Mueller HBMAG Operating Instructions Manual

Mueller HBMAG Operating Instructions Manual

Electromagnetic flow meters

Advertisement

Quick Links

MUELLER HBMAG
Electromagnetic Flow Meters
ELECTROMAGNETIC WATER METER
OPERATING INSTRUCTIONS

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the HBMAG and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Mueller HBMAG

  • Page 1 MUELLER HBMAG Electromagnetic Flow Meters ELECTROMAGNETIC WATER METER OPERATING INSTRUCTIONS...
  • Page 2 WARNING Mueller products may only be used for the applications described in the catalog and in the relevant technical documentation. If products and components from other manufacturers are used, these must be recommended or approved by Mueller. Proper transport, storage, installation, assembly, commissioning, operation and maintenance are required to ensure that the products operate safely and without any problems.
  • Page 3: Table Of Contents

    Electromagnetic Flow Meters TRADEMARKS All names identified by ® are registered trademarks of Mueller. The remaining trademarks in this publication may be trademarks whose use by third parties for their own purposes could violate the rights of the owner. DISCLAIMER OF LIABILITY We have reviewed the contents of this publication to ensure consistency with the hardware and software described.
  • Page 4 9.3 Transmitter ............................................... 28 9.4 Power Supply ..............................................29 9.5 Output Characteristics ..........................................30 9.6 Meter Uncertainty ............................................32 9.7 The effect of temperature HbMAG ...................................... 33 9.8 Dimensions and Drawings ......................................... 34 A. Appendix ................................................... 37 A.1 Unit Conversion Tables ..........................................37 A.2 Parameter Lists .............................................
  • Page 5 MUELLER HBMAG Electromagnetic Flow Meters A.2.6 500 - 599 ..............................................43 A.2.7 600 - 799 ............................................... 44 A.3 Sizing Sensor ................................................ 49 A.3.1 Sizing table 3" ... 48" (DN 80 ... 1200) ................................... 49 A.4 Certificates ................................................49 A.4.1 Certificates ..............................................49 A.5 Spare parts / Accessories ..........................................
  • Page 6: Introduction

    Literature CD The Operating Instructions are available on the CD-ROM INSPECTION shipped with the device, and on the Internet on the Mueller Check for mechanical damage due to possible improper homepage, where further information on the range of Mueller handling during shipment. All claims for damage are to be HbMAG may also be found: Product information on the made promptly to the shipper.
  • Page 7: Safety Notes

    CAUTION Material compatibility Mueller can provide assistance with the selection of the proper meter. However, the full responsibility for the selection rests with the customer and Mueller can take no responsibility for any failure due to material incompatibility. 2.3 LITHIUM BATTERIES Lithium batteries are primary power sources with high energy content designed to represent the highest possible degree of safety.
  • Page 8: Description

    Mueller HbMAG flow meters are suitable for indoor and to optimize water supply and billing. outdoor installations. The Mueller HbMAG is configured to achieve up to 6 years of • Make sure that pressure and temperature specifications battery life in typical revenue applications with the integral indicated on the device type plate / label are not exceeded.
  • Page 9: Sensor Installation

    MUELLER HBMAG Electromagnetic Flow Meters INLET AND OUTLET CONDITION To achieve most accurate flow measurement it is essential to have certain straight inlet and outlet pipe lengths as shown (Di: sensor diameter). min. 5x D min. 3x D FIGURE 4 - 2 REMOTE INSTALLATION The installation consists of two steps: Sensor installation.
  • Page 10: Orienting The Sensor

    MUELLER HBMAG Electromagnetic Flow Meters 4.2.2 ORIENTING THE SENSOR The curves are applicable to water. HORIZONTAL PIPES EXAMPLE p [psi] [mbar] 1.50 0.75 0.60 0.45 0.30 0.15 V=8/ms (25ft/sec) V=7/ms 0.075 Sensor must be mounted as shown in upper part of figure.
  • Page 11 MUELLER HBMAG Electromagnetic Flow Meters TORQUES ADVICE FOR GASKET SELECTION • Only use flat rubber gaskets. NOMINAL SIZE CLASS 150 AWWA • Thickness 1 ... 6 mm (0.0...0.02 ft) dependent on INCH F / LBS F / LBS gap / tolerance.
  • Page 12: Potential Equalization

    Built/in grounding electrode Grounding rings mounted on HBMAG 4.4 GROUNDING The sensor body must be grounded using grounding / bonding straps and/or grounding rings to protect flow signal against stray electrical noise and/or lightning.
  • Page 13: Cathodic-Protected Pipes

