” window appears. Browse to the CD-Rom. 7. Double click the “Windows2000” folder. Installation...
  • Page 21 8. Mark the “orex02.inf” file; then click “Open” 9. The Installation Wizard finds the driver for the PcCR 1417. Click “Next”. Installation...
  • Page 22 10. Click “Yes” for confirming the driver file 11. The installation wizard completes the installation of the driver. Installation...
  • Page 23: Installing The "Scanner Interface

    4.4.3 Installing the “Scanner Interface” 1. Insert the “Scanner Interface” CD into the CD- Rom drive. Double-click on “My Computer” and choose the CD-Rom drive. The content of the CD will appear on a new window: 2. Double-click the “setup.exe” file. 3.
  • Page 24 4. The software is now being installed. 5. At the end of the installation process, you are prompted to copy the configuration files. Click “yes”. 6. The “Scanner Interface” software is now installed. Click “Finish” to restart the computer. 4-10 Installation...
  • Page 25: Or-Acquire Installation

    4.4.4 Or-Acquire Installation The Or-Acquire is supplied on a different CD than the one with the “Scanner Interface”. For installation of the Or-Acquire, refer to the detailed explanations in the Installation Manual located on the CD with Or-Acquire. Pay attention: 1.
  • Page 26 4. Click “Next” to continue. 4-12 Installation...
  • Page 27 Choose “I accept the terms…”, and click “Next Choose “Complete”, and click “Next” 7. Click “Install”. Installation 4-13...
  • Page 28 8. Click “Finish”. 4-14 Installation...
  • Page 29: Licensing The Gop Filters

    1. Insert the “Scanner Interface” CD into the CD- Rom drive. 2. Double click the “CVLicense.exe” file. Figure 6: CVLicense Window Before GOP Licensing 3. A new window appears. Double click “CVIE- Orex-Soft” (3rd line). Installation 4-15...
  • Page 30: Figure 7: Setting The Gop License Key

    4. The licensing window appears. Enter the code supplied by Orex into the “Key” dialog box. If you have not received the code, please contact your distributor. The code is unique, and depends on the SN of the Dongle, which appears on the dongle as a 4-digit number at the bottom line.
  • Page 31: Figure 8: Cvlicense Window After Gop Licensing

    Figure 8: CVLicense Window After GOP Licensing 6. The previous screen appears again. This time the “CVIE-Orex-Soft” on the 3rd line is valid. Click “OK”. Installation 4-17...
  • Page 32: Installation Report

    4.5 Installation Report In order for the warranty of the PcCR 1417 to take effect, and in order to verify correct installation by Orex, you are required to fill out the installation form and fax/e-mail to Orex service department. The installation form is located in Appendix C...
  • Page 33: Scanner Interface Software

    5. Scanner Interface Software 5.1 General The Scanner Interface Software is a module that enables the FSE to access to the PcCR 1417 advanced setup. Following is a list of screens, and their functionality: 5.2 Interface Description 5.2.1 Main Screen This screen is used as the interface between the end user and the scanner.
  • Page 34 For entering the Setup interface. The login screen is displayed: • User password – None • Technician password – 9591 • Orex Technician password – Contact Orex Service Dept. 7. Exit For exiting the PcCR 1417 interface. Scanner Interface Software...
  • Page 35: Setup Anatomical

    5.2.2 Setup Anatomical Note: Each setup relates to a specific part, to be scanned from a plate that originated from a specific x-ray machine. Figure 10: Setup Menu, Anatomical tab Title Parameter Description Unit Set The PcCR 1417 can support up to 4 different x-ray units.
  • Page 36 Title Parameter Description Organ List Organ List Select the required organ from the drop- down list. The following organs are available: • Skull • Upper Extremities • Chest • Abdomen • Spine • Lower Extremities • Abdomen Sub Organ List Select the required sub-organ from the available list;...
  • Page 37 After the Offset calibration, refer to the User Manual. Do not change this value! Orex Filters Setup If you are using Orex filters, check the On checkbox; then enter the relevant values to display the best image. This option is usually not used.
  • Page 38 For using settings pre-defined by an external software. Show Histogram For showing the image histogram at the end of the scanning process. Use Orex Window For performing automatic window Leveling leveling process on the image. It is recommended not to use this option.
  • Page 39: Setup Image

    5.2.3 Setup Image Figure 11: Setup Menu, Image tab Title Parameter Description Data Correction Method Data Correction For changing the data correction method. Method The recommended method is Linearization. Browse For loading a pre-defined set of values; located by browsing. U test Not used.
  • Page 40 Title Parameter Description Image Format For setting the image file format. Use ONLY Raw Data only (16 bit) Demo Scan For using an available image file for scanning, if no plate (cassette) is available. Rotate canvas Horizontal flip For flipping the image. Vertical flip Small Plate Resolution...
  • Page 41: Setup Configuration

