5.3.2 Texmate DI-50E Panel Meter Setup
STEP A CODE SETUP: Instrument Model No. MVP-DI50E (Multi- Pycnometer)
5.3.2.1
Press Program and Up-arrow at the same time:
This enters the main programming mode. Display toggles between [bri]
and [5].
Press Program to set bri to 5. If bri is not set to 5 then use up/ down arrow
and
set bri to 5.
5.3.2.2
Press Program button
Display toggles between [CAL] and [000].
5.3.2.3
Press Program button again
Display toggles between [Cod_1] and [000].
Change Cod_1 to 001 for model#24000-2-0-D. Enter 061 for
model#20033-1
5.3.2.4
Press Program button
Display toggles between [DiSP] and [003].
Change DiSP to 004. (This will move the decimal)
5.3.2.5
Press Program button
Display toggles between [Cod_1] and [061]
Change Cod_1 to 000.
5.3.2.6
Press Program button
Display toggles between [Cod_2] and [XXX]
Change Cod_2 to 001.
5.3.2.7 Press Program button
Display toggles between [Cod_3] and [XXX]
Change Cod_3 to 000.
5.3.2.8 Press Program button
Display toggles between [Cod_4] and [XXX]
60 Hz
Change Cod_4 to 000.
50 Hz
5.3.2.9
Press Program button
Display toggles between [Cod_5] and [XXX]
Change Cod_5 to 000.
5.3.2.10 Press Program button until you have reached Cod_8.
Display toggles between [Cod_8] and [XXX]
Change Cod_8 to 002.
5.3.2.11 Press Program button
Display toggles between [Cod_9] and [XXX]
Change Cod_9 to 000.
5.3.2.12 Press Program button
Display will indicate instrument's current pressure.
Verify the decimal is set as XX.XXX
Note: Set Cod_4 to 000 for