    MUELLER HBMAG Electromagnetic Flow Meters 4.6 POTTING AND DIRECT BURIAL COMBINATION OF METAL AND PLASTIC PIPES Use straps for metal pipe and grounding rings for plastic pipe. COMBINATION OF METAL AND PLASTIC PIPES CAUTION M6 x 16 Do not pot meter before electrical connections have been made.
  • Page 14: Transmitter Installation

    MUELLER HBMAG Electromagnetic Flow Meters SUGGESTIONS FOR DIRECT BURIAL OF REMOTE SENSOR PIPE MOUNTING FIGURE 4 - 5 PIPE MOUNTING - HORIZONTAL Remote sensor is protected to IP68 / NEMA 6P and can FIGURE 4 - 4 PIPE MOUNTING - VERTICAL be buried.
  • Page 15: General Safety Requirements

    MUELLER HBMAG Electromagnetic Flow Meters 5.1 GENERAL SAFETY REQUIREMENTS Verify that model and serial numbers shown on labels of sensor and transmitter are matched properly. WARNING Ensure that cable is safety installed to avoid damage of cable and connectors. Please note the different connector The pertinent regulations must be observed for electrical installation.
  • Page 16 Male battery connector with black and red wires; black wire is ground. Male battery connector has to be Battery backup input connected to female connector on backup battery. Functional ground Black wire with terminal must be connected to HbMAG encapsulation with a screw. CABLE INSTALLATION ~10 mm ~100 mm Choose the correct glands for the selected cable type, see Accessories (Page 55) for glands selection.
  • Page 17: Outputs

    Positive pulse logic Negative pulse logic Other radio interface cable has to be a 3-wire cable with a shield connected to HbMAG housing (mounting cable shield is Signal shown to the right). Pulse output can be configured as volume, alarm or call-up, see HbMAG Commissioning (Page 13).
  • Page 18: Connection Of Add-On Modules

    A brief press (less than 2 seconds) will advance screen to next index or menu. Refer to the relevant Quick Start or Operating Instructions available at the HbMAG literature CD or on the internet at: A short press (2 to 5 seconds) will enter a menu or escape www.MuellerSystems.com.
  • Page 19: Display Of Symbols

    MUELLER HBMAG Electromagnetic Flow Meters 6.2 DISPLAY OF SYMBOLS Menu bar Top area of display shows status bar. Index for actual menu and acceptable value indication Status bar Default operator menu Actual tariff account and reset information Meter information menu...
  • Page 20 MUELLER HBMAG Electromagnetic Flow Meters 6.4 OUTPUT OPERATOR MENU The operator menu consists of several indexes described in the following. Operator menu Meter menu Service menu Data logger menu Index: Info Index: Info Index: Info Index: Info 1: Battery reset...
  • Page 21 MUELLER HBMAG Electromagnetic Flow Meters INDEX 2 disappear when alarm condition has been corrected and they are reset via communication interface. Totalizer 2 Fault evaluation and service guidelines are made in service section. After all faults have disappeared, display shows total hours of faults until meter was reset.
  • Page 22: Data Protection

    MUELLER HBMAG Electromagnetic Flow Meters DISPLAY TEST Call-up reset window (index 0) is only shown when call-up function is activated. "r" indicates that it can be reset by a long press on key. When releasing key while "r" is flashing, call-up function will be reset and window disappears.
  • Page 23: Internal Data Handling

    Status information can only be reset while hardware lock key 6.7 BATTERY-POWERED OPERATION is mounted. The HbMAG is factory-configured for 6 years of typical operation on internal battery pack (1 D-cell). High or low temperature, frequent use of IrDA communication, high pulse output rate, and high excitation frequency in leakage detection mode will reduce actual operation time.
  • Page 24 Temperature is also measured to control and adjust its influence on the battery capacity. For the HbMAG, the internal battery pack has a nominal capacity of 33 Ah giving a typical operation time of 6 years in a revenue application. Nominal capacity of external battery packs is 66 Ah and operation time is limited to battery lifetime - typically 10 years.
  • Page 25: Service And Maintenance

    7.2 HBMAG SERVICE GUIDELINES BATTERY CONFIGURATION The HbMAG battery-operated water meter is based on a very Battery figures (generated as customer parameter list, see reliable measurement technology and the advanced alarm section Customer-selected parameter list in chapter Setting basic monitoring and diagnostics provide valuable information parameters (Page 43) show power management information.
  • Page 26: Replacing Transmitter Or Pcb Board

