PcCR 1417 ACL4 System
Automatic Cassette Loading
Version
Installation Manual
October, 2003
Ver. 02

Advertisement

Table of Contents
loading

Summary of Contents for OREX PcCR 1417 ACL4

  • Page 1 PcCR 1417 ACL4 System Automatic Cassette Loading Version Installation Manual October, 2003 Ver. 02...
  • Page 2 The company’s mission is to become a leading provider of compact personal CR systems. Following the successful penetration of the global dental market, the company changed its name from Digident to OREX and expanded its markets and product lines to serve the dental, medical, military, industrial and veterinary fields.
  • Page 3 Safety Summary LIFTING HAZARD The PcCR 1417 scanner weighs 40 Kg (88lb). • Do not try to lift the scanner by yourself. • Always seek assistance from another person. • Lifting equipment that is too heavy may result in serious injury to personnel and/or damage to equipment.
  • Page 4: Table Of Contents

    Table of Contents Terms and Conventions..............1-1 Packaging..................2-1 Site Preparation................3-1 General ....................3-1 Computer Requirements..............3-1 Space Requirements ................3-1 Electrical Requirements ..............3-2 Environmental Requirements............. 3-3 3.5.1 Temperature ................ 3-3 3.5.2 Light ..................3-3 Installation ...................4-1 General ....................4-1 System Configuration On-Site ............4-2 Equipment Required For Complete Installation.........
  • Page 5 List of Figures Figure 1: Closed Box.......................2-1 Figure 2: Open Box .........................2-1 Figure 3: PcCr 1417 Dimensions ....................3-2 Figure 4: System Configuration ....................4-2 Figure 5: Dongle Parallel Plug ....................4-3 Figure 6: CVLicense Window Before GOP Licensing............4-19 Figure 7: Setting the GOP License Key ................4-20 List of Figures...
  • Page 7: Terms And Conventions

    1. Terms and Conventions The following abbreviations are used throughout this manual: Term Meaning Field Service Engineer Photo-Stimulate Phosphor Universal Serial Bus Terms and Conventions...
  • Page 9: Packaging

    The PcCR 1417 is shipped in a custom designed box that protects the scanner. If the PcCR 1417 needs to be returned to Orex or one of its dealers, the scanner must be repacked in its original box, and must include the CD and the USB cable.
  • Page 11: Site Preparation

    3. Site Preparation 3.1 General Inform the client about the site preparation requirements prior to installing the PcCR 1417. Without these requirements being met, you cannot complete the installation, and the client’s ability to use the PcCR 1417 properly can not be guaranteed.
  • Page 12: Electrical Requirements

    50 cm 80 cm 100 cm Figure 3: PcCr 1417 Dimensions 3.4 Electrical Requirements Voltage: 90 V–250 V AC, 200VA Frequency: 50Hz–60Hz Fuses: Two fuses are located next to the On/Off switch at the back of the scanner. Each fuse protects one of the power supplies.
  • Page 13: Environmental Requirements

    3.5 Environmental Requirements 3.5.1 Temperature Maintain the room temperature between 18ºC and 30ºC, normal lab conditions; relative humidity 80%, non-condensing. 3.5.2 Light The PcCR 1417 is sealed and protected from light penetration. However, it is recommended to keep the PcCR 1417 in standard room light, away from the window.
  • Page 15: Installation

    Installation Report (Appendix B). 4.1 General On-site installation of the PcCR 1417 shall be carried out by FSEs authorized by Orex, and will follow the instructions below: who do the following: 1. Verify the integrity and check the shock indicators on the package.
  • Page 16: System Configuration On-Site

    PC – Make sure that the PC you are using meets the requirements described in Section 3.2, Computer Requirements Scanner interface – Software supplied by Orex with every system GOP – Enhancements for the scanner interface software for the purpose of image...
  • Page 17: Equipment Required For Complete Installation

    4.3 Equipment Required For Complete Installation 1. PC (refer to “computer requirements” for details) 2. Cables: Power cable (not included with the PcCR 1417), USB cable (included with the PcCR 1417) 3. “Scanner interface” CD 4. “Onyx” CD 5. Dongle (parallel plug) for licensing GOP filters . 6.
  • Page 18: Installation Process

    4.4 Installation process The installation process includes the following: 4.4.1 Connecting the cables 4.4.2 Installing drivers for PcCR 1417 4.4.3 Installing “scanner interface” 4.4.4 Installing “Onyx” . 4.4.6 Installing the Dongle (license for GOP) 4.4.4 Licensing the GOP filters 4.4.1 Connecting the cables 1.
  • Page 19 3. Choose “Search for a suitable…”; then click “Next” 4. Leave only the “specify a location” check-box selected; then click “Next” Installation...
  • Page 20 5. Insert the “scanner interface” CD in to the CD- Rom drive. Click the “Browse” button. Chose the CD- Drive location 6. A “Locate file” window appears. Browse to the CD-Rom. 7. Double click the “Windows2000” folder. Installation...
  • Page 21 8. Mark the “orex02.inf” file; then click “Open” 9. The Installation Wizard finds the driver for the PcCR 1417. Click “Next”. Installation...
  • Page 22 10. Click “Yes” for confirming the driver file 11. The installation wizard completes the installation of the driver. Installation...
  • Page 23: Installing The "Scanner Interface" (And Service Pack Number Xxx If Needed)

