Mostec M3020R Operating Manual

Mostec M3020R Operating Manual

Ph/mv-controller

Advertisement

Quick Links

Operating Manual
pH/mV-Controller
Type M3020R

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the M3020R and is the answer not in the manual?

Questions and answers

Summary of Contents for Mostec M3020R

  • Page 1 Operating Manual pH/mV-Controller Type M3020R...
  • Page 2 Operating Manual Warranty Mostec warrants this product to be free of manufacturing defects for a 2-year period after the original date of purchase. Within this period, defective products will be repaired free of charge provided that the defect occurred during normal operation. This warranty does not cover damage to the product resulting from ordinary usage such as front panel scratches, broken control elements and corrosion, etc.
  • Page 3: Table Of Contents

    N. Controller with two identical set points, inactive zone = 10% p-band [W] ......15 O. Controller mode – Adapting the controller to a titration curve ..........16 P. Factory settings ........................17 Q. Mechanical Dimensions......................18 Side view:..........................18 Rear view:..........................19 R. Technical Data........................20 V1.50 © 2015 MOSTEC AG, CH-4410 Liestal...
  • Page 4: Front Elements And Keyboard

    2............6,8,14 F1 / F2 no function on home screen access to calibration menu pH7 and mV/pH....6,13 relay operation manual/auto........14 setup menu, enter............5 main menu, exit............6 increase values............5 to 14 decrease values.............5 to 14 V1.50 © 2015 MOSTEC AG, CH-4410 Liestal...
  • Page 5: Setup Menu

    + and – sets the temperature to +/- 0.1°C. press ENT 10x USB on/off Turn on and eject the USB data logger. 0.00...14.00pH or -1000...1000mV press ENT 11x Alarm 1 press ENT 12x Alarm 2 0.00...14.00pH or -1000...1000mV V1.50 © 2015 MOSTEC AG, CH-4410 Liestal...
  • Page 6: Customer Menu

    SP1-SP2: hysteresis Pump monitoring Time setup See section H USB settings Date setup Data log period Remove thumb drive Keyboard lock/unlock See section I Language settings Device settings Custom label Factory setting Device info V1.50 © 2015 MOSTEC AG, CH-4410 Liestal...
  • Page 7: Measurement Settings

    + and – between 0.0...120.0°C, confirm the selection with ENT Sets stabilization for measured Change with + and – between Digital filter 0...100 digit value 1 digit = 1mV or 0.01pH, confirm the selection with V1.50 © 2015 MOSTEC AG, CH-4410 Liestal...
  • Page 8: Controller Settings

    Should be 0.00s 17/17 Change characters with + and -, switch position with Relay 1-2 label F3 and F4, confirm label with ENT. Shows a custom label for the relay 1 and 2 V1.50 © 2015 MOSTEC AG, CH-4410 Liestal...
  • Page 9: Relay Settings

    + and –, resolution in range from 5 to 200 is +/- 1mV or 0.01pH 14/14 Relay 3-4 reference Switch with + between PT100 and pH/mV as reference for relay 3+4, confirm the selection with ENT V1.50 © 2015 MOSTEC AG, CH-4410 Liestal...
  • Page 10: Alarm Settings

    Change the value with + and - , to set the maximum pump run time to alarm (from 0s to 10800s, whereat 0s means “off”) Note: An alarm message can be acknowledged by pressing ENT on main screen. V1.50 © 2015 MOSTEC AG, CH-4410 Liestal...
  • Page 11: Usb Settings

    04.05.2011 12:12:55 04.05.2011 12:13:03 04.05.2011 12:13:04 04.05.2011 12:13:06 04.05.2011 12:13:07 A. USB error code E: xx error description drive not formatted drive wrong formatted drive is write protected sector size not supported V1.50 © 2015 MOSTEC AG, CH-4410 Liestal...
  • Page 12: Device Settings

    F4, confirm label with ENT. Sets all values to default, Factory setting confirm with ENT or leave with ESC Exit with ENT Shows firmware version, serial Device info number and calibrated point. V1.50 © 2015 MOSTEC AG, CH-4410 Liestal...
  • Page 13: Calibrate The Ph Sensor

    0.6pH off the buffer value, value because the pH-electrode is considered defective. Save data? Press ENT to save the calibrated ''pH7 = 2.3mV” values or discard the values with ''mV/pH = 57.5mV (20°C)'' V1.50 © 2015 MOSTEC AG, CH-4410 Liestal...
  • Page 14: Operating Mode Of The Limits

    4 / b ase ta n k In le t Valve up Proce ss vesse l Pro cess- with stirrer liqu id p H-0 / acid ta nk Va lve d o wn Ou tlet V1.50 © 2015 MOSTEC AG, CH-4410 Liestal...
  • Page 15: Controller With Two Different Set Points [W]

    2 the base valve. Within the p-band (X or X ), the controller handles the valves (acid or base) by its internal control mechanism. [pH] ≈ Totband 10% der P-Bandbreite V1.50 © 2015 MOSTEC AG, CH-4410 Liestal...
  • Page 16: Controller Mode - Adapting The Controller To A Titration Curve

    P-band process liquid to pH 4 (pH 4 + pH 2 = pH 6 ) acid- end of control valve process 2sec. T-on, 15sec. T-off (in middle of the P-band) V1.50 © 2015 MOSTEC AG, CH-4410 Liestal...
  • Page 17: Factory Settings

    Alarm SP2: value 6.50pH Alarm SP2: behavior lower limit Alarm SP2: delay Alarm Rel: norm / inv normal SP1 – SP2: hysteresis Pump monitoring 3'600s (1h) USB settings: Clock setup: Date setup: DDMMYY Log period: V1.50 © 2015 MOSTEC AG, CH-4410 Liestal...
  • Page 18: Mechanical Dimensions

    M3020R Operating Manual Q. Mechanical Dimensions Side view: V1.50 © 2015 MOSTEC AG, CH-4410 Liestal...
  • Page 19: Rear View

    M3020R Operating Manual Rear view: V1.50 © 2015 MOSTEC AG, CH-4410 Liestal...
  • Page 20: Technical Data

    – special measuring range / signal output – custom functions – transparent door Ip55 – custom home screen / logo – 2 additional relay outputs (contact 3/4) – 2-wire transmitter input (* 19 = Supply, 20 = input) V1.50 © 2015 MOSTEC AG, CH-4410 Liestal...