BEST System Installation
Note that it is not necessary to connect the Ethernet cable, because the
PreDAC
devices operate as slaves of the
communication is still available, but it operates in read-only mode.
The fiber optic cables from the peripheral devices must then be connected to the
Central Unit. Similar as for the
transceivers should be plugged into the dedicated slots on the rear side of the
Central Unit, but in this case the correct orientation of the SFP transceiver is
upside-down.
On the rear side of the
numbering, as can be seen in
while SFP D is the one on the left).
Figure 12: SFP connection
BEST
TetrAMM
BEST
Central Unit there are 4 SFP slots. Channel incremental
, is right-to-left (SFP A is the one the right
Figure 13
Reserved connectors
SFP D
SFP C
SFP B
Dedicated 4 SFP connectors
and 2 triggers
Figure 13: BEST Control unit SFP connection
Control and Interface unit. The Ethernet
and
PreDAC
devices the SFP optic
for future use
SFP A