Calibration - NDT James Instruments Chlorimeter C-CL-3000 Operator's Manual

Table of Contents

Advertisement

Safety
done, the display readings will all be negative numbers.)

Calibration

Upon Power Up the meter will display the main menu:
MEASURE
Use the
navigate to the CALIBRATE menu screen.
Press
Meter will now display: 0.005% Solution
Remove Black Cover Cap and place the probe filled with the
electrode wetting agent in the 0.005% solution. Wait until the mV
value (xxx.x in above diagram) is stable and OK appears in lower
right corner of the display.
Press
Meter will now display: Saving Value
Then:
Rinse the probe in distilled water and thoroughly blot dry. Now
place the probe filled with the electrode wetting agent in the 0.01%
solution. Wait until the mV value (xxx.x in above diagram) is stable
and OK appears in lower right corner of the display.
Press
www.ndtjames.com 14
(plus arrow) or
(enter) to select.
xxx.x mV NOTOK
(enter)
0.01% Solution
xxx.x mV NOTOK
(enter)
(minus arrow) buttons to