    4.4.3 Installing the “Scanner Interface” (and service pack number xxx if needed). 1. Insert the “Scanner Interface” CD into the CD- Rom drive. Double-click on “My Computer” and choose the CD-Rom drive. The content of the CD will appear on a new window: 2.
  • Page 24 3. The installation of “Scanner Interface” begins. Click “Next”. 4-10 Installation...
  • Page 25: Installing The "Scanner Interface" Service Pack (Example For Service Pack 1)

    4. The “Scanner Interface” software is now installed. Click “Finish” to restart the computer. Installing the “Scanner Interface” Service pack (Example for service pack 1) Each service pack will include the previous service pack. (Example – service pack 4 will include the previous SP1-3 . Install only The scanner interface with the last SP ) .
  • Page 26 2. Double-click the “setup.exe” file. 3. The installation of “Scanner Interface” SP1 begins. Click “Next”. 4-12 Installation...
  • Page 27 4. The “Scanner Interface”SP1 software is now installed. Click “Finish” to restart the computer. Installation 4-13...
  • Page 28: Onyx" Installation

    4.4.4 “Onyx” Installation The “Onyx” is supplied on a different CD than the one with the “Scanner Interface”. For installation of the “Onyx”, refer to the detailed explanations in the Viztek- Onyx-Rad User Manual. Pay attention: 1. The “Onyx” needs a Dongle in order to work. If the Dongle is not installed, or is not connected to the parallel port, the software cannot run.
  • Page 29 2. Insert the “Scanner Interface” CD into the CD- Rom drive. Browse to the CD-Rom drive on the windows explorer, and double click the “Dongle Driver” folder. 3. Double click the “RainbowSSD5.39.2.exe” file. Installation 4-15...
  • Page 30 4. Click “Next” to continue. 5.Choose “I accept the terms…”, and click “Next 4-16 Installation...
  • Page 31 6. Choose “Complete”, and click “Next” 7. Click “Install”. Installation 4-17...
  • Page 32 8. Click “Finish”. 4-18 Installation...
  • Page 33: Licensing The Gop Filters

    4.4.6 Licensing the GOP Filters This section can be applied ONLY after installing the Dongle as described on the previous section. 1. Insert the “Scanner Interface” CD into the CD- Rom drive. 2. Double click the “CVLicense.exe” file. Figure 6: CVLicense Window Before GOP Licensing Installation 4-19...
  • Page 34: Figure 7: Setting The Gop License Key

    3. A new window appears. Double click “CVIE- Orex-Soft” (5rd line). 4. The licensing window appears. Enter the code supplied by Orex into the “Key” dialog box indicate on the Dongle . (long Alpha Numeric no. ) If you have not received the code, please contact your distributor.
  • Page 35 6. The previous screen appears again. This time the “CVIE-Orex-Soft” Status on the 5rd line is valid. Click “OK”. Installation 4-21...
  • Page 36 4.5 Installation Report In order for the warranty of the PcCR 1417 to take effect, and in order to verify correct installation by Orex, you are required to fill out the installation form and fax/e-mail to Orex service department. The installation form is located in Appendix C...
  • Page 37: Troubleshooting

    2. If the erase lamps are OK, increase the erasing factor by 5 from the Scanner Interface – Setup menu Configuration White lines appear on Contact Orex service department image. 10. Horizontal lines appear Contact Orex service department on image 11.
  • Page 38 Place the system away from intense light Preparation source (window). Error List IMPORTANT Inform OREX directly or through your vendor about any error messages that may appear. Do not attempt to repair the malfunction by yourself! Error Message Cause/Description Update software please Memory of Operation Firmware is empty.
  • Page 39 Single Media Archive required? Yes / No When using the PcCr 1417 with a main PACS on site, the images are Dicom sent to the PACS. Please verify with Orex the compatibility of the PACS to Orex software. Appendix A: Installation Prerequisites 6-25...
  • Page 41: Appendix B: Installation Report

    Appendix B: Installation Report INSTALLATION REPORT Reporting distributor: Company name: Technician installing the PcCr 1417: Name: Tel. No.: e-mail: Site information: Name: Address: Tel. No.: Additional information: Installation pre-requisites: Date: SN of system installed Computer requirements met ? Space requirements met ? Electrical requirements met ? Environmental requirements met ? * See Installation pre-requisites, Appendix B...