ICS Schneider Messtechnik PPS 1210 Operating Instruction

High pressure calibrator

Advertisement

Quick Links

high pressure calibrator PPS 1210 / L
ICS Schneider Messtechnik GmbH
Briesestraße 59
D-16562 Hohen Neuendorf / OT Bergfelde
Operating instruction
Tel.: 03303 / 504066
Fax: 03303 / 504068
info@ics-schneider.de
www.ics-schneider.de

Advertisement

Table of Contents
loading

Summary of Contents for ICS Schneider Messtechnik PPS 1210

  • Page 1 Operating instruction high pressure calibrator PPS 1210 / L ICS Schneider Messtechnik GmbH Tel.: 03303 / 504066 info@ics-schneider.de Briesestraße 59 Fax: 03303 / 504068 www.ics-schneider.de D-16562 Hohen Neuendorf / OT Bergfelde...
  • Page 2: Table Of Contents

    Table of Contents Components of high pressure calibrator Notes on the operating instructions 1. Description of the device 2. General safety warnings 3. Operating the calibrator 4. Description of the functions 5. Menu navigation for calibrators 6. Commissioning 7. Maintenance / disposal 8.
  • Page 3: Components Of High Pressure Calibrator

    Screwed sealing plug Manual booster pump Pressure connection for test object (only PPS1110) 10 Drain valve 11 Screw compressor Additional components of High Pressure Calibrator PPS 1210 Heatable pressure port Pressure pipe Temperature controller On / Off switch for pressure port heating...
  • Page 4: Notes On The Operating Instructions

    Notes on the operating instructions Commissioning Intended use A pressure-resistant connection for The high pressure calibrator may - The operating instructions are in- the test object is required in order to only be used to generate pressure tended for specialist workers and use the high pressure calibrator.
  • Page 5: Operating The Calibrator

    3. Operating the calibrator last valid pressure value that was If you can no longer reach the desired measured will flash on the display pressure, please consult the section on Operating the high pressure calibrator (overload warning) "Maintenance" to find out how to vent is described starting on page 6.
  • Page 7: Maintenance / Disposal

    Maintenance / disposal End leak measurement early shown, and press ENTER to activate. You can now use SELECT to choose Venting the pressure system To end a leak measurement early, the function you want, and ENTER to press the ENTER button and confirm Release the pressure completely and execute the function.
  • Page 8: Evalution With The Pc

    Disposal This product must not be disposed of as normal household waste at the end of its useful lifetime. To prevent possible damage to the environment or to health due to uncontrolled waste disposal, this product must be separated from other waste and recycled correctly in order to ensure sustainable use of the raw materials.
  • Page 9: Technical Data

    Technical data Pressure range (FS) 0…700 bar (others on request) Overpressure 840 bar (FS+20%) Resolution of display 1mbar Accuracy, error band (1) (10…40 °C) < 0,05 %FS Accuracy, error band (1) (0…50 °C) < 0,1 %FS Leak rate* 700 bar: -2 bar @ 10 min. Number of digits on display 5 digits Measurement interval...
  • Page 10: Declaration Of Conformity

    Konformitätserklärung Declaration of Conformity Déclaration de Conformité...
  • Page 11 OPERATING INSTRUCTION Temperature regulator General: The temperature regulator is a high quality control instrument for both simple ON/OFF two-position regulation or, if desired, PID control systems with autotuning function. In the ON/OFF mode, the degree of hysteresis is freely programmable. In the ON/OFF mode, the degree of hysteresis is freely programmable The temperature regulator has been designed for heating and cooling operations in the range -200°C to +1,750°C.
  • Page 12 Technical data: Dimensions .... approx 48x48x105 mm (W x H x D) Sampling period: ....<3 sek. Installation depth approx 100mm Max output load ....approx 2A (resistive load, mechanical relay) Mounting cutout 45 x 45mm (DIN 1/16) Max switch voltage ....230 V AC Display viewing area ..
  • Page 13 Changing the set value and alarm temperature: When the regulator is switched on Press the button SU appears on the PV display, and the last figure on the SV display flashes Use the arrow keys to set the control set value to the desired temperature, and confirm your selection via the button The value is saved, and the regulator jumps to the set value setting for the second alarm output.
  • Page 14 Using the menus: The temperature regulator is operated via two menus (A and B). Menu A contains the settings for the set value (see also page 19: ‘Changing the set value and alarm temperature’). These settings are easily accessible to any operator by simply pressing the button to call them up. Repeated pressing of the button moves to the next menu item.
  • Page 15 consec. Parameter P a r a m e t e r n a m e adjustment range description code menu B: "Advanced parameters" K type thermal element / Possible measurement range: - 5 0 . . . + 1 3 5 0 ° C S - type thermal element / Possible measurement range: - 5 0 .
  • Page 16 ‘Automatic adjustment’ (Autotuning) function The control panel features a programme that automatically finds the optimum PID settings for the connected load. For this programme to work, it is important that the regulator is ready for operation, the load is connected and that a non-critical temperature set value has already been set.