U S E R I N S T R U C T I O N S I N D I V I D U A L C D S ® J O I N T S F O R D Y N A M I C R E D R E S S I O N , C D S ® J O I N T 3 6 0 ° , C D S ® A D A P T E R J O I N T 3 6 0 °...
Individual Instructions For Use Contents 1. Introduction ..................1.1. Foreword ................... 1.2. Customer information ................ 1.3. Mode of operation ................1.4. Intended use ..................1.5. Scope of supply ................. 1.6. Declaration of conformity ..............1.7. Indications ..................1.8. Contra-indications ................1.9.
Page 3
Instructions For Use 2.4.3. Set and adjust the spring tension to the strength required by the patient ....................2.5. Redression range ................2.5.1. Adjust or change the position of the redression range on the CDS Joints ® ..........2.5.2. Adjust or change the position of the redression range on the CDS Nano joints ®...
Page 4
Individual Instructions For Use 5.5. Install the cover ................6. Coniungi connection unit ..............6.1. Disconnect the Coniungi connection unit .......... 6.2. Connect the Coniungi connection unit ..........6.3. Coniungi drilling templates ............... 6.3.1. Drilling template 16 mm .............. 6.3.2. Drilling template 24 mm ..............
1.1. Foreword The joints provide protection and management during the healing process and allow a The individual range of the albrecht GmbH gradual, controlled return to full physiological includes: mobility. • CDS Joint ®...
• Product or go directly to albrecht GmbH before using • Instructions For Use the device. This applies in particular to the safety instructions. Additionally for the CDS Joint ®...
Instructions For Use 1.7. Indications the knee is contra-indicated • knee stretch deficit after lower leg The doctor decides on the treatment based on amputation his diagnosis. • treatment after biceps tendon rupture In general, the device is used is to treat the •...
Individual Instructions For Use immobilisation The orthosis is only intended to be in contact with intact skin. • managed movement of the elbow joint Micro joint ® 1.9. Safety instructions • Joint contraction, among other things, as a result of: The optimal effect of the CDS Joint...
Instructions For Use 1.10. Warranty avoided. • The devices of the individual range are not We guarantee the devices of the individual intended for single use, but for multiple use range for the application period of 3 months. by only one person. joint Nano joint/...
Individual Instructions For Use 2. CDS Joint / CDS Nano joint ® ® 2.1. Selection of joints according to position of use and direction of action Remarks — For treatment in the direction of flexion: The Joints ® Nano joints are basically ®...
Instructions For Use 2.2. Install the anchor on the joint 2.2.1. CDS Joint (max. torque approx. 5.2 Nm) ® Loosen the central screw on the joint using the lock nut wrench. Secure the anchor to the joint using the central screw. Use the black countersunk screws to position the anchor in the position you want.
Note: The explained steps must be performed immediately with any anchor that is assembled for the corresponding joint. For simplicity, only the straight anchor is shown. Note: For each individual special construction joint there is a suitable follower from albrecht GmbH. 00800 03032900 – INTERNATIONAL SERVICE HOTLINE...
Instructions For Use 2.2.3. CDS Nano joint (max. torque approx. 2.2 Nm) ® Loosen the central screw on the joint using the screwdriver supplied. Make sure that the Teflon washer is positioned correctly. The Teflon disc causes pressure on the joint support, which prevents the joint support from having too much play.
Note: The explained steps must be performed immediately with any anchor that is assembled for the corresponding joint. For simplicity, only the straight anchor is shown. Note: For each individual special construction joint there is a suitable follower from albrecht GmbH. 00800 03032900 – INTERNATIONAL SERVICE HOTLINE...
VA material. You should always use a HSS/E drill bit to machine the material and drill at a speed of 400-500 revolutions per minute to prevent the material from hardening. Suitable HSS/E drills can also be purchased from albrecht GmbH. www.albrechtgmbh.com...
Individual Instructions For Use 2.4. Spring tension 2.4.1. Activating the spring tension To activate the spring tension, set the switch to "on". Move the joint to Extension or Flexion against the desired direction of action of the spring force until you feel a slight resistance. Overcome this to activate the spring force. 2.4.2.
Instructions For Use 2.4.3. Set and adjust the spring tension to the strength required by the patient The set spring force is indicated by a scale from 0 to 15 and can be seen through the viewing window on the CDS housing,.
Page 18
Individual Instructions For Use Drehmoment in Nm bei Beugung 90° Gelenk ® Nano Gelenk ® Einstellungsstufen 00800 03032900 – INTERNATIONAL SERVICE HOTLINE...
Instructions For Use 2.5. Redression range The redression range of the CDS Housing is 90°. ® With the CDS joint / CDS Nano joint, the position of this redression area can ® ® be freely selected by means of an adjustable joint support. This means that Joint / CDS Nano joint can be used as both an extension joint and a...
Individual Instructions For Use 2.5.1. Adjust or change the position of the redression range on the CDS Joints ® Remove the two black countersunk screws. You can adjust the position of the redression range in 15° increments. Replace and tighten the countersunk screws again. 00800 03032900 –...
Instructions For Use 2.5.2. Adjust or change the position of the redression range on the CDS Nano joints ® Loosen the central screw with the tool supplied until the anchor can be lifted completely from the gear teeth. You can adjust the position of the redression range in 10° increments. Place the anchor onto the gear teeth and tighten the central screw again.
Individual Instructions For Use 2.6. Options for limiting CDS joints ® 2.6.1. Adjusting the limiter with stop screw Before setting the limit, you must deactivate the spring tension. Set the switch to "off". Move the joint to Extension or Flexion against the desired direction of action of the spring force.
