MAMOS – UPGRADE OF ANALYSER'S FIRMWARE
Service manual
Version: 1.2
12/2022

Advertisement

loading

Summary of Contents for madur electronics MAMOS

  • Page 1 MAMOS – UPGRADE OF ANALYSER’S FIRMWARE Service manual Version: 1.2 12/2022...
  • Page 2 User manual maMoS 1. PROCEDURE OF FIRMWARE UPGRADE - CPU 1. Turn OFF the analyser and disconnect from mains 2. Open the analyser’s casing – remove 6x blanks and unscrew 6x Pozidriv PZ2 screws.
  • Page 3 User manual maMoS 3. On the CPU board → install jumper – short-circuit the two top pins. 4. Turn ON the analyser. Analyser will start in programming mode. Display will show “Waiting for CPU...” 5. Download WSD program from Analog Devices and install it on your PC:...
  • Page 4 7. Select the communication port (the one you use to communicate with mamos via PC). Connect PC ↔ mamos using USB cable. 8. Select Erase the CODE ONLY option (otherwise all mamos settings will be lost) 9. Confirm settings by hitting OK...
  • Page 5 User manual maMoS 10. Press Download button to select a file with firmware – mamos3.hex Firmware must match the analyser’s hardware configuration – loading inappropriate hex file can (in the best case scenario) disable analyser. Please consult every firmware update with madur technical support: service@madur.com...
  • Page 6 User manual maMoS 12. After upgrade program will report the status 13. Turn the analyser OFF 14. Remove the jumper (3) and turn the device ON...
  • Page 7 User manual maMoS 2. PROCEDURE OF FIRMWARE UPGRADE – DISPLAY 1. Turn OFF the analyser and disconnect from mains 2. Open the analyser’s casing – remove 6x blanks and unscrew 6x Pozidriv PZ2 screws. 3. To perform Display upgrade special cable is needed: •...
  • Page 8 User manual maMoS 4. Connect adapter to display PCB, to connector called Prog.
  • Page 9 User manual maMoS 5. Download Silicon Labs Flash Utility Silicon Labs Flash Utility 6. Select USB Debug Adapter 7. Check “Power target…” option 8. Press Connect...
  • Page 10 User manual maMoS 9. Program confirms it has been connected...
  • Page 11 User manual maMoS 10. Select tab: Download Hex File/Go/Stop 11. Select hex file (usually called Disp4x20.hex) 12. Check option Erase all code… 13. Press Download...
  • Page 12 User manual maMoS 14. Firmware is loading...
  • Page 13 User manual maMoS 15. Program is successfully loaded 16. Close flashing program 17. Disconnect Silicon Labs debugger 18. Put back the analyser. 19. Turn the analyser ON...