Page 1
User Manual 1067 Precision Resistance Decade Box Version 1.2 11-22 Time Electronics Ltd Unit 5, TON Business Park, 2-8 Morley Road, Tonbridge, Kent, TN9 1RA, United Kingdom. T: +44 (0) 1732 355993 | F: +44 (0) 1732 350198 mail@timeelectronics.co.uk | www.timeelectronics.com...
Page 2
All rights reserved. Nothing from this manual may be multiplied, or made public in any form or manner, either electronically or hard copy, without prior written consent from Time Electronics Ltd. This also applies to any schematics, drawings and diagrams contained herein.
Housed in a robust metal case the 1067 is fully screened and low thermal EMF terminals are used. The slimline design means it takes up minimum bench space and is easily transportable.
Time Electronics 1067 Resistance Decade Box User Manual Specifications Range / Resolution ...... 0 to 12 kΩ / 10 mΩ steps. Number of decades ...... 6. Each settable from 1 to 11. Accuracy ........± 0.01 % of setting ± 2 mΩ, after deduction of residual end resistance ±...
Time Electronics 1067 Resistance Decade Box User Manual Operation Safety Precautions Observe proper safety operational guidelines when working with high voltages. To minimize shock hazard connect the case terminal to an electrical ground. Always take precautions to avoid and prevent contact with live components.
Time Electronics 1067 Resistance Decade Box User Manual Operating Instructions 3.3.1 Connections Connection to the decade box is via 4 mm safety terminal posts, using 4 mm shrouded or standard plugs. Alternatively, crocodile clips or stripped wire connections can be used.
Page 8
Set the 0.01 Ω dial to 1. • For a quick reference to the setting values of the 1067, please see the Resistance Setting Table on the following page. Note: All resistance boxes have a residual resistance, meaning that when the dials are set to zero a small resistance remains.
1067 Resistance Decade Box User Manual Applications Pt100 Resistance Simulation Accurate simulation of low ohm values, such as PRT, can be performed by using the 1067. Resistance vs Temperature Relationship for Platinum Resistance Thermometer Detector Element (ITS-90) Ω Ω Ω...
Time Electronics’ total liability is limited to repair or replacement of the product. Note that if Time Electronics determine that the fault on a returned product has been caused by the user, we will contact the customer before proceeding with any repair.
Page 12
User Manual Returning Instruments Prior to returning your product please contact Time Electronics. We will issue a return merchandise authorization (RMA) number that is to accompany the goods returning. Further instructions will also be issued prior to shipment. When returning instruments, please ensure that they have been adequately packed, preferably in the original packing supplied.
Need help?
Do you have a question about the 1067 and is the answer not in the manual?