  • Page 1 FDY10 Battery Tester User Manual Version: 2.0...
  • Page 2 SUMMARIZE: FDY10-S Battery tester structural design in reasonable, easy to carry, may detection battery capacity, can measure battery discharge time and discharge AH. Comprehensive detection and measure two function new type measure equipment . Can adaptive from 1V to 20V voltage range different varieties different types, measure battery varieties: Ni-Cd, Li-Ion, Pb, Ni-Mh, Ag-Zn, LMP, Li-po, Pb...
  • Page 3: Specification

    SPECIFICATION: Set the battery discharge current: If the measured battery voltage is low, user can free to set current between 0.1A ~ 10A; If the battery voltage over 10V, user can free to set current between 0.4A ~ 10A. Step value in setting mode is 0.1A, press and hold can faster set step in 1A, If open the constant resistance function mode, the setting is invalid, actual discharging current is determined by the resistance of the battery voltage and external power, please refer the detail in OPERATION section.
  • Page 4: Operation

    Star/Stop: Press "OK" key for start, press again stop. Time and capacity is accumulate, press and hold key for reset the memory, or auto-reset when power off. Discharge warning’s: the display flick and buzzer warning when discharge stop, press any key to stop warning. External power resistance cooling model: when discharge process the resistance is very hot! Please avoid touch and place in safety place.
  • Page 5 5.2.- Function 2 (setup battery discharge current): Press key and key decrease and increase current value, press and hold can change value step quickly. 5.3.- Function 3 (save and exit): After setup finish, press key change the display show " ES-y", press "OK" key, means save change and exit setup mode.
  • Page 6 flashing mean in running mode. 7.- In the process of discharge, battery voltage is decrease, when battery voltage less than configuration voltage, tester will automatic stop discharge. If use quickly test model, when time counter to zero will stop battery discharge, display flashing and BUZZER alarm user to check the result.
  • Page 7 Do not use this test in over measurement range!! Under full load discharge mode, the high power resistors temperature is very high, please placed in a safe place, to prevent scalding accidents. PARKING LIST: FDY10-S 1PCS High-power resistance (1 /200W) 1PCS...