Ventilation unit

Advertisement

Pocket Guide
Babylog® VN500
Software version 2.n

Advertisement

loading

Summary of Contents for Drager Babylog VN500

  • Page 1 Pocket Guide Babylog® VN500 Software version 2.n...
  • Page 2: Table Of Contents

    QUICK GUIDE BABYLOG® VN500 02 | Content System overview Operating concept Getting started Operation Alarms Monitoring The Babylog® VN500 Pocket Guide is not a replacement or substitute for the Instructions for Use. It is for informational purposes only. Any use of the device requires full understanding and strict observation of the Instruction for Use.
  • Page 3 | 03 Intended use Babylog VN500 ventilation unit The Babylog VN500 ventilation unit of the Infinity Acute Care System is intended for the ventilation of pediatric patients and neonates. Babylog VN500 provides mandatory ventilation modes and ventilation modes for supporting spontaneous breathing and also airway monitoring.
  • Page 4: System Overview

    QUICK GUIDE BABYLOG® VN500 04 | System overview Intensive care ventilator Control and display unit Infinity C500 Ventilation unit Gas supply unit GS500 Power supply unit PS500 Trolley with 4 brakes Babylog® VN500...
  • Page 5 | 05 System overview Cockpit Front Panel Alarm bar Touch screen Rotary knob Audio paused ON/OFF button touch – turn – confirm button AC power LED – lights up when the device is connec- ted to AC power Battery LED –indicates battery status...
  • Page 6 QUICK GUIDE BABYLOG® VN500 06 | System Overview Cockpit Back Panel Alarm bar VESA mounting holes USB ports Serial communication ports Single-link Digital Isolated network Visual Interface (DVI) connectors...
  • Page 7 | 07 System Overview Ventilation unit, front panel Operation display Flap of ventilation Expiration valve Expiratory flow sensor Inspiration valve Emergency air intake Nebulizer port Water trap...
  • Page 8: Operating Concept

    QUICK GUIDE BABYLOG® VN500 08 | Operating concept Main screen Header bar Monitoring area with curves, loops, trends and measu- red values Main menu bar with buttons for opening dia- log windows and activat- ing functions Therapy bar with the therapy controls for the ventilation parameters of the active ventilation mode...
  • Page 9 | 09 Operating concept Main screen System data, Therapy status: Alarm status: e. g., state of Therapy type, Alarms, messages Alarm silence charge of the Patient ventilation mode, and instructions and display of batteries category additional settings for the user remaining time...
  • Page 10 QUICK GUIDE BABYLOG® VN500 10 | Operating concept Alarm priorities A certain priority, indicating the urgency, is assigned to each alarm. The following table shows the differences between the alarm priorities with respect to identification and the action required. Colour Alarm priority Action required High !!!
  • Page 11 | 11 Operating concept Touch – Turn – Confirm Selecting and setting parameters and functions Touch Turn Confirm the control element. the rotary knob to the right or by pressing the rotary knob. The control element to the left to make the setting. The setting is applied, and turns yellow.
  • Page 12: Getting Started

    QUICK GUIDE BABYLOG® VN500 12 | Getting started Switching on Babylog VN500 and Infinity C500 Prerequisites: – Babylog VN500, Infinity C500, Ventilation unit, PS500 and GS500 are left side: reprocessed and assembled Toggle switch ready for operation. – The mains power supply and the gas supply are connected.
  • Page 13 | 13 Getting started Selecting the breathing circuit and the breathing gas humidifier The breathing circuit and the breathing gas humidifier can only be selected in standby mode. 1. Touch the Start/ Standby... button in the main menu bar. 3./4. 2.
  • Page 14 4./5. 1. Touch the Breathing circ. check tab. 2. Confirm with the rotary knob. 3. When requested by Babylog VN500 in the instruction field: Seal the patient connection port, e. g., with a sterile glove. 4. Confirm with OK.
  • Page 15 | 15 Getting started Selecting a patient Admitting a new patient For a new patient, Babylog VN500 determines the ventilation parameters’ start-up settings based on the patient category or the body weight (height in ped. patients). Ventilation parameter start-up settings by...
  • Page 16 QUICK GUIDE BABYLOG® VN500 16 | Getting started Selecting the Tube or NIV application mode Babylog VN500 can switch between non- invasive ventilation and tube ventilation. The application mode can only be selected in standby mode. 1. Touch the Start/Standby ... button in the main menu bar.
  • Page 17: Operation

    | 17 Operation Setting ventilation Opening the Ventilation settings dialog window. The Ventilation settings dialog window can be opened as follows: 1. Touch the Ventilation settings... button in the main menu bar. 2. Touch the ⬆ button in the therapy bar.
  • Page 18 QUICK GUIDE BABYLOG® VN500 18 | Operation Setting ventilation parameters 1. Prerequisite: The General settings page is opened. 2. Touch the corresponding therapy control, e. g., RR. The colour turns yellow. 3. Set the value by turning the rotary knob and push to confirm. 4.
  • Page 19 Babylog VN500 opens the Start/ Standby dialog window. Touch the Standby button and confirm with the rotary knob. Babylog VN500 is in standby mode. 2. Touch the O Therapy button. The message field displays the information to use specific masks for O therapy.
  • Page 20: Alarms

    Display of alarms – Optical alarm signals Alarm bar Alarm message field Blinking parameter field Display of alarms – Acoustic alarm signals Babylog VN500 generates different alarm tone sequences to indicate alarms acoustically. The acoustic alarm signal can be suppressed for 2 minutes maximum.
  • Page 21 | 21 Alarms Displaying current alarms To display the current alarms, proceed as follows: 1. Touch the alarm message in the header bar. 2. Touch the Alarms ... button in the main menu bar. 3. Touch the Current alarms tab. Acknowledging all alarm messages that are no longer active: 4.
  • Page 22: Monitoring

    QUICK GUIDE BABYLOG® VN500 22 | Flow monitoring Calibrating the neonatal flow sensor Manual calibration of the neonatal flow sensor is necessary: − During the device check and before use – At least once a day – After replacing the neonatal flow sensor –...
  • Page 23 Flow monitoring Starting calibration of the neonatal flow sensor 1. Touch the Sensors Parameters ... button in the main menu bar. Babylog VN500 opens the Sensors/ Parameters dialog window. The Neonatal flow sensor page appears by default. 2. Select the sensor type being used: –...
  • Page 24 Not all products, features, or services are for sale in all countries. Mentioned Trademarks are only registered in certain countries and not necessarily in the country in which this material is released. Go to www.draeger.com/trademarks to find the current status. CORPORATE HEADQUARTERS Drägerwerk AG &...