Instructions For Use 2.6.2. Adjusting the limiter with stop wedge Insert the stop wedge at the desired position Secure the stop wedge with the supplied screw. The spring tension must be activated before inserting the stop wedge. www.albrechtgmbh.com...
Individual Instructions For Use 2.7. Setting the fine adjustment for Joint 360° plus and CDS Adapter joint 360° plus ® ® With the CDS Joint 360° plus / CDS Adapter joint 360° plus , you can adjust the ® ® limit (stop) continuously with the fine adjustment.
Instructions For Use 3. CDS Nano Hybrid joint ® The redression range of the CDS Housing is 90°. ® With the CDS Nano hybrid joint, the position of this redression area can be ® freely selected by means of an adjustable joint support. 3.1.
Individual Instructions For Use 3.2. Activating the spring tension To activate the spring tension, set the switch to "on". Move the brace to Extension or Flexion against the desired direction of action of the spring force until you feel a slight resistance. Overcome this to activate the spring force. 3.3.
Instructions For Use 3.4. Adjust the spring tension to the strength required by the patient The set spring force is indicated by a scale from 0 to 15 and can be seen through the viewing window on the CDS housing,. The areas above 15 and ®...
Page 28
Individual Instructions For Use Drehmoment in Nm bei Beugung 90° Nano Gelenk ® Einstellungsstufen 00800 03032900 – INTERNATIONAL SERVICE HOTLINE...
Instructions For Use 4. CDS Micro joint ® By clamping or relaxing the spring, the force can be adjusted without tools and to individual requirements. 4.1. Adjusting and changing the spring tension (max. torque approx. 0.4 Nm) Fold up the clamping ring. By turning the clamping ring in the + direction, you increase the spring force.
Individual Instructions For Use 5. ROM Nano adapter joint By positioning the stop pins accordingly, you can limit the extension and the flexion stops in 10° increments. 5.1. Remove the cover Rotate the cover 90° to the stop. Lift the cover off the joint. 00800 03032900 –...
Instructions For Use 5.2. Adjusting the extension stop To adjust the extension stop, put the brace in Flexion. Position the stop pin in the position you want. 5.3. Adjusting the flexion stop To adjust the flexion stop, put the brace in Extension. Position the stop pin in the position you want.
Individual Instructions For Use 5.4. Immobilise To immobilise the brace, first adjust the extension stop. Move the support against the extension stop and then position the flexion stop on the support. 5.5. Install the cover Make sure that both stop pins of the joint are in the pin slots. Replace the cover and rotate it 90°...
Instructions For Use 6. Coniungi connection unit 6.1. Disconnect the Coniungi connection unit Press the locking bolt on the Coniungi and at the same time pull the orthoses apart. 6.2. Connect the Coniungi connection unit Ensure that the parts are correctly aligned. Slide the two parts into one another until you hear that theConiungi is locked...
Individual Instructions For Use 6.3. Coniungi drilling templates 6.3.1. Drilling template 16 mm All M3 threads 6.3.2. Drilling template 24 mm All M3 threads 00800 03032900 – INTERNATIONAL SERVICE HOTLINE...
Instructions For Use 6.4. Coniungi connection support As an option, a connection support can be ordered for the Coniungi, which facilitates the work with the Coniungi connection element. Coniungi connection unit is delivered assembled with screws on the support. 1 Coniungi connector support 2 Coniungi connection unit mounted on the connector support...
Individual Instructions For Use 7. Nano hand piece 7.1. Mounting the Nano hand piece and adjusting the palm plate Shorten the joint support (CDS Nano anchor straight elongated hole) to the desired length. ® Loosen the fixing screw from the hand rest. Position the black mounting washer on the joint support.
Instructions For Use 7.2. Adjusting the finger plate The finger plate can be adjusted in 15° increments. Loosen the pad below the palm of your hand. Remove the pad. Pull the two elements of the palm rest apart. Move the finger plate into the desired position and position the locking pin. Move the pad below the palm of your hand.
Individual Instructions For Use 7.3. Optional disassembly of thumb plate The thumb plate can be removed if not needed. To do this, loosen the two screws with the tool supplied Lift off the plastic cover. Please note that the thumb plate is made of PE material and is thermally deformable.
Instructions For Use 8. Cleaning, servicing and disinfecting The devices of the Individual range are designed maintenance-free. To ensure proper functioning during the treatment period, the devices should be cleaned Hand wash 30°C regularly (at least every 3 months) or, if necessary, according to the following instructions.
Individual Instructions For Use 9. Technical data / materials Joint / CDS Nano joint / CDS Nano hybrid joint ® ® ® Designation Material Support material Aluminium / steel ROM Nano adapter Joint Designation Material Support material Aluminium / steel Micro Joint Designation Material...
Instructions For Use 10. Disclosure The devices of the individual range are not intended for single use, but for multiple use by only one person. We advise against passing the brace on. If, however, this is desired, please observe the care and cleaning instructions before passing on and have the brace checked for safe and perfect operation by an authorised specialist dealer.
Page 42
Individual Instructions For Use 00800 03032900 – INTERNATIONAL SERVICE HOTLINE...
Page 43
Instructions For Use www.albrechtgmbh.com...
Page 44
PATENTS: EP 0 841 044 / US 5,954,677 / EP 3 352 713 / DE 10 2008 049 854 / DE 10 2015 012 320 / EP 314 6944 FURTHER PATENTS PENDING VERSION: EN 08.2022 Medical device Manufacturer a l b r e c h t G m b H a l b r e c h t G m b H C H I E M S E E S T R A S S E 8 1 D - 8 3 2 3 3 B E R N A U A M C H I E M S E E...