    Check O-ring for damage or deformity. – Smear O-ring with acid-free lubricating gel. Since the HbMAG does not have a removable EEprom, special care Push locking tab and loosen strip. must be taken when replacing a damaged or defective transmitter or PCB board to ensure proper operation and continued accuracy.
  • Page 27: Power Up With Battery Reset, Date And Time Set Up

    MUELLER HBMAG Electromagnetic Flow Meters 7.5 POWER UP WITH BATTERY RESET, DATE 7.6 VERIFICATION AND TIME SET UP Verification mode increases measurement frequency to provide maximum measurements per second. This function is especially useful to minimize calibration rig time when validating flowmeter METER VERSION accuracy.
  • Page 28: Using Sealing

    Electromagnetic Flow Meters 7.7 USING SEALING ADDITIONAL SUPPORT Please contact your local Mueller representative if you have The HbMAG can be sealed to provide tamper detection. additional questions call Customer Care at 1-800-423-1323. Sealing device 7.9 RETURN PROCEDURES Contact Mueller Customer Care at 1-800-423-1323 or your local Mueller sales representative or distributor to arrange all product returns and dispositions.
  • Page 29 MUELLER HBMAG Electromagnetic Flow Meters FAULT NAME / TEXT DESCRIPTION CAUSE REMEDY CODES Error in the coil circuit. Coil • Coils have been current has not settled within disconnected could be Coil current Check cable and wiring installation. Alarm remains active until the specified period.
  • Page 30 MUELLER HBMAG Electromagnetic Flow Meters FAULT NAME / TEXT DESCRIPTION CAUSE REMEDY CODES Change volume per pulse to a higher value-see Technical data Duty cycle of output A has for pulse selection. Reminder: basic version is limited to 50 Hz •...
  • Page 31: Built-In Functions

    8.3 FLOW SIMULATIONS Default settings for Electrode Impedance Limit is 25,000 ohm The HbMAG has a built-in flow simulator (FT551 & FT552) corresponding to a water conductivity of 20 µS/cm (10 000 to verify and adjust pulse output to any connected device or ohm ≈...
  • Page 32: Technical Data

    1.5625 Hz for sensor size 28" ... 48" Liner EPDM Electrode and Hastelloy C276 grounding FIGURE 9 - 1 3" HBMAG W/ INTEGRAL DISPLAY For further features, see features list in appendix (Page 57). METER HBMAG Standard Accuracy ± 0.4% of rate ± 2 mm/s...
  • Page 33: Transmitter

    MUELLER HBMAG Electromagnetic Flow Meters 9.3 TRANSMITTER TRANSMITTER HBMAG Integral (compact) or remote with factory-mounted cable in 25' lengths with Installation NEMA 6P/ IP68 connectors. Connection is made at the transmitter bottom. Top housing Stainless steel (AISI 316) Material Bottom...
  • Page 34: Power Supply

    MUELLER HBMAG Electromagnetic Flow Meters 9.4 POWER SUPPLY TECHNICAL SPECIFICATIONS POWER SUPPLY HBMAG Auto-detection of power source with displayed symbol for remaining power. In battery mode, excitation frequency is manually selected. Internal battery pack 21D_Cell 3.6 V / 16 Ah...
  • Page 35: Output Characteristics

    MUELLER HBMAG Electromagnetic Flow Meters 9.5 OUTPUT CHARACTERISTICS OUTPUT B AS CALL-UP OUTPUT This chapter describes how the HbMAG outputs work. Call up output Call up status PR [s] = PF [Hz] Call up reset On - Off When output B is configured as "call-up", the output is PW 5, 10, 50, 100 &...
  • Page 36 Note: You can select other units than the defaults via the software. The pulse output will only be enabled if the meter selected is ordered with the pulse output as an option. PULSE OUTPUT, VOLUME SELECTION (HBMAG) GUIDELINES FOR MIN.VOLUME PER PULSE AT QN...
  • Page 37: Meter Uncertainty

    Mutual Recognition Arrangement) ensuring international traceability and recognition of the test results worldwide. A calibration certificate is shipped with every sensor and calibration The selected calibration determines the accuracy of the water meter. An HbMAG calibration results in max. ±0.4% of rate ±2 mm/s uncertainty.
  • Page 38: The Effect Of Temperature Hbmag

