Series5 Mattress
Version 5: 30.09.2022

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Series5 and is the answer not in the manual?

Questions and answers

Summary of Contents for OSKA Series5

  • Page 1 User Guide Series5 Mattress Version 5: 30.09.2022...
  • Page 2 Mattress Pump (This Pump is not an OSKA Product.) Contents Know your symbols European conformity marking European conformity marking Know your symbols This manual contains different See the user manual for use Non sterile About your mattress typefaces and symbols to make...
  • Page 3 The system consists of a foam shell with a electrocution, fire or injury to persons high-density foam topper serving as the support The Series5 is not for use by those with surface underneath the patient. The foam shell also 01. Always unplug this unit immediately after 01.
  • Page 4: Design Features

    (MVTR). Covers easily design wipe clean with standard, hospital-grade cleaners. Cover of Series5 Mattress toggle switch mattress features bi-directional stretch top fabric designed to allow full integration of the user into the surface. Top is made from...
  • Page 5 “Off ”. Plug power cord into wall outlet. Press On/ Off switch “ON”. should be tagged and inspected routinely. 01. Place the Series5 mattress on the bed frame with the airline connectors at the foot end of the bed. The mattress has a black bottom...
  • Page 6: Using The Pump

    horizontal position. Bending the If “Low Pressure” indicator light comes on after support surface and the body at the midpoint initial set-up or when moving mattress or control when elevating the HOB concentrates the body unit, first check that all airlines are properly Series5 Mattress User Guide...
  • Page 7 Protect from damage. Avoid pressure management performance. Use of Series5 mattress is only one element of temperature extremes (below freezing or above care in the prevention and treatment of pressure 48°C).
  • Page 8 1.3 Some surface wrinkling may result from training. It is advisable to mark any equipment cleaning procedures. This should have no Training/practise The Series5 comes with evacuation straps within adverse effect on the fabric’s properties. used for training as “Training equipment” and it‘s base.
  • Page 9: General Guidelines

    remove the filter, and the surface. OSKA V-Guard Technology covers 5.1 Regularly check the inside of the cover for any cutting, or from being worn through. They can hand wash using warm water and mild detergent.
  • Page 10: Specifications

    Defective control unit Call for service Classification: Class II, Type BF No AP or APG protection Patient exceeds weight limit Call OSKA for alternative products Weight: 2.1kg Mattress not inflating Control unit is not turned on Turn control unit on 29.21cm x 17.78cm x 11.43cm...
  • Page 11 Centre of mattress (for a user sleeping alone), or on Patient Right Side (for two users, or for a single user who cannot/prefers not to move to centre of mattress). Air cylinders not connected to digital control unit must be maintained at desired pressure through regular adjustment/inflation with manual pump kit, item # SERVICE-00. Series5 Mattress User Guide...
  • Page 12 03. Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the Pump, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result. Series5 Mattress User Guide...
  • Page 13 30 A/m 30 A/m Power frequency magnetic measures may be necessary, such as reorienting or relocating the device. (50/60Hz) fields should be at levels magnetic field characteristic of a typical b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m. IEC61000-4-8 location in a typical commercial or hospital environment Series5 Mattress User Guide...
  • Page 14 NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people. Series5 Mattress User Guide...
  • Page 15 Series5 Mattress User Guide...
  • Page 16 Series5 Mattress User Guide...
  • Page 17 Pressure Care Experts Edward House, 5 Penner Road, Havant, PO9 1QZ 02394 318 318 | ask@oska.uk.com oska.uk.com Series5 Mattress User Guide...