NDT James Instruments Chlorimeter C-CL-3000 Operator's Manual page 15

Table of Contents

Advertisement

Meter will now display
Then:
Rinse the probe in distilled water and thoroughly blot dry. Place the
probe filled with electrode wetting agent in the 0.05% solution. Wait
until mV value (xxx.x in above diagram) is stable and the same as
above.
Press
(enter)
Meter will now display: Saving Value
Then:
Rinse the probe in distilled water and thoroughly blot dry. Place the
probe filled with electrode wetting agent in the 0.1% solution. Wait
until mV value (xxx.x in above diagram) is stable.
Press
(enter)
Meter will now display: Saving Value
Then:
Rinse the probe in distilled water and thoroughly blot dry. Place the
probe filled with electrode wetting agent in the 0.3% solution. Wait
until mV value (xxx.x in above diagram) is stable.
Press
(enter)
Saving Value
0.05% Solution
xxx.x mV NOTOK
0.1% Solution
xxx.x mV NOTOK
0.3% Solution
xxx.x mV NOTOK