    Incorporation in pipe section Inlet section 10 x inlet pipe diameter Outlet section 5 x inlet pipe diameter Flow conditions: Fully developed flow profile 9.7 THE EFFECT OF TEMPERATURE HBMAG IMPERIAL (PRESSURES IN PSI) SIZES > 12" TEMPERATURE °F FLANGE SPEC.
  • Page 39: Dimensions And Drawings

    MUELLER HBMAG Electromagnetic Flow Meters 9.8 DIMENSIONS AND DRAWINGS METER DIMENSIONS DIMENSIONS FOR HBMAG 131 (5.16) 161 (6.34) 161 (6.34) 131 (5.16) 115 (4.53) 155 (6.10) 155 (6.10) 104 (4.09) 152 (5.98) 104 (4.09) 152 (5.98) 104 (4.09) DN 25 + 40 (1" + 1 1/2") & DN 350 - DN 1200 (14" - 48") DN 50 - 300 (2 - 12")
  • Page 40 MUELLER HBMAG Electromagnetic Flow Meters REMOTE VERSION 71 (2.80) 31.5 (1.24) 21.5 (0.85) 13 (0.51) 96 (3.78) Ø8 (0.31) 104 (4.09) Ø12 (0.47) Ø8 (0.31) 152 (5.98) 77 (3.03) 158 (6.22) FIGURE 9 - 2 DIMENSIONS IN MM (INCH), WEIGHT 3.5 KG (8 LBS) FLANGE DIMENSIONS ØPCD...
  • Page 41 MUELLER HBMAG Electromagnetic Flow Meters 44 (1.73) 4 (0.16) 4 (0.16) Ø6 (0.31) Ø12 (0.47) 76.5 (3.01) 47 (1.85) EXTERNAL BATTERY PACK Note: Physical orientation of battery pack may influence battery capacity. Optimal battery capacity is achieved with battery pack in an upright position as shown.
  • Page 42: Appendix

    127132.8 A.2 PARAMETER LISTS HbMAG is delivered with factory settings that are not stored as default values. Because defaults values are not present in the meter, an automatic return to factory values is not possible. The default settings are available at www.MuellerSystems.com Display information is indicated in the table by menu and index number.
  • Page 43 MUELLER HBMAG Electromagnetic Flow Meters A.2.1 1 - 99 FT ID METER DISPLAY VIEW PARAMETER / DATA TYPE FACTORY SETTINGS DATA RANGE NUMBER VERSION FIXED PARAMETER OR METER DATA THAT NOT ARE CHANGEABLE Max. 14 characters. Only numbers are visible...
  • Page 44 MUELLER HBMAG Electromagnetic Flow Meters FT ID METER DISPLAY VIEW PARAMETER / DATA TYPE FACTORY SETTINGS DATA RANGE NUMBER VERSION FIXED PARAMETER OR METER DATA THAT NOT ARE CHANGEABLE 0 ... 255, binary presented with information 1 for bit 0...
  • Page 45 MUELLER HBMAG Electromagnetic Flow Meters FT ID METER DISPLAY VIEW PARAMETER / DATA TYPE FACTORY SETTINGS DATA RANGE NUMBER VERSION FIXED PARAMETER OR METER DATA THAT NOT ARE CHANGEABLE Coil current fault appears 2000-01-01 T 00:00:00 Coil current fault disappears...
  • Page 46 MUELLER HBMAG Electromagnetic Flow Meters FT ID METER DISPLAY VIEW PARAMETER / DATA TYPE FACTORY SETTINGS DATA RANGE NUMBER VERSION FIXED PARAMETER OR METER DATA THAT NOT ARE CHANGEABLE Consumption fault appears 2000-01-01 T 00:00:00 Consumption fault disappears 2000-01-01 T 00:00:00...
  • Page 47 T Calibration date Calibration date hours:minutes:seconds Calibration factor Sensor-related HbMAG: 6.25 Hz (DN3 ... 200 (3"... 8") 3.125 Hz (DN250 ... 600 (10" ... 24") Gain correction Sensor-related (DN250 ... 600 (10" ... 24") 1.5625 Hz (DN700 ... 1200 (28"...
  • Page 48 MUELLER HBMAG Electromagnetic Flow Meters FT ID METER DISPLAY VIEW PARAMETER / DATA TYPE FACTORY SETTINGS DATA RANGE NUMBER VERSION FIXED PARAMETER OR METER DATA THAT NOT ARE CHANGEABLE Even 1 stop, Odd 1 stop, None 1 stop, None Parity...
  • Page 49 MUELLER HBMAG Electromagnetic Flow Meters FT ID METER DISPLAY VIEW PARAMETER / DATA TYPE FACTORY SETTINGS DATA RANGE NUMBER VERSION FIXED PARAMETER OR METER DATA THAT NOT ARE CHANGEABLE Repair checksum Yes / No Device Product ID 10779 A.2.7 600 - 799...
  • Page 50 MUELLER HBMAG Electromagnetic Flow Meters FT ID METER DISPLAY VIEW PARAMETER / DATA TYPE FACTORY SETTINGS DATA RANGE NUMBER VERSION FIXED PARAMETER OR METER DATA THAT NOT ARE CHANGEABLE Log period 3 fault status See 613 Log period 3 status information...
  • Page 51 MUELLER HBMAG Electromagnetic Flow Meters FT ID METER DISPLAY VIEW PARAMETER / DATA TYPE FACTORY SETTINGS DATA RANGE NUMBER VERSION FIXED PARAMETER OR METER DATA THAT NOT ARE CHANGEABLE Date of log period 11 Log period 11 totalized (1) Log period 11 totalized (2)
  • Page 52 MUELLER HBMAG Electromagnetic Flow Meters FT ID METER DISPLAY VIEW PARAMETER / DATA TYPE FACTORY SETTINGS DATA RANGE NUMBER VERSION FIXED PARAMETER OR METER DATA THAT NOT ARE CHANGEABLE Log period 18 totalized (2) Log period 18 fault status See 613...
  • Page 53 MUELLER HBMAG Electromagnetic Flow Meters FT ID METER DISPLAY VIEW PARAMETER / DATA TYPE FACTORY SETTINGS DATA RANGE NUMBER VERSION FIXED PARAMETER OR METER DATA THAT NOT ARE CHANGEABLE Log period 25 status information See 614 Date of log period 26...
  • Page 54: Sizing Sensor

