FDX4343R
7. Installing to the Equipment
Installing FPI housing to the equipment by according to the technical data sheet or the
specification. It will prevent dropping off of FPI from the equipment.
8. Shipping and Handling
The weight of FPI is mentioned at technical data sheet or the specification. Please take into
account on this weight for preventing any dangers during usage.
9. Unnecessary X-ray, Qualified Personnel
The FPI is used combined with X-ray tube assembly. Only Qualified personnel for X-ray handling
procedures can use the FPI.
For set-ups and services, only knowleged enigineers or special service personnels are able to
handle the FPI.
10. X-ray Shield
When using this device, unnecessary X-ray irradiation must be avoided.
(1) Please follow the "minimum distance between the X-ray tube focal point and X-ray incidence
plane" and "maximum transmission tube voltage" specified in the technical data.
(2) This devise has X-ray shield only for electronic circuit protection and does not comply with the
international safety standard IEC 60601-1-3. Therefore, it requires additional X-ray protection
measures for X-ray system to comply with the IEC 60601-1-3.
11. X-ray Specifications
Please follow the ratings below.
(1) Maximum transmission tube voltage .......................................................................... 150kV
(2) Maximum X-ray incident dose rate ............................................. 8.73 x 10
(3) Minimum distance between the X-ray tube focal point and X-ray incidence plane ...... 73cm
12. Stop Operation
A limited operating life and the possibility of random failures are inherent to FPI and shall be
taken into account for the protection of personnel and equipment. The operation of FPI under
inappropriate conditions, either due to the lack of care or knowledge, can lead to grave risk to
the life and limb of personnel, independent of the risk of FPI and/or equipment damage.
13. Contact to Patient
The FPI is not for equip to the patients. Install the FPI not to contact or touch the patients.
14. Installation
Be sure to install the FPI not to touch the patients during operation.
-20-
-3
Gy/min [1R/min]