Repair instructions
Front frame, complete
Dräger Medical AG & Co. KGaA • EvitaXL
1
General
In case of repair, replace the "complete front frame".
The "front frame, complete", comprises the front frame,
the label (EvitaXL), and the front membrane.
In case of repair, remove the keypad labeling strip
from the faulty front frame and use it in the new front
frame!
1.1
Replacing the front frame
1 Switch off the device.
2 Unplug the power plug of the device from the socket-
outlet.
3 Disconnect the compressed-gas supply hoses from the
central gas supply.
Electrostatic discharge may damage electrostatic
sensitive devices. When handling electrostatic
sensitive devices, use a static-dissipative mat
and a static dissipative wrist strap.
4 Create ESD conditions.
5 Remove the control unit from the interlock of the device
and place it on a static-dissipative mat.
General
1

Advertisement

loading
Need help?

Need help?

Do you have a question about the EvitaXL and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Dräger Medical EvitaXL

  • Page 1 4 Create ESD conditions. 5 Remove the control unit from the interlock of the device and place it on a static-dissipative mat. Dräger Medical AG & Co. KGaA • EvitaXL...
  • Page 2 9 Remove the rear panel and put it aside. Fig. 1 Plug-in connector 10 Pull out the cable connector (Fig. 3/1). 7 Place the control unit on the static-dissipative mat. 8 Remove the screws (Fig. 2/1). Dräger Medical AG & Co. KGaA • EvitaXL...
  • Page 3 (Fig. frame with a sealing; if the sealing sticks to the display, the touchscreen could be unintentionally removed in the process and might fall down!). Dräger Medical AG & Co. KGaA • EvitaXL...
  • Page 4 16 Remove the loudspeaker connector from the Adapter PCB. 17 Remove the rotary transducer connector from the Adapter PCB. 18 Remove the Adapter PCB screws. 19 Remove the Adapter PCB from the faulty front frame and place it aside. Dräger Medical AG & Co. KGaA • EvitaXL...
  • Page 5 24 Remove the nut and the washer (Fig. 7/1). 28 Attach the new “EvitaXL” label to the provided indented space in the top right corner of the front frame. 29 Turn the new front frame around. 30 Secure the Adapter PCB to the new front frame using the screws and washers.
  • Page 6 37 Push the keypad labeling strip from the old frontframe (Fig. 8) into the front membrane as far as it will go (Note: the labeled side must be visible!). Fig. 8 Keypad labeling strip Dräger Medical AG & Co. KGaA • EvitaXL...
  • Page 7 41 Secure the installation frame to the front frame using the screws (Fig. 11/1). Fig. 9 Attaching the front membrane 39 Place the “Touchscreen” (Fig. 10) into the front frame (Note: the rough side of the “Touchscreen” must face outward!). Dräger Medical AG & Co. KGaA • EvitaXL...
  • Page 8 43 Hold the flex cable near the connection socket and push it slowly and carefully into the connection socket of the Adapter PCB (Fig. 12). Dräger Medical AG & Co. KGaA • EvitaXL...
  • Page 9 (Fig. 15/1). 45 Place the rear panel onto the front frame and secure it with the screws (Fig. 14/1). Dräger Medical AG & Co. KGaA • EvitaXL...
  • Page 10 49 Switch the device on. 50 Perform “electrical safety tests” and “function tests” as described in the PMS procedure. For the USA and Canada: test according to CAN/CSA-22.2 No. 601.1- M90. Dräger Medical AG & Co. KGaA • EvitaXL...