Installation; Receiving; Selection Of Mounting Location - Sentry ISOLOK SAL-DBB Installation, Operation & Maintenance Manual

Table of Contents

Advertisement

Sample Container with Adapter
Dangerous gas! The gases being emitted from the bottle vent may be hazardous and toxic upon exposure . The vent line
should be directed to a charcoal canister, flare, or other sub atmospheric region for collection and treatment of sample
vapors .
Always ensure control is OFF before mounting or exchanging sample containers .
Sample collection containers should always be clean and completely dry before being used . Polypropylene containers
(bottles) are common .
Screw Mount Containers
Standard sampler body is provided with a 1-inch tri-clamp discharge port . Adapters to fit threaded sample collection
bottles are available from Sentry Equipment . These adapters have a suitable connection and mount onto the discharge
port .
Separate Containers
Suitable piping connections for metal or plastic components can be arranged to conduct sample portions to a separate
container . Whenever such piping is installed, avoid long runs with multiple fittings, joints, or other areas where solids or
crystals could build up and block flow .
 NOTE
Select a sample container size that will not be completely filled, even on longest sampling times . Use care when setting a
cycle rate that may nearly fill a container .

Installation

Receiving

ƒ Examine the crate and all contents for any shipping damage immediately after receipt .
ƒ Take pictures of any suspected damage .
ƒ Report damages to the delivery company at once . This is the responsibility of the consignee .

Selection of mounting location

ƒ Select a location where the material in the process stream is well mixed .
ƒ Ensure the sample collection container will clear vertical lines (or other obstructions) when mounted onto the
sampler .
Sentry ISOLOK SAL-DBB Sampler