Model 84032-FI and Model 84032P-FI Gate
Bias Option, Field Installation Addendum
I. Introduction
The gate bias option provides bias voltage (up to 100 V) for gated Hall measurements, gated IV curves
and FET device characterization. This option includes the Keithley 6487 Picoammeter voltage source,
rack mount, and associated cables.
This addendum provides instructions for installing and connecting your gate bias option. All other
information can be found in the user's manual.
II. Unpacking
The Model 84032-FI is shipped in a box with the rack mount kit. Remove the packaging and set the
Keithley 6487 and all parts in a workspace near the console. The box contains the following:
1 Keithley 6487 Picoammeter voltage source
1 license file update software disc
1 half rack mount kit
1 dual half rack mount kit
Fuses for 100/120V operation and 220/240V operation
1 IEEE-488 cable
1 BNC-to-triaxial adapter
2 power cords
1 interlock adapter cable - 843-016 (Models 8404/8407 only, prior to 02/14/2014)
Lake Shore Cryotronics, Inc. 575 McCorkle Blvd Westerville, Ohio 43082
Tel: (614) 891-2244 Fax: (614) 818-1600 e-mail: info@lakeshore.com
Revision 1.1
P/N 119-845
www.lakeshore.com
September 1, 2015

Advertisement

loading

Summary of Contents for Lakeshore 84032-FI

  • Page 1 II. Unpacking The Model 84032-FI is shipped in a box with the rack mount kit. Remove the packaging and set the Keithley 6487 and all parts in a workspace near the console. The box contains the following: ...
  • Page 2: Installation

    III. Installation Follow these instructions to prepare your console and install the gate bias option. You will need a #2 Phillips head screwdriver to perform this installation. Insert the Line Fuse The fuse drawer supplied with the Keithley 6487 holds the instrument line fuse and line voltage selection module.
  • Page 3 Models 8404/8407 only: Prepare the Location for the Keithley 6487 Picoammeter Voltage Source The system will arrive with blank filler panels on the console, which you will need to remove to prepare the location for the Keithley 6487. Use a #2 Phillips head screwdriver to remove the blank filler panels and ensure there is support for the instrument.
  • Page 4 Instructions: 1. If there is a blank filler panel in the position where the instrument will be installed, use a #2 Phillips head screwdriver to remove it (see Figure 2). 2. Use the support angle located at the bottom of the console (behind the blank filler panels) to support the new instrument: a.
  • Page 5 Model 8425 only: Prepare the Location for the Keithley 6487 Picoammeter Voltage Source The system will arrive with a blank filler panel on the console, which you will need to remove to prepare the location for the Keithley 6487. Use a #2 Phillips head screwdriver to remove the blank filler panel and ensure there is support for the instrument.
  • Page 6: Mount The Instrument

    Mount the Instrument The Keithley 6487 should be installed using one of the provided rack mount kits. The half rack mount kit contains mounting ears, panel, handles (optional), and screws that adapt the front panel to fit into an 88.9 mm (3.5 in) tall full rack space. Refer to Figure 4 for installation details. The dual half rack mount kit contains inner and outer mounting ears, handles (optional), and screws that adapt the front panel to fit into an 88.9 mm (3.5 in) tall full rack space beside an existing instrument.
  • Page 7 Connect the Interlock Adapter Cable (Models 8404/8407 only, prior to 02//14/2014) 1. Lake Shore has discovered that some of the connection cables to the interlock connector on the Keithley 6487 Picoammeter Voltage Source were made incorrectly. The pins outs were not in the right location, but Lake Shore has solved this problem by making an adapter cable with the correct pin outs.
  • Page 8: Make The Electrical Connections

    Make the Electrical Connections The electrical connections for the Keithley 6487 consist of the IEEE-488 cables, banana jumpers, triaxial cable, power cord, and possibly the interlock adapter cable. 1. Verify the instrument arrives with connections as shown in Figure 9. 2.
  • Page 9 Figure 10: Keithley 6487, as installed Prepare the Software Configure the IEEE-488 address for the Keithley 6487. Refer to the Keithley user manual section on GPIB interface to configure the instrument. The address should be set to 22. Then, follow these instructions to install your new software and update the license for the gate bias option: 1.
  • Page 10 Figure 11: New license window 5. Insert the HMS 8400 software DVD into the DVD drive. The DVD contains the license file for the gate bias option. 6. Navigate to the license file, select it, and click Open. 7. When the license file is installed, the software will show the dialog box in Figure 12. 8.
  • Page 11: Testing And Verification

    The most direct and efficient means of contacting Lake Shore is to complete the online service request form at http://www.lakeshore.com/service/Pages/default.aspx. Provide a detailed description of the problem and the required contact information. You will receive a response within 24 hours, or the next business day in the event of weekends or holidays.

This manual is also suitable for: