HI 6600 series Quick Start Guide 
HI 6600 series is a modular system of weight processors that can deliver 
up to 30 channels of fast, stable, high‐resolution weight values to PLCs, 
PACs and DSCs over a single fieldbus network. 
 
 
 
This Quick Start Guide is intended for users that are already familiar 
with setting up Hardy Process Solutions weighing instruments.  
 
 
A complete User's Guide can be found under Documents and Programs 
online:  www.hardysolutions.com     
 
Navigate to the Product menu> Weighing Instruments > Weight 
Processors > HI 6600 > Docs & Programs 
 
 
 
 
 
 

Advertisement

loading
Need help?

Need help?

Do you have a question about the HI 6600 Series and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Hardy Process Solutions HI 6600 Series

  • Page 1   HI 6600 series Quick Start Guide  HI 6600 series is a modular system of weight processors that can deliver  up to 30 channels of fast, stable, high‐resolution weight values to PLCs,  PACs and DSCs over a single fieldbus network.            This Quick Start Guide is intended for users that are already familiar  with setting up Hardy Process Solutions weighing instruments.       A complete User’s Guide can be found under Documents and Programs  online:  www.hardysolutions.com        Navigate to the Product menu> Weighing Instruments > Weight  Processors > HI 6600 > Docs & Programs       ...
  • Page 2   DIN Rail Installation:  1. Place the top slot of the instrument onto the DIN rail first.  2. Rotate the instrument down until the bottom tab of the instrument  enguges with the DIN rail, then press firmly on the bottom on the  instrument until it snaps in place.    Note: Heavy Gage DIN rail recommended.  3. Check for proper DIN rail engagement by gently lifting up on the   bottom and making sure the instrument is securly in place.    4. Repeat with additional units as required for your system.  5. To remove an instrument from a DIN rail, disconnect wiring then lift  up on the bottom‐front corner of the module until it snaps off.         ...
  • Page 3   Note:  The maximum distance between the first and the last  module in the system is 500 feet (150 meters).     The Gateway should placed at the end of a system             Wiring:  1. After units are securely mounted on DIN rail, remove terminal  connectors making note of the wiring markings found on the  connector labels located on the instruments.          Hardy Gateway Module                          Weight Processing Module      2. Cable HardyNet Ports located on the bottom of the units in series  using Cat5e cable with RJ‐45 connectors.    3. Cable the DC power located on the bottom of the units using the  Module I‐I Diagram and by following the notes for Power Input  terminations.   ...
  • Page 4     Power Input Termination Notes:  a) It is NOT necessary to provide power to every module.  Power is  distributed from the Gateway Module to other modules on the  system via the HardyNet Port and the Cat5e cabling.    b) You must use a power‐limited 12‐24 VDC power supply (Class 2)  on the DC input wiring.  DC power should be supplied by a clean  primary line, directly from the DC power source.    c) Make sure the VDC source is off before connecting any wires to  the module and before plugging in the terminal connectors.    d) Connect the 24 VDC voltage wire, ground wire and shield wire  to the connector that plugs into the DC voltage socket located  on the bottom of the Gateway module.  The factory installed  jumper connects the Earth ground and the internal ground  making them common, and should remain in place.        4. Cable the load cells or  other type of strain‐ gauge based sensors  to the Weight ...
  • Page 5 Processing Modules using the I‐I diagram and by following the notes  for Weight Scale Input Termination:        5. Cable the communication ports on located on the Gateway terminal  according to the protocol used and by following the I‐I diagram for  Hardy Gateway Module available on the Hardy website.     6. Connect the optional display to the display port located on the  Hardy Gateway module.    7. Power up system.    Initial Set‐Up:    Set Up may be performed through an optional front panel display or   through the integrated Webserver.  If setting up using the Webserver,   connect an Ethernet cable between the Hardy Gateway Module and   follow the Communications>Ethernet TCP/IP instructions shown   below before proceeding.    Discovering Weight Processing Modules present on the system:     The system requires a process of discovery that allows the Gateway    Module to identify and test communications with all Weight Processor   Modules connected on the HardyNet.  The discovery process is   automatically initated upon power up of the system; verify all modules   on the system have be registered by the Gateway by checking the  ...