    V = Q x 283.67 / (Pipe I.D.)2 (V: [ft/s]; Q: [MGD]; Pipe 10⁶ I.D.: [inch]) 10⁴ A.4 CERTIFICATES A.4.1 CERTIFICATES 10⁴ All certificates are available by calling Mueller Customer Care at 1-800-423-1323. 10³ 10⁴ A.5 SPARE PARTS / ACCESSORIES 10³ A.5.1 ORDERING 10⁴...
  • Page 55 25' Itron in line connector for all Itron Endpoints 25' 3 wire unterminated 22 gauge wire for retrofits SPARE PARTS DESCRIPTION IMAGE Service tool kit package with various components for service and replacement HbMAG compact replacement kit Remote cable set Se HbMAG remote replacement kit...
  • Page 56 (manually selected): – Max. selectable excitation frequency of 1/₁5 Hz – HbMAG default excitation frequency is selected for typically 6 years' operation in a revenue application: 1/₁5 Hz for 3" … 6" 1/30 Hz for 8" … 24" 1/60 Hz for 28" ... 48"...
  • Page 57 MUELLER HBMAG Electromagnetic Flow Meters DATA LOGGER (FT600...FT739) METER STATUS (FT120) • Logging of 26 records: selectable as daily, weekly or Monitoring of important revenue parameters and data: monthly logging • Changing totalizers 1 and 2 • Each logging includes: •...
  • Page 58: Features

    MUELLER HBMAG Electromagnetic Flow Meters DIAGNOSTIC • Continuous self test including – Coil current to drive the magnetic field – Signal input circuit – Data calculation, handling and storing • Features – Alarm statistics and logging for fault analyzing –...
  • Page 59: Index

    MUELLER HBMAG Electromagnetic Flow Meters I. INDEX Laws and directives, 2 Lithium batteries Add-On Modules Safety, 2 Electrical connection, 13 Mains supply, 9 Comissioning Maintenance, 26 With Flow Tool, 13 Material compatibility, 2 Compliance, 2 Contact person, 1 Customer Support Hotline, 27...
  • Page 60 © 2020 Mueller Systems, LLC. All Rights Reserved. The trademarks, logos and service marks displayed in this document are the property of Mueller Water Products, Inc., its affiliates or other third parties. Products above marked with a section symbol (§) are subject to patents or patent applications. For details, visit www.mwppat.com. These products are intended for use in potable water applications. Please contact your Mueller Sales or Customer Service Representative concerning any other application(s).

Table of Contents