Quick Reference Guide
Intermittent
operation
Cardinal Health™ Kangaroo™ Joey
Enteral Feeding Pump
Below are some of the probable causes of alarm conditions and how to correct them.
Symptom Probable cause
Pump in Holding mode
Hold error
for more than 10 minutes
Occlusion in pump-patient
Flow error
line (downstream occlusion)
Empty bag or occlusion in
Feed error
bag-pump line
(upstream occlusion)
Pump Set
Pump Set in use beyond
use >24 Hrs
recommended length of time
Pump Set
Pump Set not properly
dislodged
loaded to pump
Battery low
Battery charge too low
Pump Set tubing not
properly loaded on rotor,
Rotor error
or other unusual rotor
operating condition
General error caused by
System error
many factors; see Manual
for error list
For additional operating instructions and
information, see the Operating Manual for
the Kangaroo™ Joey Enteral Feeding Pump
"Intermittent Operation."
Programming and Menu
Selection buttons
Corrections
Press } Continue to return to Holding screen, or press
Power button to turn off. Several options
and hold the
are available from the Holding screen, such as } Run and
} Adjust Settings.
Check for and correct the occlusion in the line
between pump and patient. Replace Pump Set if
error can't be resolved.
Check for empty bag and refill, or check for and correct
the occlusion in the line between the pump and feed bag.
Replace Pump Set if error can't be resolved. Check the
valve pocket in the Pump Set loading area for dampness
or dirt.
Informational warning message blinks on upper left of
screen. It's recommended that the Pump Set be replaced.
Check black ring retainer (MISTIC) and valve; reload Pump
Set and restart pump to continue. Replace Pump Set if
error can't be resolved.
Immediately plug pump into AC outlet to recharge battery
and to provide power for pump.
Check that the tubing is loaded on the rotor and
that it's not damaged, torn, etc. Reload tubing or
replace Pump Set.
Power down pump and attempt restart to clear error.
Call customer service with screen code if error
can't be resolved.
Status LEDs
Red: Error
Yellow: Hold/Pause
Green: Normal operation
Power button

Advertisement

loading
Need help?

Need help?

Do you have a question about the Kangaroo Joey Enteral and is the answer not in the manual?

Questions and answers

Summary of Contents for Cardinal Health Kangaroo Joey Enteral

  • Page 1 Status LEDs Intermittent Red: Error Yellow: Hold/Pause Green: Normal operation operation Cardinal Health™ Kangaroo™ Joey Enteral Feeding Pump Power button Programming and Menu Selection buttons Below are some of the probable causes of alarm conditions and how to correct them.
  • Page 2 * For additional operating instructions and information, see the Operating Manual for the Kangaroo™ Joey Enteral Feeding Pump “Intermittent Operation.” © 2020 Cardinal Health. All Rights Reserved. CARDINAL HEALTH, the Cardinal Health LOGO and KANGAROO are trademarks of Cardinal Health and may be registered in the US and/or in other countries.