  • Page 6 channel number count located in the HardyNet sub‐menu of the   Gateway module menu structure.    The Channel Identification and channel order number may also be  modifed in the HardyNet sub‐menu section during system set up.    Once the discovery process is completed, proceed to setting up each   individual channel by following the basic SET UP instructions for each   channel outlined below.    Suggested steps when setting up the instrument for the first time:     Choose a specific Weight Processor Module to set up by selecting a  channel either from the front panel display or on the webserver.    Channel    1) Choose channel to set‐up      SET UP (minimum)  1) Select Unit of Measure    2) Input Total Scale Capacity    3) Select Graduation Size  4) Select Decimal Point Position    5) Input Motion Tolerance              Filter ...
  • Page 7 Calibration:   Section used to calibrate the instrument to the  sensor(s) used.  C2 eCAL – Electronic Calibration     1) Remove all weight from the scale.  2) Set gravity correction for location of weighing system  3) Press Do C2 Cal.    4) Place a verification weight on the scale to ensure calibration is  successful.    Hard Cal‐ Traditional Calibration      1) Input the Span Weight.  The span weight should be about 80%  of the scale capacity.  2) Input Cal Low Weight.  With nothing on the scale, the Cal Low  Weight is 0, with 5 lbs on the scale, the Cal Low Weight would  be 5.  3) Press Do Cal Low to set the low point on the calibration curve.  4) Place the Span Weight onto the scale.    5) Press Do Cal Hi.    Refer to User’s Guide if adjustments to other calibration‐related  parameters are required.    Communication:  Section used for setting up communications to  the instrument.  Ethernet TCP/IP   ...
  • Page 8 4) Using the unit’s Fixed IP address (the default is  192.168.000.100), the PC must now be assigned a unique IP  address (for example, 192.168.000.101)    There are two simple rules for the IP Address:     It must have the same network identifier as the computer.            It must have a different node identifier than the computer.       5) On a PC running windows, open Internet Properties (TCP/IP).    6) Click in the ‘Use the Following IP Address’ checkbox; then enter  the following into the TCP/IP Properties dialog.          IP Address = 192.168.0.101    Subnet Mask = 255.255.255.000  7) Select OK on the TCP/IP Properties dialog box. The computer is  now fully configured. To return the computer to the original  network settings, return to the ‘Internet Properties (TCP/IP)  dialog, select ‘Obtain an IP address automatically,’ and click OK.    8) The HI 6500 series instrument is now configured to  communicate with the PC. Enter the HI 6500 series instrument  IP address into the Windows PC Web browser to access the  embedded web browser.   For example: http://192.168.000.100  See the HI 6600 User’s Guide for setting up EtherNet/IP,  Ethernet‐UDP  and Modbus‐RTU communication protocols.         ...
  • Page 9 MENU TREE       ...
  • Page 10 Diagnostics:  Section used for diagnostics of system. Refer to User’s  Guide if troubleshooting and use of diagnostics menu is  required.  Display:              Section used for adjusting various aspects of the display,   including contrast, brightness,  and split screen feature.  Filter:  Section used to stabilize weigh readings in an unstable  environment.   WAVERSAVER:  Press the enter button and use the  arrow keys to select the appropriate setting.   WAVERSAVER eliminates unwanted noise and is  typically set at 1.0 Hz.  NumAverages:  Set the number of readings to average  to provide a stable weight.    Information:  Section used for checking instrument serial number,  and firmware revision.  Language:  Section used for changing language.  Operations:  Section used for setting up parameters related to Tare,  Count and Zero of scale.  Security:  It is highly recommended to see the User’s Guide for  setting up security functions on the instrument.   Improper set up of the security settings could cause a  user to be locked out from using the device.  Setup:  Section used for setting up operation parameters of the  instrument.  Capacity:  Sets the normal operating capacity of scale.  Decimal Point:  Sets to decimal point for the weight resolution.  Grads:    Sets the graduation size of the display increments.  ...
  • Page 11 Notes     ...
  • Page 12         Hardy Process Solutions sincerely appreciates your business. We encourage  input about the performance and operation of our products from our  customers. Should you not understand any information in this guide or  experience any problems with this product, please contact our Technical  Support Dept. at:    Phone: (858) 278‐2900   Toll Free: 1‐800‐821‐5831   FAX: (858) 278‐6700   E‐Mail: hardysupport@hardysolutions.com    Or visit our web site at: http://www.hardysolutions.com    Proudly made in the U.S.A            PN: 0596‐0334‐01 Rev. A 07/15 ...