Toshiba FDX4343R Technical Data Manual page 22

X-ray flat panel imager

Advertisement

FDX4343R
25. Temperature Sensor
The FPI is equipped with temperature sensor. Alarm appears when it detects abnormal internal
temperature. Stop operating FPI when it appears then check surrounding temperature.
26. Cables
Users must use cables attached with the FPI.
Be sure to connect correctly and do not loosen cables.
27. Ethernet
Cross type Ethernet cable must be used. Also, do not connect Ethernet cable to common
networks, must be directly connected to PC.
Do not disconnect Ethernet connection while DC24V is operating and supplying to FPI.
28. Fuses
Fuses are installed inside of power supply unit. Stop immediately using the FPI when fuse
breaks.
29. Defects
It is natural due to characteristics of TFT that the FPI has defects. Please calibrates/corrects
defects upon usage of FPI. Defects can be corrected by using defect maps.
30. Vibration Noise
Do not apply vibration noise during operation. It may becomes the cause of image deterioration
or noise.
31. Service/Maintenance
Service/Maintenance should be carried out periodically according to the technical data sheet or
the specification sheets.
32. Shipment
FPI should be removed from equipment then put into polyethylene-type bag and pack in TETD
packing box to be protected. Keep FPI input plate faced upwards during shipment.
33. Scrap
FPI has lead plate for the prevention of the X-ray leakage. The lead powder or vapor is harmful
for human health. Dispose the scrapped products according to the requirement of local
regulation. If you have any questions, please contact to our local sales representatives for
further information.