    5.2.4 Setup Configuration Figure 12: Setup Menu, Configuration tab Title Parameter Description Local Scanner Settings Configuration file Opens the configuration file with the Notepad. Origin X X-axis home, where the scanner begins to scan the plate. Origin Y Y-axis home, where the plate begins its journey inside the scanner.
  • Page 42 The default operating mode of the Scan scanner. After scanning, on its way back into the cassette, the plate is erased. Auto For Orex internal use only. Load/Unload Auto Scan For automatically starting a scan when you insert a cassette into the scanner.
  • Page 43 ACL4 nominal Must be chosen for nominal work. ACL4 test For Orex internal use only. Origin Settings For fine-tuning the home position. After a scan is acquired, you can center the image, by either clicking...
  • Page 44 Title Parameter Description 10 Inch Choose when setting Origin for 10-inch wide images. 14 Inch Choose when setting Origin for 14-inch wide images. For saving the parameters. This Save HAS to be done before selecting another tab. 5-12 Scanner Interface Software...
  • Page 45: Setup Operation Update

    5.2.5 Setup Operation Update Figure 13: Setup Menu, Operation Update tab Title Parameter Description Update Main Firmware Software Update Select the type of scanner (APL, ACL, ACL2 or ACL4); then click the Software update button to update the software to the scanner.
  • Page 46: Setup Diagnostics

    5.2.6 Setup Diagnostics Figure 14: Setup Menu, Diagnostics tab Title Parameter Description Optical Erase Lamp Turn on/off the Erase lamp Lamp1 Indicates that lamp1 is on. Lamp2 Indicates that lamp2 is on. Laser Turns on/off the Laser unit Turns on the Rotational motor Rotational Motor Speed Indicates the speed of the rotational...
  • Page 47 Title Parameter Description Plate Control Sensors Sensors Tray – Cassette is docked Cassette hold – Cassette is locked Loader – Loader has reached plate W0 – Plate is approaching rollers; has to slow down Roller – Feeding rollers Z0-Z4 – Positioning sensors Moves the plate In or Out of the scanner.
  • Page 48: Setup Automatic Tests

    5.2.7 Setup Automatic Tests This menu is used primarily, at the end of the Orex production line, for burn-in tests. Figure 15: Setup Menu, Automatic Tests tab Title Parameter Description Scanner Test Test Counter Displays the number of tests that have been run.
  • Page 49 Title Parameter Description Automatic Automatically Loads and unloads the Load/Unload plate. This test is usually run with the Automatic Scan test. Number of tests Sets the number of tests to run. Scanner Interface Software 5-17...
  • Page 50: Setup Advance

    Resets the number of scans to “0”. RAIS 2 Dual Scanner mode Scanner index Interface Settings Anatomical Check for the body image on the main Interface screen. Uncheck for test-labeled buttons. The labels can be renamed by using the Orex Editor. 5-18 Scanner Interface Software...
  • Page 51 C. Memory backup Backup of data on serial EPROM The buttons perform the following functions Save Copy B Copy A Copy B Copy C Orex recommends that these function shall be performed only by Orex technicians. Scanner Interface Software 5-19...
  • Page 52: Setup About

    5.2.9 Setup About Figure 17: Setup Menu, About tab Title Parameter Description Software Version 1. Motion Currently used version of the Motion board 1.1 USB (FPGA) Currently used version of the USB (FPGA). 2. Firmware (boot) Currently used version of the Firmware (boot) 3.
  • Page 53 Title Parameter Description Scanner ID Serial number DV number Scan counter Scanner Interface Software 5-21...
  • Page 54: Image Diagnostics

    5.2.10 Image Diagnostics After scanning the desired image, it is displayed on the mini viewer. This is used for final calibration of the filters assigned to the displayed part (sub-organ). As opposed to the Setup screens, the results can be seen immediately. Figure 18: Image Diagnostic screen (Mini Viewer) Title Parameter...
  • Page 55 Title Parameter Description Data Correction For changing the data correction method. method The recommended method is Linearization. Data minimum For changing the Data Minimum. Brightness slider Contrast slider Process For viewing the result of the new setup For resetting the image Reset Histogram To view the histogram of the displayed...
  • Page 57: Troubleshooting

    2. If the erase lamps are OK, increase the erasing factor by 5 from the Scanner Interface – Setup menu Configuration White lines appear on Contact Orex service department image. 10. Horizontal lines appear Contact Orex service department on image 11.
  • Page 59: Appendix A: Installation Prerequisites

    Single Media Archive required? Yes / No When using the PcCr 1417 with a main PACS on site, the images are Dicom sent to the PACS. Please verify with Orex the compatibility of the PACS to Orex software. Appendix A: Installation Prerequisites...
  • Page 61: Appendix B: Installation Report

    8. Appendix B: Installation Report INSTALLATION REPORT Reporting distributor: Company name: Technician installing the PcCr 1417: Name: Tee. No.: e-mail: Site information: Name: Address: Tel. No.: Additional information: Installation pre-requisites: Date: SN of system installed Computer requirements met ? Space requirements met ? Electrical requirements met ? Environmental requirements met ? * See Installation pre-requisites, Appendix B...