11. MAINTENANCE

To prolong the working life of the bed and for trouble free operation it is
recommended that routine inspections should be performed by a person
who is suitably qualified and well acquainted with the design, use and care
of your FL250 Bed.
If inspection reveals that the safety of the bed is jeopardised in any way, it
should be IMMEDIATELY WITHDRAWN FROM SERVICE, tagged as such and
the owner notified.
NOTE: Other than general maintenance activities all other service or repair
work must be performed by qualified and trained persons who are approved
by Aidacare.
NOTE: For easy access to all parts of the bed and for safe handling,
use the EasyMove Transport System.
The bed must be disconnected from the mains power supply
before starting any maintenance activity.
WARNING
General
Examine all the warning signs on the bed, examine the bed for obvious signs of
1
damage. All aspects of the equipment should operate as intended. Check that
all nuts, bolts and other fasteners are tight and are not missing.
Examine flexible cables and conduits for cuts, cracks, abrasions or
2
other deterioration.
Check that the power supply plug is not damaged. If either the power cable or
3
plug is damaged then both the cable and plug must be replaced as a complete
assembly by an approved service agent.
Brake System
1
Activate brakes in the locked position and apply pressure to the bed foot end.
2
If the bed castors rotate the brakes should be adjusted accordingly.
If the bed does not stay stationary the bed should be removed from service
3
until the issues are rectified.
Battery Backup
Check battery indicator light on attendant control panel to monitor condition
1
of battery back-up.
48
aidacare.com.au

12. TROUBLESHOOTING

IMPORTANT!
Prior to making any adjustments to the electronics of the bed, the control box
should always be reset to rule out any calibration issues.
Calibration issues can occur when power outages happen or when beds are
unplugged from the wall during use.
To complete a reset:
1.
Hold down the top two buttons of the handset (Top Left + Top Right)
2. After 5 seconds continue to hold and the bed will enter reset sequence
This will result in all actuators retracting to their minimum length which
manifests as the bed flattening and descending to the lowest position
3. Once the bed reaches this position a faint relay "click" can be heard and the
handset LED light will flash several times indicating reset procedure is complete
4. If the bed does not perform normally after being reset, please explore
the trouble shooting guide suggestions
Effect
Possible Cause
Verification
Bed does not
Braking system
Brake pedal is not engaged
stay in place
is not in Total
in lowest position
Brake position
Brake pedal is in lowest
Brake System is
position, but bed does not
malfunctioning
stay in place
Bed does
Bed is not
Check if the power lead is
not move
plugged in and
plugged into mains power
up or down
optional battery
Swap actuators
is flat
Swap handsets
Actuator is faulty
Swap NCP
Handset is faulty
NCP is Faulty
Backrest
Actuator code is
Check if cables have
does not
unplugged
come loose from control
move up
box (under top frame at
Actuator is faulty
or down
head end)
Handset is faulty
Swap handsets
NCP is Faulty
Swap Actuator
Swap NCP
Corrective Action
Brake pedal is not engaged
in lowest position
Contact your sales
representative or a qualified
Aidacare service technician
If unplugged, plug power
leads into mains power.
Allow 3 hours to charge
battery (if fitted)
Conduct electronic system
reset - refer to section 6.4
Contact your sales
representative or a qualified
Aidacare service technician
If loose, fully plug cables
into control box
Conduct electronic system
reset - refer to section 6.4
Contact your sales
representative or a qualified
Aidacare service technician
49